Disposable screen-printed electrode modified with bismuth-PSS composites as high sensitive sensor for cadmium and lead determination

被引:56
|
作者
Maria-Hormigos, Roberto [1 ]
Jesus Gismera, Ma [1 ]
Procopio, Jesus R. [1 ]
Teresa Sevilla, Ma [1 ]
机构
[1] Univ Autonoma Madrid, Dept Analyt Chem & Instrumental Anal, Fac Sci, Avda Francisco Tomas Y Valiente 7, E-28049 Madrid, Spain
关键词
Bismuth bulk modified electrodes; Screen-printed electrode; Polystyrene sulfonate composites; Differential pulse anodic stripping voltammetry (DPASV); Cd(II) and Pb(II) determination; ANODIC-STRIPPING VOLTAMMETRY; MERCURY FILM ELECTRODES; TRACE-METALS; POLYSTYRENE SULFONATE; CARBON ELECTRODE; GLASSY-CARBON; CD(II); PB(II); OXIDE; NANOSTRUCTURES;
D O I
10.1016/j.jelechem.2016.02.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Bismuth was incorporated in screen-printed carbon electrodes (SPCEs) modified with polystyrene sulfonate (PSS) and carbon nanopowder (CnP) composites. Different strategies, based on bulk approaches using bismuth oxide particles and bismuth (III) solutions, were assayed. The features of the modified electrodes were evaluated by differential pulse anodic stripping voltammetry (DPASV) of Pb(II) and Cd(II) solutions. The best results were obtained when bismuth was incorporated as bismuth oxide particles during the preparation of the PSS-CnP aqueous suspension used to modify the electrodes. Using this optimal modified sensor, the DPASV measurement conditions for Pb(II) and Cd(II) determination were optimized and their figures of merit were evaluated. Measurements were performed using two experimental approaches: the "drop method", by putting the test solution on the surface of the modified SPCE, and the "immersion method" performed by immersing the device in a stirred test solution. The limits of detection (S/N = 3) were 0.27 mu g L-1 (1.3 x 10(-9) M) for Pb(II) and 0.10 mu g L-1 (9.0 x 10(-10) M) for Cd(II), using the "drop method", and 0.029 mu g L-1 (1.4 x 10(-10) M) for Pb(II) and 0.012 mu g L-1 (1.1 x 10(-19) M) for Cd(II) employing the immersion procedure. The optimal bismuth modified SPCE was used for Pb(II) and Cd(II) determination in natural water samples with successful results. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 122
页数:9
相关论文
共 50 条
  • [31] Disposable screen-printed bismuth electrode modified with multi-walled carbon nanotubes for electrochemical stripping measurements
    Shanghai Key Laboratory of Functional Materials Chemistry, Research Centre of Analysis, Test, East China University of Science and Technology, No. 130 Meilong Road, Shanghai 200237, China
    不详
    Anal. Sci., 12 (1237-1241):
  • [32] Disposable Screen-printed Bismuth Electrode Modified with Multi-walled Carbon Nanotubes for Electrochemical Stripping Measurements
    Xiangheng Niu
    Hongli Zhao
    Minbo Lan
    Analytical Sciences, 2011, 27 : 1237 - 1241
  • [33] Disposable Screen-printed Bismuth Electrode Modified with Multi-walled Carbon Nanotubes for Electrochemical Stripping Measurements
    Niu, Xiangheng
    Zhao, Hongli
    Lan, Minbo
    ANALYTICAL SCIENCES, 2011, 27 (12) : 1237 - 1241
  • [34] Disposable nitrite sensor based on hemoglobin-colloidal gold nanoparticle modified screen-printed electrode
    Xu, XX
    Liu, SQ
    Li, B
    Ju, HX
    ANALYTICAL LETTERS, 2003, 36 (11) : 2427 - 2442
  • [35] Disposable screen-printed enzyme sensor for simultaneous determination of starch and glucose
    Hu, T
    Zhang, XE
    Zhang, ZP
    BIOTECHNOLOGY TECHNIQUES, 1999, 13 (06) : 359 - 362
  • [36] High Sensitive On-Site Cadmium Sensor Based on AuNPs Amalgam Modified Screen-Printed Carbon Electrodes
    Zhang, Lei
    Li, Da-Wei
    Song, Wei
    Shi, Lei
    Li, Yang
    Long, Yi-Tao
    IEEE SENSORS JOURNAL, 2010, 10 (10) : 1583 - 1588
  • [37] Sensitive electrochemical detection of horseradish peroxidase at disposable screen-printed carbon electrode
    Lee, Ai-Cheng
    Liu, Guodong
    Heng, Chew-Kiat
    Tan, Swee-Ngin
    Lim, Tit-Meng
    Lin, Yuehe
    ELECTROANALYSIS, 2008, 20 (18) : 2040 - 2046
  • [39] Disposable sputtered-bismuth screen-printed sensors for voltammetric monitoring of cadmium and lead in atmospheric particulate matter samples
    Palomo-Marin, M. R.
    Rueda-Holgado, F.
    Marin-Exposito, J.
    Pinilla-Gil, E.
    TALANTA, 2017, 175 : 313 - 317
  • [40] Selenocystine modified screen-printed electrode as an alternative sensor for the voltammetric determination of metal ions
    Puy-Llovera, Jaume
    Perez-Rafols, Clara
    Serrano, Nutiria
    Diaz-Cruz, Jose Manuel
    Arino, Cristina
    Esteban, Miguel
    TALANTA, 2017, 175 : 501 - 506