Biomechanical Comparison of Fully Threaded Solid Cortical Versus Partially Threaded Cannulated Cancellous Screw Fixation for Lisfranc Injuries

被引:15
|
作者
Rozell, Joshua C. [1 ]
Chin, Matthew [1 ]
Donegan, Derek J. [1 ]
Hast, Michael W. [1 ]
机构
[1] Univ Penn, Dept Orthopaed Surg, Biedermann Lab Orthopaed Res, 3737 Market St,Ste 1050,10th Fl, Philadelphia, PA 19104 USA
关键词
FRACTURE-DISLOCATIONS; TARSOMETATARSAL JOINT; INTERNAL-FIXATION; MANAGEMENT; REDUCTION;
D O I
10.3928/01477447-20180103-03
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
There currently exists an array of operative strategies to manage Lisfranc injuries. Modular fixation systems present surgeons with a choice between fully threaded solid cortical (FSC) and partially threaded cannulated cancellous (PCC) bone screws when using a transarticular screw approach. It is currently unknown how screw design influences fixation strength in Lisfranc reconstructions. The purpose of this study was to evaluate the biomechanical differences of FSC and PCC screws using a cadaveric model of a simulated Lisfranc injury and controlled benchtop experiments. Ten matched pairs of cadaveric feet received an acute Lisfranc injury and were repaired with FSC or PCC screws. Diastasis was measured between the medial and intermediate cuneiforms and the first and second metatarsals during simulations of partial weight bearing. Three-point bending and axial pull-out tests were performed to characterize screw mechanics that could not be measured within the cadaveric model. Screw design did not affect cuneiform or metatarsal diastasis. Neither screw loosening nor deformation was observed following cadaveric testing. Bending tests indicated FSC screws had higher ultimate strength, but there was no significant difference in yield load. Partially threaded cannulated cancellous bone screws exhibited superior axial pull-out strength. Fully threaded solid cortical and PCC screws provide equal amounts of fixation strength during partial weight bearing and similar resistance to deformation under bending loads. Partially threaded cannulated cancellous screws may simplify the operative procedure and minimize nonoptimal screw placement. If a clinician so desires, PCC screws may be used in lieu of FSC screws without sacrificing fixation strength.
引用
收藏
页码:E222 / E227
页数:6
相关论文
共 40 条
  • [1] Biomechanical comparison of the combined partially threaded cancellous lag screw and partially threaded cancellous lag screw
    Xu, Daqiang
    Sun, Peidong
    Wang, Jian
    Zhao, Weidong
    Yang, Huilin
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (02): : 2574 - 2581
  • [2] Fully Threaded Versus Partially Threaded Screws: Determining Shear in Cancellous Bone Fixation
    Downey, Michael W.
    Kosmopoulos, Victor
    Carpenter, Brian B.
    Journal of Foot & Ankle Surgery, 2015, 54 (06): : 1021 - 1024
  • [3] Biomechanical comparison of a fully threaded, variable pitch screw and a partially threaded lag screw for internal fixation of type II dens fractures
    Magee, William
    Hettwer, Werner
    Badra, Mohammed
    Bay, Brian
    Hart, Robert
    SPINE, 2007, 32 (17) : E475 - E479
  • [4] Comparison of Partially Threaded and Fully Threaded 4mm Cancellous Screws in Fixation of Medial Malleolar Fractures
    Sayyed-Hosseinian, Sayyed-Hadi
    Bagheri, Farshid
    Ebrahimzadeh, Mohammad Hosein
    Moradi, Ali
    Golshan, Sogol
    ARCHIVES OF BONE AND JOINT SURGERY-ABJS, 2020, 8 (06): : 710 - 715
  • [5] Fully threaded headless compression screw versus partially threaded cancellous lag screw in medial malleolus fractures: clinical and radiological outcomes
    Tugrul Bulut
    Merve Gursoy
    Hakan Ertem
    European Journal of Trauma and Emergency Surgery, 2021, 47 : 179 - 185
  • [6] Biomechanical properties of a posterior fully threaded positioning screw for cannulated screw fixation of displaced neck of femur fractures
    Schaefer, Thomas K.
    Spross, Christian
    Stoffel, Karl K.
    Yates, Piers J.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 (11): : 2130 - 2133
  • [7] Fully threaded headless compression screw versus partially threaded cancellous lag screw in medial malleolus fractures: clinical and radiological outcomes
    Bulut, Tugrul
    Gursoy, Merve
    Ertem, Hakan
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2021, 47 (01) : 179 - 185
  • [8] Biomechanical Study of 16-mm Threaded, 32-mm Threaded, and Fully Threaded SCFE Screw Fixation
    Dragoni, Massimiliano
    Heiner, Anneliese Dorothy
    Costa, Simone
    Gabrielli, Armando
    Weinstein, Stuart L.
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2012, 32 (01) : 70 - 74
  • [9] Biomechanical comparison of fully and partially threaded screws for fixation of slipped capital femoral epiphysis
    Miyanji, Firoz
    Mahar, Andrew
    Oka, Richard
    Pring, Maya
    Wenger, Dennis
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2008, 28 (01) : 49 - 52
  • [10] Lag Screw Fixation of Medial Malleolar Fractures: A Biomechanical, Radiographic, and Clinical Comparison of Unicortical Partially Threaded Lag Screws and Bicortical Fully Threaded Lag Screws
    Ricci, William M.
    Tornetta, Paul
    Borrelli, Joseph, Jr.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (10) : 602 - 606