The "Sound of Silence" in a Neonatal Intensive Care Unit-Listening to Speech and Music Inside an Incubator

被引:22
|
作者
Bertsch, Matthias [1 ]
Reuter, Christoph [2 ]
Czedik-Eysenberg, Isabella [2 ]
Berger, Angelika [3 ]
Olischar, Monika [3 ]
Bartha-Doering, Lisa [3 ]
Giordano, Vito [3 ]
机构
[1] Univ Mus & Performing Arts Vienna, Dept Mus Physiol, Vienna, Austria
[2] Univ Vienna, Musicol Dept, Vienna, Austria
[3] Med Univ Vienna, Comprehens Ctr Pediat, Dept Pediat & Adolescent Med, Div Neonatol Pediat Intens Care & Neuropediat, Vienna, Austria
来源
FRONTIERS IN PSYCHOLOGY | 2020年 / 11卷
关键词
incubator; NICU; noise; music; preterm; transfer spectrum; recordings; ENVIRONMENTAL NOISE; EAR; INFANTS; NICU;
D O I
10.3389/fpsyg.2020.01055
中图分类号
B84 [心理学];
学科分类号
04 ; 0402 ;
摘要
Background: The intrauterine hearing experience differs from the extrauterine hearing exposure within a neonatal intensive care unit (NICU) setting. Also, the listening experience of a neonate drastically differs from that of an adult. Several studies have documented that the sound level within a NICU exceeds the recommended threshold by far, possibly related to hearing loss thereafter. The aim of this study was, first, to precisely define the dynamics of sounds within an incubator and, second, to give clinicians and caregivers an idea about what can be heard "inside the box." Methods: Audio recordings within an incubator were conducted at the Pediatric Simulation Center of the Medical University Vienna. They contained recorded music, speech, and synthesized sounds. To understand the dynamics of sounds around and within the incubator, the following stimuli were used: broadband noise with decreasing sound level in 10 steps of 6 dB, sine waves (62.5, 125, 250, 500, 1000, 2000, 4000, 8000, and 16,000 Hz), logarithmic sweep (Chirp) over the frequency band 20 Hz to 21 kHz, singing male voice, singing, and whispering female voice. Results: Our results confirm a protective effect of the incubator from noises above 500 Hz in conditions of "no-flow" and show almost no protective effect of an incubator cover. We, furthermore, observed a strong boost of low frequencies below 125 Hz within the incubator, as well as a notable increase of higher frequency noises with open access doors, a significant resonant effect of the incubator, and a considerable masking effect of the respiratory support against any other source of noise or sound stimulation even for "low-flow" conditions. Conclusion: Our study reveals high noise levels of air supply at high flow rates and the boost of low frequencies within the incubator. Education of medical staff and family members as well as modifications of the physical environment should aim at reducing noise exposure of preterm infants in the incubator. Audiovisual material is provided as Supplementary Material.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Noise at the Neonatal Intensive Care Unit and inside the incubator
    Pinheiro, Eliana Moreira
    Guinsburg, Ruth
    de Araujo Nabuco, Marco Antonio
    Kakehashi, Tereza Yoshiko
    REVISTA LATINO-AMERICANA DE ENFERMAGEM, 2011, 19 (05): : 1214 - 1221
  • [2] Noise exposure of the newborn in the incubator at the neonatal intensive care unit
    Bieber, C.
    Thoma, R.
    KLINISCHE PADIATRIE, 2007, 219 (02): : 105 - 105
  • [3] Religion, Spirituality, and the Intensive Care Unit The Sound of Silence
    Balboni, Tracy A.
    Balboni, Michael J.
    Fitchett, George
    JAMA INTERNAL MEDICINE, 2015, 175 (10) : 1669 - 1670
  • [4] Music Therapy in Neonatal Intensive Care Unit
    Imseytoglu, Diler
    Yildiz, Suzan
    FLORENCE NIGHTINGALE JOURNAL OF NURSING-FLORENCE NIGHTINGALE HEMSIRELIK DERGISI, 2012, 20 (02): : 160 - 165
  • [5] There is a lack of contraindications for live music listening for preterm and term infants in the neonatal intensive care unit
    Andrea M. Cevasco-Trotter
    Ellyn H. Evans
    Anna C. Baird-Galloway
    Journal of Perinatology, 2022, 42 : 513 - 514
  • [6] There is a lack of contraindications for live music listening for preterm and term infants in the neonatal intensive care unit
    Cevasco-Trotter, Andrea M.
    Evans, Ellyn H.
    Baird-Galloway, Anna C.
    JOURNAL OF PERINATOLOGY, 2022, 42 (04) : 513 - 514
  • [7] Sound levels in a neonatal intensive care unit significantly exceeded recommendations, especially inside incubators
    Parra, Johanna
    de Suremain, Aurelie
    Audeoud, Frederique Berne
    Ego, Anne
    Debillon, Thierry
    ACTA PAEDIATRICA, 2017, 106 (12) : 1909 - 1914
  • [8] Sound transmission into incubators in the neonatal intensive care unit
    Robertson A.
    Cooper-Peel C.
    Vos P.
    Journal of Perinatology, 1999, 19 (7) : 494 - 497
  • [9] Sound pressure levels in the Neonatal Intensive Care Unit
    Peixoto, Priscila Vendramini
    Nabuco de Araujo, Marco Antonio
    Kakehashi, Tereza Yoshiko
    Pinheiro, Eliana Moreira
    REVISTA DA ESCOLA DE ENFERMAGEM DA USP, 2011, 45 (06) : 1308 - 1313
  • [10] The infant incubator in the neonatal intensive care unit: unresolved issues and future developments
    Antonucci, Roberto
    Porcella, Annalisa
    Fanos, Vassilios
    JOURNAL OF PERINATAL MEDICINE, 2009, 37 (06) : 587 - 598