Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems

被引:15
|
作者
Alshaikhi, Jailan [1 ,2 ,3 ]
Doolan, Paul J. [4 ]
D'Souza, Derek [2 ]
Holloway, Stacey McGowan [1 ,5 ]
Amos, Richard A. [1 ]
Royle, Gary [1 ]
机构
[1] UCL, Dept Med Phys & Biomed Engn, London, England
[2] Univ Coll London Hosp NHS Fdn Trust, Dept Radiotherapy Phys, London, England
[3] Saudi Particle Therapy Ctr, Riyadh, Saudi Arabia
[4] German Oncol Ctr, Dept Med Phys, Limassol, Cyprus
[5] NIHR Univ Coll London Hosp Biomed Res Ctr, London, England
关键词
particle therapy; proton therapy; treatment planning; MONTE-CARLO; THERAPY; RADIOTHERAPY; INTERPLAY;
D O I
10.1002/mp.13382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose In pencil beam scanning proton therapy, target coverage is achieved by scanning the pencil beam laterally in the x- and y-directions and delivering spots of dose to positions at a given radiological depth (layer). Dose is delivered to the spots on different layers by pencil beams of different energy until the entire volume has been irradiated. The aim of this study is to investigate the implementation of proton planning parameters (spot spacing, layer spacing and margins) in four commercial proton treatment planning systems (TPSs): Eclipse, Pinnacle(3), RayStation and XiO. Materials and Methods Using identical beam data in each TPS, plans were created on uniform material synthetic phantoms with cubic targets. The following parameters were systematically varied in each TPS to observe their different implementations: spot spacing, layer spacing and margin. Additionally, plans were created in Eclipse to investigate the impact of these parameters on plan delivery and optimal values are suggested. Results It was found that all systems except Eclipse use a variable layer spacing per beam, based on the Bragg peak width of each energy layer. It is recommended that if this cannot be used, then a constant value of 5 mm will ensure good dose homogeneity. Only RayStation varies the spot spacing according to the variable spot size with depth. If a constant spot spacing is to be used, a value of 5 mm is recommended as a good compromise between dose homogeneity, plan robustness and planning time. It was found that both Pinnacle(3) and RayStation position spots outside of the defined volume (target plus margin). Conclusions All four systems are capable of delivering uniform dose distributions to simple targets, but their implementation of the various planning parameters is different. In this paper comparisons are made between the four systems and recommendations are made as to the values that will provide the best compromise in dose homogeneity and planning time. (c) 2019 American Association of Physicists in Medicine
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [41] Effects of spot size errors in DynamicARC pencil beam scanning proton therapy planning
    Rana, Suresh
    Rosenfeld, Anatoly B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (23):
  • [42] A method for improving simplified pencil beam dose in IMRT treatment planning
    Riegel, A
    Chen, Y
    Galvin, J
    MEDICAL PHYSICS, 2004, 31 (06) : 1856 - 1856
  • [43] Dosimetric Impact of Dose Rate Fall-Off in FLASH Proton Pencil Beam Scanning Treatment
    Bentefour, E. H.
    Sawant, A.
    Mossahebi, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : S32 - S33
  • [44] Robust Evaluation of Treatment Planning Techniques in Pencil Beam Scanning Proton Therapy for Cranio-Spinal Irradiation
    Rana, S.
    Romaguera, T.
    Gutierrez, A.
    Bennouna, J.
    MEDICAL PHYSICS, 2017, 44 (06) : 2866 - 2867
  • [45] Accurate proton treatment planning for pencil beam crossing titanium fixation implants
    Righetto, Roberto
    Clemens, Luis Pazos
    Lorentini, Stefano
    Fracchiolla, Francesco
    Algranati, Carlo
    Tommasino, Francesco
    Dionisi, Francesco
    Cianchetti, Marco
    Schwarz, Marco
    Farace, Paolo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 : 28 - 38
  • [46] Dosimetry and Dose Rate Evaluation of Transmission Proton Pencil Beam Scanning FLASH Treatment Planning for Hypofractionated Lung Cases: A Comparison Study with Intensity Modulated Proton Therapy
    Wei, S.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [47] Measuring the Impact of Interplay on Target Dose Homogeneity in Pencil Beam Scanning Proton Irradiation
    El-Sherif, O.
    Johnson, J.
    Kruse, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E484 - E484
  • [48] Impact of dose engine algorithm in pencil beam scanning proton therapy for breast cancer
    Tommasino, Francesco
    Fellin, Francesco
    Lorentini, Stefano
    Farace, Paolo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 50 : 7 - 12
  • [49] Quantitative Assessment of 3D Dose Rate for Proton Pencil Beam Scanning FLASH Radiotherapy and Its Application for Lung Hypofractionation Treatment Planning
    Kang, Minglei
    Wei, Shouyi
    Choi, J. Isabelle
    Simone, Charles B.
    Lin, Haibo
    CANCERS, 2021, 13 (14)
  • [50] REDUCING DOSE TO DENTOFACIAL STRUCTURES: A PENCIL-BEAM SCANNING PROTON THERAPY AND VOLUMETRIC-ARC THERAPY PHOTON TREATMENT PLANNING COMPARISON STUDY
    Davies, Lucy Siew Chen
    Charlwood, Frances
    Hol, Marinka
    Aznar, Marianne C.
    Davey, Angela
    Eccles, Cynthia
    Foster-Thomas, Emma
    Gaito, Simona
    Gaze, Mark
    Indelicato, Danny
    Mandeville, Henry
    Slater, Olga
    Sitch, Peter
    Tang, Vivian
    Whitfield, Gillian
    Pan, Shermaine
    PEDIATRIC BLOOD & CANCER, 2023, 70 : S68 - S69