Improving dose delivery accuracy with EPID in vivo dosimetry: results from a multicenter study

被引:19
|
作者
Esposito, M. [1 ]
Piermattei, A. [2 ]
Bresciani, S. [3 ]
Orlandini, L. C. [4 ]
Falco, M. D. [5 ]
Giancaterino, S. [5 ]
Cilla, S. [6 ]
Ianiro, A. [6 ]
Nigro, R. [7 ]
Botez, L. [3 ]
Riccardi, S. [7 ]
Fidanzio, A. [2 ]
Greco, F. [2 ]
Villaggi, E. [8 ]
Russo, S. [9 ]
Stasi, M. [10 ]
机构
[1] Azienda Sanitaria USL Toscana Ctr Florence, Med Phys Unit, SC Fis Sanitaria Firenze Empoli, Radiat Oncol Dept, Via DellAntella 58, I-50012 Florence, Italy
[2] Fdn Policlin Univ A Gemelli IRCCS, Radioterapia Oncol Ematol, UOC Fis Sanitaria, Rome, Italy
[3] Candiolo Canc Inst FPO IRCCS, Med Phys, Turin, Italy
[4] Sichuan Canc Hosp, Dept Radiat Oncol, Chengdu, Peoples R China
[5] Univ G dAnnunzio, Dipartimento Radioterapia, Chieti, Italy
[6] Fdn Ric & Cura Giovanni Paolo II, Med Phys Unit, Campobasso, Italy
[7] OGP S Camillo Lellis, Rieti, Italy
[8] AUSL Piacenza, Piacenza, Italy
[9] Azienda Sanitaria USL Toscana Ctr Florence, SC Fis Sanitaria Firenze Empoli, Florence, Italy
[10] SC Fis Sanitaria, AO Ordine Mauriziano Torino, Turin, Italy
关键词
Error detection; Quality assurance; Portal dosimetry; Online dosimetry; Online measurements;
D O I
10.1007/s00066-021-01749-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose To investigate critical aspects and effectiveness of in vivo dosimetry (IVD) tests obtained by an electronic portal imaging device (EPID) in a multicenter and multisystem context. Materials and methods Eight centers with three commercial systems-SoftDiso (SD, Best Medical Italy, Chianciano, Italy), Dosimetry Check (DC, Math Resolution, LCC), and PerFRACTION (PF, Sun Nuclear Corporation, SNC, Melbourne, FL)-collected IVD results for a total of 2002 patients and 32,276 tests. Data are summarized for IVD software, radiotherapy technique, and anatomical site. Every center reported the number of patients and tests analyzed, and the percentage of tests outside of the tolerance level (OTL%). OTL% was categorized as being due to incorrect patient setup, incorrect use of immobilization devices, incorrect dose computation, anatomical variations, and unknown causes. Results The three systems use different approaches and customized alert indices, based on local protocols. For Volumetric Modulated Arc Therapy (VMAT) treatments OTL% mean values were up to 8.9% for SD, 18.0% for DC, and 16.0% for PF. Errors due to "anatomical variations" for head and neck were up to 9.0% for SD and DC and 8.0% for PF systems, while for abdomen and pelvis/prostate treatments were up to 9%, 17.0%, and 9.0% for SD, DC, and PF, respectively. The comparison among techniques gave 3% for Stereotactic Body Radiation Therapy, 7.0% (range 4.7-8.9%) for VMAT, 10.4% (range 7.0-12.2%) for Intensity Modulated Radiation Therapy, and 13.2% (range 8.8-21.0%) for 3D Conformal Radiation Therapy. Conclusion The results obtained with different IVD software and among centers were consistent and showed an acceptable homogeneity. EPID IVD was effective in intercepting important errors.
引用
收藏
页码:633 / 643
页数:11
相关论文
共 50 条
  • [21] Radiotherapy delivery error detection with electronic portal imaging device (EPID)-based in vivo dosimetry
    Yang, Xueying
    Geng, Lisheng
    Huang, David
    Li, Kaiwen
    Zhuang, Hongqing
    Cai, Jing
    Yang, Ruijie
    CHINESE MEDICAL JOURNAL, 2023, 136 (08) : 998 - 1000
  • [22] 3D DOSE VERIFICATION FOR LUNG CANCER TREATMENTS BY IN AQUA VIVO EPID DOSIMETRY
    Wendling, M.
    McDermott, L.
    Mans, A.
    Sonke, J. J.
    van Herk, M.
    Mijnheer, B.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S160 - S160
  • [23] Statistical analysis of epid-based in-vivo dosimetry results for 3DCRT
    Fidanzio, A.
    Azario, L.
    Greco, F.
    Placidi, E.
    Piermattei, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S64 - S64
  • [24] Evaluation of EPID based transit dosimetry on the implementation of In-vivo dosimetry and a step forward for dose guided adaptive radiation therapy
    Wilson, D.
    Sankar, A.
    Warnock, J.
    Goodall, L.
    EUROPEAN JOURNAL OF CANCER, 2013, 49 : S226 - S226
  • [25] Clinical Results of an EPID-Based In-vivo Dosimetry for Lung Cancer Radiation Therapy
    Giancaterino, Stefano
    De Nicola, Andrea
    Adorante, Nico
    Di Tommaso, Monica
    Trignani, Marianna
    Nuzzo, Marianna
    Genovesi, Domenico
    Falco, Maria Daniela
    JOURNAL OF THORACIC ONCOLOGY, 2017, 12 (01) : S1510 - S1511
  • [26] Clinical results of an EPID-based in-vivo dosimetry for prostate cancer treated by VMAT
    Falco, M. D.
    Giancaterino, S.
    De Nicola, A.
    Perrotti, F.
    Menna, S.
    Fidanzio, A.
    Piermattei, A.
    Genovesi, D.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S706 - S706
  • [27] A novel method for sub-arc VMAT dose delivery verification based on portal dosimetry with an EPID
    Cools, Ruud A. M.
    Dirkx, Maarten L. P.
    Heijmen, Ben J. M.
    MEDICAL PHYSICS, 2017, 44 (11) : 5556 - 5562
  • [28] A Preliminary Experience of Using EPID Transit Dosimetry for Monitoring Daily Dose Variations in Radiation Treatment Delivery
    Yao, R.
    Chisela, W.
    MEDICAL PHYSICS, 2015, 42 (06) : 3345 - 3345
  • [29] EPID-BASED IN VIVO DOSIMETRY SYSTEM FOR SBRT-VMAT: MEASURED VERSUS PLANNED DOSE
    McCowan, Peter M.
    van Beek, Timothy
    vanUytven, Eric
    Butler, Jim
    Loewen, Shaun K.
    Nashed, Maged
    Quon, Harvey
    McCurdy, Boyd
    RADIOTHERAPY AND ONCOLOGY, 2016, 120 : S2 - S3
  • [30] Evaluation of dose delivery accuracy of Gamma Knife by polymer gel dosimetry
    Watanabe, Yoichi
    Akimitsu, Tomohide
    Hirokawa, Yutaka
    Mooij, Rob B.
    Perera, G. Mark
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (03): : 133 - 142