A Wireless 8-Channel ECG Biopotential Acquisition System for Dry Electrodes

被引:0
|
作者
Pu, Xiaofei [1 ]
Wan, Lei [1 ]
Sheng, Yun [2 ]
Chiang, Patrick [1 ,3 ]
Qin, Yajie [1 ]
Hong, Zhiliang [1 ]
机构
[1] Fudan Univ, State Key Lab ASIC & Syst, Shanghai 201203, Peoples R China
[2] Nanosense Corp, Shanghai 201203, Peoples R China
[3] Oregon State Univ, Corvallis, OR 97331 USA
关键词
ECG dry metal electrode; wireless; chopper modulation; instrumentation amplifier; CIRCUIT;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A wireless 8-channel biopotential acquisition system for capturing electrocardiogram (ECG) using dry electrodes is presented. The ECG system consists of copper electrodes, a micropowered 8-channel custom ASIC, and an off-the-shelf microprocessor and bluetooth radio. Each analog channel of the custom ECG front-end is composed of a chopper-modulated instrumentation amplifier (CMIA) with chopping spike filter (CSF), a programmable gain amplifier (PGA), and a output buffer. Implemented in standard a 0.35 gm CMOS technology, the ECG front-end consumes 101 mu A from a 2.7 V supply, occupying 5 mm(2) of chip area. Measurement results show an input impedance of 1 G Omega, an input-referred noise of 0.97 mu Vrms (0.5 similar to 100 Hz), and a CM of 114 dB. Finally, a complete wireless 8-channel ECG monitoring system incorporating this analog front-end is demonstrated, showing successful recordings of a capture ECG waveform using a smart phone.
引用
收藏
页码:140 / 142
页数:3
相关论文
共 50 条
  • [21] An 8-channel skin impedance measurement system for acupuncture research
    Thong, Tran
    Colbert, Agatha P.
    Larsen, Adrian P.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 861 - +
  • [22] A WIRELESS FLEXIBLE WEARABLE BIOPOTENTIAL ACQUISITION SYSTEM UTILIZING PARYLENE BASED MICRONEEDLE ARRAY
    Huang, Dong
    Wang, Hancong
    Li, Junshi
    Chen, Yufeng
    Li, Zhihong
    2019 20TH INTERNATIONAL CONFERENCE ON SOLID-STATE SENSORS, ACTUATORS AND MICROSYSTEMS & EUROSENSORS XXXIII (TRANSDUCERS & EUROSENSORS XXXIII), 2019, : 298 - 301
  • [23] Performance Evaluation of Woven Conductive Dry Textile Electrodes for Continuous ECG Signals Acquisition
    Rajanna, Rama Reddy
    Sriraam, Natarajan
    Vittal, Prabhu Ravikala
    Arun, Uma
    IEEE SENSORS JOURNAL, 2020, 20 (03) : 1573 - 1581
  • [24] 8-CHANNEL TIME MEASUREMENT SYSTEM WITH 2-NS RESOLUTION
    DORING, FW
    HOLLENBERG, K
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1981, 14 (05): : 638 - 640
  • [25] A high-performance 8 nV/√Hz 8-channel wearable and wireless system for real-time monitoring of bioelectrical signals
    Konstantinos Petkos
    Simos Koutsoftidis
    Thomas Guiho
    Patrick Degenaar
    Andrew Jackson
    Stephen E. Greenwald
    Peter Brown
    Timothy Denison
    Emmanuel M. Drakakis
    Journal of NeuroEngineering and Rehabilitation, 16
  • [26] A Cost-Effective Multichannel Wireless ECG Acquisition System
    Ahamed, Md. Asif
    Ahmad, Mohiuddin
    2018 10TH INTERNATIONAL CONFERENCE ON ELECTRICAL AND COMPUTER ENGINEERING (ICECE), 2018, : 397 - 400
  • [27] Design of ECG Acquisition System Based on Bluetooth Wireless Communication
    Wang, Xiaoni
    2014 5TH IEEE INTERNATIONAL CONFERENCE ON SOFTWARE ENGINEERING AND SERVICE SCIENCE (ICSESS), 2014, : 1019 - 1022
  • [28] Live Demonstration: A Wearable Wireless ECG Acquisition and Specification System
    Liang, Ming-Chun
    Hsieh, Cheng-Han
    Hong, Jia-Hua
    Chien, Shih-Yu Chang
    Lee, Shuenn-Yuh
    2014 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2014, : 438 - 438
  • [29] Development of Low Cost Wireless ECG Data Acquisition System
    Ahamed, Md. Asif
    Asraf-Ul-Ahad, Md.
    Sohag, Md. Hanif Ali
    Ahmad, Mohiuddin
    2015 INTERNATIONAL CONFERENCE ON ADVANCES IN ELECTRICAL ENGINEERING (ICAEE), 2015, : 72 - 75
  • [30] A Wireless ECG Acquisition and Classification System for Body Sensor Networks
    Hong, Jia-Hua
    Lee, Shuenn-Yuh
    Liang, Ming-Chun
    Hsieh, Cheng-Han
    Chien, Shih-Yu Chang
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 5183 - 5186