Digital pulse shape analysis for the capture-gated liquid scintillator BC-523A

被引:27
|
作者
Flaska, Marek [1 ]
Pozzi, Sara A. [1 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
关键词
Capture-gated detector; Liquid scintillator; Neutron spectroscopy; Special nuclear material; PLASTIC SCINTILLATOR; FAST-NEUTRONS; GAMMA-RAYS; DISCRIMINATION; SPECTROMETER;
D O I
10.1016/j.nima.2008.10.030
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The BC-523A detector is a boron-10-loaded liquid scintillation detector that is sensitive to not only fast neutrons and gamma rays, but also to thermal neutrons. This detector has recently attracted the attention of several researchers and has shown promise for neutron spectroscopy. Several efforts have been made to accurately characterize this detector for several neutron and gamma-ray energies. In this paper, we show, for the first time, that neutron capture pulses can be visually distinguished from neutron scattering pulses by using an offline digital pulse shape discrimination method based on charge integration. In addition, new Monte Carlo simulation and measurement results for time to capture are compared and reasonable agreement is achieved. This work has applications in fields such as nuclear safeguards, nuclear nonproliferation, and national security. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:221 / 225
页数:5
相关论文
共 50 条
  • [41] Comparison of model fitting and gated integration for pulse shape discrimination and spectral estimation of digitized lanthanum halide scintillator pulses
    McFee, J. E.
    Mosquera, C. M.
    Faust, A. A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 828 : 105 - 115
  • [42] Digital pulse shape discrimination between fast neutrons and gamma rays with para-terphenyl scintillator
    Chepurnov, A. S.
    Kirsanov, M. A.
    Klenin, A. A.
    Klimanov, S. G.
    Kubankin, A. S.
    3RD INTERNATIONAL CONFERENCE ON PARTICLE PHYSICS AND ASTROPHYSICS (ICPPA 2017), 2017, 934
  • [43] Compton continuum suppression by digital pulse shape analysis
    Aspacher, B.
    Coldwell, R.L.
    Rester, A.C.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993, A330 (1-2) : 243 - 253
  • [44] Pulse shape analysis of liquid scintillators for neutron studies
    Marrone, S
    Cano-Ott, D
    Colonna, N
    Domingo, C
    Gramegna, F
    Gonzalez, EM
    Gunsing, F
    Heil, M
    Käppeler, F
    Mastinu, PF
    Milazzo, PM
    Papaevangelou, T
    Pavlopoulos, P
    Plag, R
    Reifarth, R
    Tagliente, G
    Tain, JL
    Wisshak, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 490 (1-2): : 299 - 307
  • [45] Neutron measurements on Joint European Torus using an NE213 scintillator with digital pulse shape discrimination
    Esposito, B
    Bertalot, L
    Marocco, D
    Riva, M
    Kaschuck, Y
    Skopintsev, S
    Zimbal, A
    Reginatto, M
    Schuhmacher, H
    Adams, JM
    Murari, A
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3550 - 3552
  • [46] Comparative analysis of pulse shape discrimination methods in a 6Li loaded plastic scintillator
    Balmer, Matthew J. I.
    Gamage, Kelum A. A.
    Taylor, Graeme C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 788 : 146 - 153
  • [47] Compton suppression in BEGe detectors by digital pulse shape analysis
    Mi, Yu-Hao
    Ma, Hao
    Zeng, Zhi
    Cheng, Jian-Ping
    Li, Jun-Li
    Zhang, Hui
    APPLIED RADIATION AND ISOTOPES, 2017, 121 : 96 - 100
  • [48] Hardware Acceleration of Digital Pulse Shape Analysis Using FPGAs
    Gonzalez, Cesar
    Ruiz, Mariano
    Carpeno, Antonio
    Pinas, Alejandro
    Cano-Ott, Daniel
    Plaza, Julio
    Martinez, Trino
    Villamarin, David
    SENSORS, 2024, 24 (09)
  • [49] Digital Pulse-Shape Analysis for Ge Compton Camera
    Fukuchi, Tomonori
    Motomura, Shinji
    Takeda, Sin'ichiro
    Hiromura, Makoto
    Fukunaka, Ayako
    Haba, Hiromitsu
    Watanabe, Yasuyoshi
    Enomoto, Shuichi
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 851 - 852
  • [50] Digital system on partial discharge pulse shape detection and analysis
    Dept. of Elec. Eng., Tsinghua Univ., Beijing 100084, China
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 2001, 41 (03): : 65 - 68