Monte Carlo calculation of the ionization chamber response to 60Co beam using PENELOPE

被引:17
|
作者
Yi, Chul-Young
Hah, Suck-Ho
Yeom, Min Sun
机构
[1] Korea Res Inst Stand & Sci, Ionizing Radiat Grp, Taejon 305600, South Korea
[2] Korea Inst Sci & Technol, Supercomp Ctr, Dept Supercomp Applicat & Technol, Taejon 305606, South Korea
关键词
Monte Carlo; PENELOPE; ion chamber; cavity dose; Co-60;
D O I
10.1118/1.2188822
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The calculation of the ionization chamber response is one of key factors to develop a primary standard of air kerma. Using Monte Carlo code PENELOPE, we simulated the cavity response of the plane parallel ionization chamber to the monoenergetic Co-60 beam incident normally on a flat surface of the chamber. Two simulation techniques, namely, the uniform interaction technique and the reentrance technique, were introduced. The effect of the input parameters such as C1 (average angular deflection in a single step between hard elastic events), C2 (maximum average fractional energy loss in a step), s(max) (maximum step length) and W-cc (the lower energy of secondary electrons created as a result of a hard collision) on the simulated cavity dose was evaluated. We found that the simulated cavity response of the graphite and solid air chambers obtained by PENELOPE to the monoenergetic Co-60 beam could be consistent with the value expected from the cross-sections of PENELOPE to within 0.2% (one standard deviation) when W-cc and s(max) were selected carefully. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1213 / 1221
页数:9
相关论文
共 50 条
  • [1] Experimental and Monte Carlo evaluation of an ionization chamber in a 60Co beam
    Perini, A. P.
    Neves, L. P.
    Santos, W. S.
    Caldas, L. V. E.
    8TH BRAZILIAN CONGRESS ON METROLOGY (METROLOGIA 2015), 2016, 733
  • [2] Monte Carlo Calculation to Determine Tandem Ionization Chamber Response to X-Rays Beam Using PENELOPE
    Pianoschi, T.
    Goes, E.
    Nicolucci, P.
    Costa, A.
    Pela, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [3] A prototype of an ionization chamber for gamma radiation beams of 60Co: Experimental and Monte Carlo preliminary results
    Vahabi, Seyed Milad
    Zafarghandi, Mojtaba Shamsaie
    Ghasemi, Mohammadreza
    Alipoor, Anita
    Shahvar, Arjang
    RADIATION MEASUREMENTS, 2013, 59 : 284 - 287
  • [4] Calculation of beam quality correction factors for various thimble ionization chambers using the Monte Carlo code PENELOPE
    Erazo, Fabian
    Lallena, Antonio M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (02): : 163 - 170
  • [5] Application of a Pencil Ionization Chamber (0.34 cm3 Volume) for 60Co Beams: Experimental and Monte Carlo Results
    Neves, Lucio P.
    Perini, Ana P.
    Fernandez-Varea, Jose M.
    Caldas, Linda V. E.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) : 746 - 750
  • [6] Accurate simulation of ionisation chamber response with the Monte Carlo code PENELOPE
    Sempau, Josep
    Andreo, Pedro
    XXXIII BRAZILIAN WORKSHOP ON NUCLEAR PHYSICS, 2011, 1351
  • [7] Calculation of scatter calibration factor for 60Co therapy unit with Monte-Carlo method
    Chen, Qiu-Qiu
    Chen, Yong
    Yu, Hong-Liang
    Sun, Liang
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2011, 45 (02): : 230 - 234
  • [8] Chamber-quality factors in 60Co for three plane-parallel chambers for the dosimetry of electrons, protons and heavier charged particles: PENELOPE Monte Carlo simulations
    Panettieri, Vanessa
    Sempau, Josep
    Andreo, Pedro
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (21): : 5917 - 5926
  • [9] Monte Carlo dosimetry of 60Co HDR brachytherapy sources
    Papagiannis, P
    Angelopoulos, A
    Pantelis, E
    Sakelliou, L
    Karaiskos, P
    Shimizu, Y
    MEDICAL PHYSICS, 2003, 30 (04) : 712 - 721
  • [10] Calculation of the effects of electric fields on ionization by electrons emitted in a parallel plate ionization chamber irradiated with 60Co γ-rays
    Takata, Nobuhisa
    RADIATION PHYSICS AND CHEMISTRY, 2020, 170