COMPARISON OF TWO BIOELECTRICAL IMPEDANCE ANALYSIS DEVICES WITH DUAL ENERGY X-RAY ABSORPTIOMETRY AND MAGNETIC RESONANCE IMAGING IN THE ESTIMATION OF BODY COMPOSITION

被引:90
|
作者
Wang, Ji-Guang [1 ]
Zhang, Yi [1 ]
Chen, Han-E [1 ]
Li, Yan [1 ]
Cheng, Xiao-Guang [2 ]
Xu, Li [2 ]
Guo, Zhe [2 ]
Zhao, Xing-Shan [3 ]
Sato, Tetsuya [4 ]
Cao, Qi-Yun [5 ]
Chen, Ke-Min [6 ]
Li, Biao [7 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Med, Ctr Epidemiol Studies & Clin Trials, Shanghai Inst Hypertens,Ruijin Hosp, Shanghai 200030, Peoples R China
[2] Beijing Univ, Jishuitan Hosp, Dept Radiol, Beijing 100871, Peoples R China
[3] Beijing Univ, Jishuitan Hosp, Dept Cardiol, Beijing 100871, Peoples R China
[4] Omron Healthcare, Kyoto, Japan
[5] Omron Healthcare, Shanghai, Peoples R China
[6] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Dept Radiol, Shanghai 200030, Peoples R China
[7] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Dept Nucl Med, Shanghai 200030, Peoples R China
关键词
fat; skeletal muscle; visceral fat; measurement; accuracy; SKELETAL-MUSCLE MASS; VALIDATION; APPLICABILITY; PREVALENCE; PREDICTION; OVERWEIGHT; EQUATION;
D O I
10.1519/JSC.0b013e31824f2040
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Wang, J-G, Zhang, Y, Chen, H-E, Li, Y, Cheng, X-G, Xu, L, Guo, Z, Zhao, X-S, Sato, T, Cao, Q-Y, Chen, K-M, and Li, B. Comparison of two bioelectrical impedance analysis devices with dual energy X-ray absorptiometry and magnetic resonance imaging in the estimation of body composition. J Strength Cond Res 27(1): 236-243, 2013-We compared a 4-limb bioelectrical impedance analysis (BIA) system, HBF 359 (Omron), and a 2-limb foot-to-foot device, BC 532 (Tanita), with the standard dual energy X-ray absorptiometry (DXA) and magnetic resonance imaging (MRI) methods for the measurement of body fat percentage (BF), skeletal muscle mass percentage (SMM, or fat-free mass [FFM] for BC 532), and visceral fat level (VF). Body composition was measured in 200 healthy volunteers (100 men and 100 women, mean age 48 years) by HBF 359 and BC 532 and by DXA and MRI. The agreement was assessed by correlation analysis and paired t-test. The correlation coefficients between BIA and DXA or MRI ranged from 0.71 to 0.89 for BF, SMM, and VF by HBF 359 and from 0.77 to 0.90 for BF, FFM, and VF by BC 532 in all subjects and in men and women separately (p < 0.001 for all). Compared with DXA, HBF 359 significantly (p < 0.001) underestimated BF by -5.8% in men and -9.6% in women. Compared with MRI, the corresponding underestimatons (negative) or overestimations (positive) by HBF 359 in men and women were, respectively, +1.9% (p = 0.02) and +1.7% (p = 0.10) for SMM, and +13.3% (p < 0.001) and -8.5% (p = 0.006), for VF. The corresponding values by BC 532 in men and women were -10.7 and -6.2% for BF, -1.4 and -2.5% for FFM, and +20.4 and -18.0% for VF. The BIA devices are accurate in the estimation of body composition, especially skeletal muscle mass or FFM.
引用
收藏
页码:236 / 243
页数:8
相关论文
共 50 条
  • [1] Accuracy of Bioelectrical Impedance Consumer Devices for Measurement of Body Composition in Comparison to Whole Body Magnetic Resonance Imaging and Dual X-Ray Absorptiometry
    Bosy-Westphal, Anja
    Later, Wiebke
    Hitze, Britta
    Sato, Tetsuya
    Kossel, Elke
    Glueer, Claus-Christian
    Heller, Martin
    Mueller, Manfred James
    OBESITY FACTS, 2008, 1 (06) : 319 - 324
  • [2] Comparison of bioelectrical impedance analysis and dual energy X-ray absorptiometry for assessment of body composition in children
    Okasora, K
    Takaya, R
    Tokuda, M
    Fukunaga, Y
    Oguni, T
    Tanaka, H
    Konishi, K
    Tamai, H
    PEDIATRICS INTERNATIONAL, 1999, 41 (02) : 121 - 125
  • [3] Comparison of two bioelectrical impedance devices and dual-energy X-ray absorptiometry to evaluate body composition in heart failure
    Alves, F. D.
    Souza, G. C.
    Biolo, A.
    Clausell, N.
    JOURNAL OF HUMAN NUTRITION AND DIETETICS, 2014, 27 (06) : 632 - 638
  • [4] Comparison of Body Composition by Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry in Hispanic Diabetics
    Batech, Michael
    Beeson, W. Lawrence
    Schultz, Eloy
    Salto, Lorena
    Firek, Anthony
    DeLeon, Marino
    Balcazar, Hector
    Cordero-MacIntyre, Zaida
    FASEB JOURNAL, 2010, 24
  • [5] Comparison of segmental body composition estimated by bioelectrical impedance analysis and dual-energy X-ray absorptiometry
    Wingo, Brooks C.
    Barry, Valene Garr
    Ellis, Amy C.
    Gower, Barbara A.
    CLINICAL NUTRITION ESPEN, 2018, 28 : 141 - 147
  • [6] Comparison of Anthropometry, Bioelectrical Impedance, and Dual-energy X-ray Absorptiometry for Body Composition in Cirrhosis
    Grover, Indu
    Singh, Namrata
    Gunjan, Deepak
    Pandey, Ravindra M.
    Sati, Hem Chandra
    Saraya, Anoop
    JOURNAL OF CLINICAL AND EXPERIMENTAL HEPATOLOGY, 2022, 12 (02) : 467 - 474
  • [7] ESTIMATION OF BODY-COMPOSITION FROM BIOELECTRICAL-IMPEDANCE OF BODY SEGMENTS - COMPARISON WITH DUAL-ENERGY X-RAY ABSORPTIOMETRY
    STEWART, SP
    BRAMLEY, PN
    HEIGHTON, R
    GREEN, JH
    HORSMAN, A
    LOSOWSKY, MS
    SMITH, MA
    BRITISH JOURNAL OF NUTRITION, 1993, 69 (03) : 645 - 655
  • [8] Body Composition in Heavy Smokers: Comparison of Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry
    Rom, O.
    Reznick, A. Z.
    Keidar, Z.
    Karkabi, K.
    Aizenbud, D.
    BODY METABOLISM AND EXERCISE, 2015, 840 : 1 - 11
  • [9] Body composition in male elite athletes, comparison of bioelectrical impedance spectroscopy with dual energy X-ray absorptiometry
    Svantesson, Ulla
    Zander, Martina
    Klingberg, Sofia
    Slinde, Frode
    JOURNAL OF NEGATIVE RESULTS IN BIOMEDICINE, 2008, 7
  • [10] Comparison of bioelectrical impedance analysis and dual-energy X-ray absorptiometry for the assessment of appendicular body composition in anorexic women
    Bedogni, G
    Marra, M
    Bianchi, L
    Malavolti, M
    Nicolai, E
    De Filippo, E
    Scalfi, L
    EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2003, 57 (09) : 1068 - 1072