Benefits of ambulatory axillary intra-aortic balloon pump for circulatory support as bridge to heart transplant

被引:53
|
作者
Umakanthan, Ramanan [1 ]
Hoff, Steven J. [1 ]
Solenkova, Natalia [1 ]
Wigger, Mark A. [1 ]
Keebler, Mary E. [1 ]
Lenneman, Andrew [1 ]
Leacche, Marzia [1 ]
DiSalvo, Thomas G. [1 ]
Ooi, Henry [1 ]
Naftilan, Allen J. [1 ]
Byrne, John G. [1 ]
Ahmad, Rashid M. [1 ]
机构
[1] Vanderbilt Heart & Vasc Inst, Dept Cardiac Surg, Nashville, TN 37232 USA
来源
关键词
ARTERY;
D O I
10.1016/j.jtcvs.2012.02.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Axillary intra-aortic balloon pump therapy has been described as a bridge to transplant. Advantages over femoral intra-aortic balloon pump therapy include reduced incidence of infection and enhanced patient mobility. We identified the patients who would benefit most from this therapy while awaiting heart transplantation. Methods: We conducted a single-center, retrospective observational study to evaluate outcomes from axillary intra-aortic balloon pump therapy. These included hemodynamic parameters, duration of support, and success in bridging to transplant. We selected patients on the basis of history of sternotomy, elevated panel-reactive antibody, and small body habitus. Patients were made to ambulate aggressively beginning on postoperative day 1. Results: Between September 2007 and September 2010, 18 patients underwent axillary intra-aortic balloon pump therapy. All patients had the devices placed through the left axillary artery with a Hemashield side graft (Boston Scientific, Natick, Mass). Before axillary placement, patients underwent femoral placement to demonstrate hemodynamic benefit. Duration of support ranged from 5 to 63 days (median = 19 days). There was marked improvement in ambulatory potential and hemodynamic parameters, with minimal blood transfusion requirements. There were no device-related infections. Some 72% of the patients (13/18) were successfully bridged to transplantation. Conclusions: Axillary intra-aortic balloon pump therapy provides excellent support for selected patients as a bridge to transplant. The majority of the patients were successfully bridged to transplant and discharged. Although this therapy has been described in previous studies, this is the largest series to incorporate a regimen of aggressive ambulation with daily measurements of distances walked. (J Thorac Cardiovasc Surg 2012; 143: 1193-7)
引用
收藏
页码:1193 / 1197
页数:5
相关论文
共 50 条
  • [41] LEFT HEART ASSIST SYSTEM - INTRA-AORTIC BALLOON PUMP
    BUCKLEY, MJ
    LAIRD, JD
    MADRAS, PN
    JONES, RT
    KANTROWITZ, AR
    AUSTEN, WG
    SURGICAL CLINICS OF NORTH AMERICA, 1969, 49 (03) : 505 - +
  • [42] CIRCULATORY SUPPORT BY INTRA-AORTIC BALLOON COUNTERPULSATION - RESULTS AND PRESENT INDICATIONS
    DEBRUX, JL
    BARBIERBOHM, G
    MALBEZIN, S
    HVASS, U
    PANSARD, Y
    CHASPOUX, P
    HADDANI, J
    LANGLOIS, J
    COEUR, 1984, 15 (06): : 739 - 747
  • [43] Location of an intra-aortic balloon pump
    Fletcher, SN
    Riedel, BJCJ
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2001, 15 (02) : 267 - 268
  • [44] Intra-aortic Balloon Pump Securement
    Holcomb, Brandi
    CRITICAL CARE NURSE, 2022, 42 (06) : 79 - 81
  • [45] Intra-Aortic Balloon Pump Support in the Isolated Beating Porcine Heart in Nonischemic and Ischemic Pump Failure
    Schampaert, Stephanie
    van Nunen, Lokien X.
    Pijls, Nico H. J.
    Rutten, Marcel C. M.
    van Tuijl, Sjoerd
    van de Vosse, Frans N.
    van 't Veer, Marcel
    ARTIFICIAL ORGANS, 2015, 39 (11) : 931 - 938
  • [46] Is the intra-aortic balloon pump leaking?
    Sjauw, Krischan D.
    Piek, Jan J.
    LANCET, 2013, 382 (9905): : 1616 - 1617
  • [47] INTRA-AORTIC BALLOON PUMP THERAPY
    PURCELL, JA
    PIPPIN, L
    MITCHELL, M
    AMERICAN JOURNAL OF NURSING, 1983, 83 (05) : 776 - 790
  • [48] The answer is an intra-aortic balloon pump
    Takami, Y
    Masumoto, H
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2005, 129 (05): : 1195 - 1196
  • [49] Minimally Invasive Insertion of Axillary Intra-Aortic Balloon Pump is Safe and Effective for Advanced Heart Failure
    Jain, C. Charles
    Rosenbaum, Andrew
    El Sabbagh, Abdallah
    Behfar, Atta
    JOURNAL OF CARDIAC FAILURE, 2019, 25 (08) : S174 - S174
  • [50] Percutaneous Left Axillary Artery Placement of Intra-Aortic Balloon Pump in Advanced Heart Failure Patients
    Bhimaraj, Arvind
    Agrawal, Tanushree
    Duran, Antonio
    Tamimi, Omar
    Amione-Guerra, Javier
    Trachtenberg, Barry
    Guha, Ashrith
    Hussain, Imad
    Kim, Ju
    Kassi, Mahwash
    Xu, Jiaqiong
    Suarez, Erik
    Ngo, Uy Q.
    Torre-Amione, Guillermo
    Estep, Jerry D.
    JACC-HEART FAILURE, 2020, 8 (04) : 313 - 323