Quantitative characterization of noise in the measurement chain of an optical particle counter

被引:3
|
作者
Borggräfe, P
Trampe, A
Fissan, H
机构
[1] Gerhard Mercator Univ Duisburg, Dept Proc, D-47048 Duisburg, Germany
[2] Gerhard Mercator Univ Duisburg, Aerosol Measurement Technol Div, D-47048 Duisburg, Germany
关键词
optical particle counter; noise; background light; scattering;
D O I
10.1088/0957-0233/13/4/304
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Optical particle counters are deployed for monitoring possible contamination in gases by particles. The counting and sizing of particles are affected by various kinds of interference encountered during measurement. Two sources of interference are background light in the measurement chamber and electrical noise. Such interference causes measurement errors, which stem from the measurement equipment itself and not from the measured object or from the sampling. For the evaluation of measurement error by the particle counter itself, interference for a typical optical particle counter is identified and its effects on the measured results are discussed qualitatively. For the reduction of measurement error, quantitative determination of interference is important. It is only by doing this that we can judge which kind of interference has the more significant influence on measurement error. A method has been developed which makes it possible to determine individual components of interference. This paper presents an experimental method which allows separation of components of background light consisting of light scattering by the molecules and surface. The results of these measurements can be used to optimize existing particle counters as well as new equipment design.
引用
收藏
页码:445 / 450
页数:6
相关论文
共 50 条
  • [1] AIRCRAFT MEASUREMENT WITH CONDENSATION NUCLEI COUNTER AND OPTICAL-PARTICLE COUNTER
    DREILING, V
    JAENICKE, R
    JOURNAL OF AEROSOL SCIENCE, 1988, 19 (07) : 1045 - 1050
  • [2] Measurement of the fluidic resistance of the MicroMED optical particle counter
    Scaccabarozzi, Diego
    Saggin, Bortolino
    Vaz Junior, Elimar Vieira
    Corti, Marco Giovanni
    Valnegri, Pietro
    Esposito, Francesca
    Cozzolino, Fabio
    Mongelluzzo, Giuseppe
    Franzese, Gabriele
    Porto, Carmen
    Ruggeri, Alan Cosimo
    Molfese, Cesare
    Brienza, Daniele
    Cortecchia, Fausto
    Martin-Ortega, Alberto
    Arruego, Ignacio
    Andres Santiuste, Nuria
    2022 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR AEROSPACE (IEEE METROAEROSPACE 2022), 2022, : 538 - 543
  • [3] Measurement particle-size of the hydrate smoke particles with optical particle counter
    Wang, Wai
    Zhou, Yingchun
    Tan, Tiande
    Nanjing Li Gong Daxue Xuebao/Journal of Nanjing University of Science and Technology, 1994, (03):
  • [4] MEASUREMENT OF PARTICLE SIZE DISTRIBUTIONS IN GAS FLOWS WITH AN OPTICAL PARTICLE COUNTER.
    Buettner, Helmut
    Particle Characterization, 1985, 2 (01): : 20 - 24
  • [5] Optical particle counter measurement of marine aerosol hygroscopic growth
    Snider, J. R.
    Petters, M. D.
    ATMOSPHERIC CHEMISTRY AND PHYSICS, 2008, 8 (07) : 1949 - 1962
  • [7] Optimization of the optical particle counter for online particle measurement in high-pressure gas
    Lu, Lifeng
    Wu, Xiaolin
    Ji, Zhongli
    Xiong, Zhiyi
    Wang, Mingxing
    APPLIED OPTICS, 2019, 58 (02) : 308 - 316
  • [8] Optimization of the optical particle counter for online particle measurement in pressure-changing natural gas
    Song, Xuan
    Wu, Xiaolin
    Ji, Zhongli
    Wang, Deyu
    Liu, Zhen
    APPLIED OPTICS, 2020, 59 (30) : 9581 - 9590
  • [9] LOWER DETECTION LIMIT OF AN OPTICAL-PARTICLE COUNTER IN THE MEASUREMENT OF PARTICLE-SIZE DISTRIBUTIONS
    SACHWEH, B
    BUTTNER, H
    EBERT, F
    PARTICLE & PARTICLE SYSTEMS CHARACTERIZATION, 1989, 6 (03) : 124 - 128
  • [10] Quantum noise limited measurement of particle position with optical tweezers
    Bowen, W. P.
    Tay, J. W.
    Jiang, X.
    2007 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1-4, 2007, : 213 - 214