Response characteristics of new perchlorate selective PVC membrane electrode based on a tripodal benzimidazole derivative as a neutral carrier

被引:0
|
作者
Chandra, Sulekh [1 ]
Singh, Inderjeet [2 ]
Tomar, Praveen Kumar [1 ,2 ]
Malik, Amrita [1 ]
Kumar, Avdhesh [1 ]
机构
[1] Univ Delhi, Zakir Husain Delhi Coll, Dept Chem, JL Nehru Marg, New Delhi 110002, India
[2] CCR Coll, Dept Chem, Muzaffarnagar, Uttar Pradesh, India
关键词
Perchlorate selective electrode; PVC membrane; potentiometry; tris[2-(1-ethylbenzimidazolyl)]methylamine; FLOW-INJECTION; ION CHROMATOGRAPHY; COMPLEX; LIQUID; EXTRACTION; IONOPHORES; WATER;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The potentiometric response characteristics of a new perchlorate selective poly(vinyl chloride) membrane electrode based on a tripodal benzimidazole derivative - tris[2-(1-ethylbenzimidazolyl)]methylamine as ionophore was investigated. The electrode exhibited a Nernstian response to perchlorate ions over the activity range of 1.5 x 10(-7) to 1.0 x 10(-1) M with a limit of detection of 5.2 x 10(-8) M between pH range of 2.0-10.0, giving a relatively fast response and reversibility within 5 s. The selectivity coefficient was calculated using fixed interference method (FIM). The electrode worked well for nearly 3 months. The electrode can also be used in partially non-aqueous media having up to 25% (v/v) methanol, ethanol or acetone contents with no significant change in the value of slope or working concentration range. It was successfully applied for the direct determination of perchlorate content in water and human urine samples with satisfactory results.
引用
收藏
页码:79 / 85
页数:7
相关论文
共 50 条
  • [21] Response Characteristics of Copper-Selective Polymer Membrane Electrodes Based on a Newly Synthesized Macrocyclic Calix[4]arene Derivative as a Neutral Carrier Ionophore
    Kamel, Ayman H.
    Mahmoud, Wagiha H.
    Mostafa, Marwa S.
    ELECTROANALYSIS, 2010, 22 (20) : 2453 - 2459
  • [22] BARIUM ION-SELECTIVE ELECTRODE BASED ON A NEW NEUTRAL CARRIER COMPLEX
    BOUKLOUZE, AA
    VIRE, JC
    COOL, V
    ANALYTICA CHIMICA ACTA, 1993, 273 (1-2) : 153 - 163
  • [23] URANYL-ION SELECTIVE ELECTRODE BASED ON A NEW SYNTHETIC NEUTRAL CARRIER
    SENKYR, J
    AMMANN, D
    MEIER, PC
    MORF, WE
    PRETSCH, E
    SIMON, W
    ANALYTICAL CHEMISTRY, 1979, 51 (07) : 786 - 790
  • [24] STUDY OF CALCIUM ION-SELECTIVE ELECTRODE BASED ON A NEW NEUTRAL CARRIER
    Wei XU Bin GUO You Rong NIE Zai Fu HUANG Department of Environmental Science
    ChineseChemicalLetters, 1993, (02) : 179 - 180
  • [25] A tetra-co ordinate nickel(II) complex as neutral carrier for nitrate-selective PVC membrane electrode
    Asghari, AR
    Amini, MK
    Mansour, HR
    Salavati-Niasari, M
    TALANTA, 2003, 61 (04) : 557 - 563
  • [26] A NEW LIQUID-MEMBRANE ELECTRODE FOR SELECTIVE DETERMINATION OF PERCHLORATE
    HASSAN, SSM
    ELSAIED, MM
    TALANTA, 1986, 33 (08) : 679 - 684
  • [27] Lead selective membrane electrode using cryptand(222) neutral carrier
    Amini, MK
    Mazloum-Ardakani, M
    Ensafi, AA
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 364 (08): : 690 - 693
  • [28] Lead selective membrane electrode using cryptand(222) neutral carrier
    M. K. Amini
    M. Mazloum
    A. A. Ensafi
    Fresenius' Journal of Analytical Chemistry, 1999, 364 : 690 - 693
  • [29] Perchlorate Selective PVC Membrane Electrode Based on Cobalt (II) Complex of p-Hydroxy Acetophenone Semicarbazone
    Chandra, Sulekh
    Malik, Amrita
    Tomar, Praveen Kumar
    Kumar, Avdhesh
    Sadwal, Sangeeta
    ANALYTICAL & BIOANALYTICAL ELECTROCHEMISTRY, 2011, 3 (04): : 379 - 392
  • [30] Nickel Ion-Selective PVC Membrane Electrode Based on a New t-Octyl-Calix[6]arene Derivative
    Kamel Belhamel
    Rainer Ludwig
    Mohamed Benamor
    Microchimica Acta, 2005, 149 : 145 - 150