Calibration of the multichannel sensor of a plane angle in dynamic and static modes

被引:0
|
作者
Agapov, MY [1 ]
Bournashev, MN [1 ]
Loukianov, DP [1 ]
Mironov, AV [1 ]
Filatov, YV [1 ]
机构
[1] St Petersburg State Electrotech Univ, Dept Autonomous Navigat Control & Mech, St Petersburg 197376, Russia
关键词
a dynamic goniometric bench; photoelectric angle encoder; interference zero-point unit; dynamical inaccuracy;
D O I
10.1117/12.454657
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The work concerns to the computerized techniques of quality estimation of automated top precision an-le-code converters. Study of instrumental errors of a rotor position sensor (RPS) with use of goniometric rotary bench (GRB) was made, the plane angle accuracy being got from an optical interferential flat-mirror-zero-point module of original design and a photoelectric angle encoder of RON-905 type manufactured by the "Heidenhein" Corporation (Germany) Also a research was made of inaccuracies of the goniometric rotary bench itself, and also the methods have been developed for definition of inaccuracies due to dynamical behavior of measurement of an angle, and both for their algorithmic elimination Study of systematic, random, and dynamical inaccuracies of both the standard sensor RON-905, and the tested uric, RPS, was carried out. Research of dynamic inaccuracy of said angle measurements was made by means of interrelation between angular rate of rotation and changing of the measured angle interval from the interferential zero-label to the certain selected RON label. The latter data gave rise to close algorithmic elimination of a dynamical error arisen from RON own delay time, In this way, it became possible to calibrate the tested sensor in a very broad range of rotation rate and in several modes of rotation with a steady accuracy. The developed techniques provide calibration of sensors in a case of a rotary bench movement with the composite postulated law of a turn angle against time.
引用
收藏
页码:66 / 71
页数:6
相关论文
共 50 条
  • [1] Discrimination of taste qualities using static and dynamic responses of multichannel taste sensor
    Toko, K
    Nakagawa, Y
    Obata, M
    Yahiro, T
    SENSORS AND MATERIALS, 1997, 9 (05) : 297 - 306
  • [2] Static and dynamic calibration of a MEMS calorimetric shear-stress sensor
    Weiss, Julien
    Jondeau, Emmanuel
    Giani, Alain
    Charlot, Benoit
    Combette, Philippe
    SENSORS AND ACTUATORS A-PHYSICAL, 2017, 265 : 211 - 216
  • [3] Auto-calibration of dynamic gas sensor network: Influence of static sensors
    Tsujita, W
    Ishida, H
    Moriizumi, T
    TRANSDUCERS '05, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2005, : 1951 - 1952
  • [4] Absolute calibration of nanoradian angle sensor
    Zhang, SZ
    Kiyono, S
    Uda, Y
    PROCEEDINGS OF THE THIRTEENTH ANNUAL MEETING OF THE AMERICAN SOCIETY FOR PRECISION ENGINEERING, 1998, : 526 - 529
  • [5] Calibration for a Multichannel Magnetic Sensor Array of a Magnetospinography System
    Adachi, Yoshiaki
    Higuchi, Masanori
    Oyama, Daisuke
    Haruta, Yasuhiro
    Kawabata, Shigenori
    Uehara, Gen
    IEEE TRANSACTIONS ON MAGNETICS, 2014, 50 (11)
  • [6] Static calibration and dynamic compensation of the SCORBOT robot using sensor fusion and LSTM networks
    Kuo, Yong-Lin
    Hsieh, Chia-Hang
    JOURNAL OF THE CHINESE INSTITUTE OF ENGINEERS, 2023, 46 (08) : 881 - 894
  • [7] Secondary measurement standard for calibration of dynamic pressure sensor to bridge the gap between existing static and dynamic standards
    Amer, Eynas
    Jonsson, Gustav
    Arrhen, Fredrik
    MEASUREMENT, 2025, 242
  • [8] DYNAMIC OVERSHOOT OF THE STATIC STALL ANGLE
    ERICSSON, LE
    REDING, JP
    JOURNAL OF AIRCRAFT, 1985, 22 (07): : 637 - 638
  • [9] UNCERTAINTY ASSESSMENT IN THE PLANE SOURCE - SENSOR CALIBRATION
    Malinaric, Svetozar
    JOURNAL OF ELECTRICAL ENGINEERING-ELEKTROTECHNICKY CASOPIS, 2009, 60 (02): : 112 - 116
  • [10] Precision angle sensor and its calibration techniques
    Ma, J.S.
    Wang, X.Z.
    Fang, Z.J.
    Zhang, S.Z.
    Satoshi, K.