Calculations of dose point kernels of 64Cu in different media with PENELOPE Monte Carlo code

被引:0
|
作者
Tse, Jason [1 ,2 ]
Geoghegan, Sean [3 ]
机构
[1] Austin Hosp, Med Phys Dept, Melbourne, Vic, Australia
[2] Univ Sydney, Fac Hlth Sci, Sydney, NSW, Australia
[3] Royal Adelaide Hosp, Dept Med Phys, Adelaide, SA, Australia
关键词
dose point kernel; dosimetry; radionuclide therapy; COLLAPSED CONE SUPERPOSITION; SIMULATION; ELECTRON; DOSIMETRY; VALIDATION; ALGORITHM;
D O I
10.1002/mp.13465
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The unique decay properties of copper-64 (Cu-64) has made it a radionuclide of interest in theragnostic applications of nuclear medicine. This study aims to calculate the dose point kernels (DPKs) of Cu-64 in various media with PENELOPE Monte Carlo code. Methods Monte Carlo simulations were performed using PENELOPE code (version 2014). To calculate DPKs, the simulation comprised an isotropic point radiation source positioned at the origin of a spherical object of radius 50 cm. The absorbed dose along the radial direction outwards from the point source were scored with a resolution of 20 mu m. Validations were firstly performed by calculating the DPKs of monoenergetic electrons and photons in water and the results were compared against the literature values. The continuous energy spectra of the beta minus and positron emissions from Cu-64 were numerically modeled and used as inputs to the simulation. DPKs of Cu-64 were calculated in water, soft tissue, lung tissue, and cortical bone, including all emissions types. Results The simulations have been successfully validated against literature values. The largest deviations have been observed with 10 keV monoenergetic electrons with the average and maximum dose difference of -1.01% and -10.56%. The modeled energy spectra closely compared with the average energies from Brookhaven Laboratory National Nuclear Data Centre and the combined spectral shapes from the RAdiation Dose Assessment Resource (RADAR). The DPKs of Cu-64 demonstrated different radial dose deposition in different media owing to the different physical density and effective atomic number. Conclusions The DPKs of Cu-64 have been calculated with Monte Carlo simulations in four different media. They will be useful to study the dosimetric properties of Cu-64-labeled radiopharmaceuticals and perform therapeutic dose planning.
引用
收藏
页码:2422 / 2429
页数:8
相关论文
共 50 条
  • [11] Dose calculations with the BEAM Monte Carlo code at extended SSD's
    Stathakis, S
    Kappas, C
    Papanikolaou, N
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1679 - 1680
  • [12] Cellular lethal damage of 64Cu incorporated in mammalian genome evaluated with Monte Carlo methods
    Carrasco-Hernandez, Jhonatan
    Ramos-Mendez, Jose
    Padilla-Rodal, Elizabeth
    Avila-Rodriguez, Miguel A.
    FRONTIERS IN MEDICINE, 2023, 10
  • [13] Point flux PL-estimators for Monte Carlo dose calculations
    Donskoy, EN
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 3203 - 3205
  • [15] Coincidence summing corrections using PENELOPE/PENNUC Monte Carlo code for volume sources of different densities
    Salpadimos, N.
    Karfopoulos, K.
    Potiriadis, C.
    Seimenis, I.
    APPLIED RADIATION AND ISOTOPES, 2023, 192
  • [16] Dose distributions for CCX, CCA and CIA applicators: Plaque Simulator vs. PENELOPE Monte Carlo code
    Cano-Herranz, A.
    Delgado-Soler, C.
    Anducas-Santiago, N.
    Hermida-Lopez, M.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S181 - S181
  • [17] Different methods of organ equivalent dose scoring in Monte Carlo neutron dose calculations
    Jarlskog, C. Zacharatou
    Lee, C.
    Bolch, W.
    Xu, X.
    Paganetti, H.
    MEDICAL PHYSICS, 2007, 34 (06) : 2437 - 2437
  • [18] Calculation of Tissue Dose Point Kernels Using GATE Monte Carlo Simulation Toolkit to Compare with Water Dose Point Kernel
    Khazaee, M.
    Asl, A. Kamali
    Geramifar, P.
    MEDICAL PHYSICS, 2015, 42 (06) : 3367 - 3367
  • [19] Measurements of dose point kernels using GATE Monte Carlo toolkit for personalized convolution dosimetry
    Tiwari, Ashok
    Graves, Stephen
    Sunderland, John
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [20] Benchmark dose calculations of MCNP5 and PENELOPE codes: Accuracy of low-energy photon cross-sections in Monte Carlo dose calculations
    Ye, S
    Naqvi, S
    Brezovich, I
    Pareek, P
    Popple, R
    MEDICAL PHYSICS, 2004, 31 (06) : 1731 - 1731