Early Studies of a Transmission Mechanism for MR-Guided Interventions

被引:0
|
作者
Zhao, Haoran [1 ]
Liu, Xin [2 ]
Zaid, Habib M. [2 ]
Shah, Dipan J. [3 ]
Heffernan, Michael J. [4 ]
Becker, Aaron T. [1 ]
Tsekos, Nikolaos V. [2 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77004 USA
[2] Univ Houston, Dept Comp Sci, Houston, TX 77204 USA
[3] Methodist DeBakey Heart & Vasc Ctr, Cardiovasc MRI Lab, Houston, TX USA
[4] GuidaBot LLC, Houston, TX USA
基金
美国国家科学基金会;
关键词
Solid media transmission; MR compatible; Robotic Actuator; COMPATIBLE ROBOTIC SYSTEM; MAGNETIC-RESONANCE; BRACHYTHERAPY; MOTION;
D O I
10.1109/BIBE.2017.00082
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Magnetic resonance imaging (MRI)-guided, manipulator-assisted interventions have the potential to improve patient outcomes. This work presents a force transmission mechanism, called solid-media transmission (SMT), for actuating manipulators inside MRI scanners. The SMT mechanism is based on conduits filled with spheres and spacers made of a nonmagnetic, nonconductive material that forms a backbone for bidirectional transmission. Early modeling and experimental studies assessed SMT and identified limitations and improvements. Simulations demonstrated the detrimental role of friction, which can be alleviated with a choice of low friction material and long spacers. However, the length of spacer is limited by the desired bending of the conduit. A closed-loop control law was implemented to drive the SMT. The 3rd order system fit ratio is 92.3%. A 1-m long SMT was experimentally tested under this closed-loop controller with heuristically set parameters using a customized benchtop setup. For commanded displacements of 1 to 50 mm, the SMT-actuated 1 degree of freedom stage exhibited sub-millimeter accuracy, which ranged from 0.109 +/- 0.057 mm to 0.045 +/- 0.029 mm depending on the commanded displacement. However, such accuracy required long control times inversely proportional to displacement ranging from 7.56 +/- 1.85s to 2.53 +/- 0.11s. This was attributed to friction as well as backlash which is due to suboptimal packing of the media. In MR studies, a 4-m long SMT-actuated 1 DoF manipulator was powered by a servo motor located inside the scanner room but outside the 5 Gauss line of the magnet. With shielding and filtering, the SNR of MR images during the operation of the servo motor and SMT-actuation was found to be 89 +/- 9% of the control case.
引用
收藏
页码:450 / 456
页数:7
相关论文
共 50 条
  • [1] MR-guided vascular interventions
    Atalar, E
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 4056 - 4056
  • [2] Pediatric MR-guided interventions
    Schulz, T
    Tröbs, RB
    Schneider, JP
    Hirsch, W
    Puccini, S
    Schmidt, F
    Kahn, T
    EUROPEAN JOURNAL OF RADIOLOGY, 2005, 53 (01) : 57 - 66
  • [3] MR-guided prostate interventions
    Tempany, Clare
    Straus, Sarah
    Hata, Nobuhiko
    Haker, Steven
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) : 356 - 367
  • [4] Robotic Systems for MR-guided Interventions
    Bock, M.
    Krafft, A.
    Maier, F.
    Umathum, R.
    Homagk, A. -K.
    Alt, S.
    Zhang, K.
    Pitsaer, C.
    Jenne, J.
    Semmler, W.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 6, 2009, 25 : 114 - +
  • [5] MR-guided mechanically assisted interventions
    Fichtinger, G.
    MEDICAL PHYSICS, 2006, 33 (06) : 2198 - 2198
  • [6] Toward MR-guided coronary interventions
    Serfaty, J
    Yang, X
    Aksit, P
    Solaiyappan, M
    Atalar, E
    RADIOLOGY, 2000, 217 : 422 - 422
  • [7] Advancements in MRI and MR-Guided Interventions
    Cai, Jing
    Chen, N.
    Cai, J.
    Jackson, E.
    MEDICAL PHYSICS, 2017, 44 (06) : 3313 - 3314
  • [8] MR-guided interventions of the prostate gland
    Zangos, S.
    Eichler, K.
    Thalhammer, A.
    Schoepf, J. U.
    Costello, P.
    Herzog, C.
    Mack, M. G.
    Vogl, T. J.
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2007, 16 (04) : 222 - 229
  • [9] MR-guided intravascular interventions: Techniques and applications
    Bock, Michael
    Wacker, Frank K.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) : 326 - 338
  • [10] MR-Guided Interventions of the Prostate Gland a Literature Review
    Zangos, S.
    Eichler, K.
    Thalhammer, A.
    Mack, M. G.
    Marquardt, F.
    Vogl, T. J.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2010, 182 (11): : 947 - 953