THE UTILITY OF THE C-MAC AS A DIRECT LARYNGOSCOPE FOR INTUBATION IN THE EMERGENCY DEPARTMENT

被引:31
|
作者
Sakles, John C. [1 ]
Mosier, Jarrod M. [1 ]
Patanwala, Asad E. [2 ]
Arcaris, Brittany [3 ]
Dicken, John M. [3 ]
机构
[1] Univ Arizona, Coll Med, Dept Emergency Med, 1501 N Campbell Ave,POB 245057, Tucson, AZ 85724 USA
[2] Univ Arizona, Coll Pharm, Dept Pharm Practice & Sci, Tucson, AZ 85721 USA
[3] Univ Arizona, Coll Med, Tucson, AZ USA
来源
JOURNAL OF EMERGENCY MEDICINE | 2016年 / 51卷 / 04期
关键词
C-MAC; direct laryngoscopy; EM resident intubations; URGENT ENDOTRACHEAL INTUBATION; RANDOMIZED CONTROLLED-TRIAL; VIDEO-LARYNGOSCOPY; TRACHEAL INTUBATION; AIRWAY MANAGEMENT; OROTRACHEAL INTUBATION; ESOPHAGEAL INTUBATIONS; ADVERSE EVENTS; VIDEOLARYNGOSCOPY; DEVICE;
D O I
10.1016/j.jemermed.2016.05.039
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Although the C-MAC (Karl Storz, Tuttlingen, Germany) is a video laryngoscope (VL), it can also be used as a direct laryngoscope (DL). Objective: The goal of this study was to evaluate the utility of the C-MAC as a DL for intubations in the emergency department (ED). Methods: This was an analysis of prospectively collected continuous quality-improvement data during the 6-year period from February 1, 2009 to January 31, 2015, when both the C-MAC and Macintosh DL (Mac DL) were clinically available in our ED. This analysis included adult patients who underwent rapid sequence intubation by an emergency medicine resident in the ED with a C-MAC initially used as a DL or a Mac DL. The primary outcome measure was the first pass success (FPS). Results: When the C-MAC was used as a DL, the initial DL attempt was successful in 199 of 346 (57.6%) cases. When the attempt could not be completed using the C-MAC as a DL, the operator utilized the video monitor and successfully completed the intubation using VL in 104 of 134 (77.6%) cases, thus achieving an overall FPS of 303 of 346 (87.6%). When the Mac DL was used, the FPS was 505 of 671 (75.3%). Conclusions: The C-MAC is a useful device for DL because in the event of a failed DL attempt, operators have the option of switching to the video monitor to successfully complete the intubation using VL without having to make a second attempt. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:349 / 357
页数:9
相关论文
共 50 条
  • [21] Endotracheal intubation using the C-MAC® video laryngoscope or the Macintosh laryngoscope: A prospective, comparative study in the ICU
    Ruediger R Noppens
    Stephanie Geimer
    Nicole Eisel
    Matthias David
    Tim Piepho
    Critical Care, 16
  • [22] A randomized comparative evaluation of C-MAC video-laryngoscope with Miller laryngoscope for neonatal endotracheal intubation
    Goel, Sachin
    Choudhary, Ripon
    Magoon, Rohan
    Sharma, Ridhima
    Usha, G.
    Kapoor, Poonam M.
    Bagga, Deepak
    JOURNAL OF ANAESTHESIOLOGY CLINICAL PHARMACOLOGY, 2022, 38 (03) : 464 - 468
  • [23] Endotracheal intubation using the C-MAC® video laryngoscope or the Macintosh laryngoscope: A prospective, comparative study in the ICU
    Noppens, Ruediger R.
    Geimer, Stephanie
    Eisel, Nicole
    David, Matthias
    Piepho, Tim
    CRITICAL CARE, 2012, 16 (03):
  • [24] A comparison of the video laryngoscope to the direct laryngoscope for endotracheal intubation in emergency department patients
    Sun, Yongbo
    Zhang, Shiming
    Lin, Bin
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2020, 13 (11): : 9151 - 9158
  • [25] Comparison of aerosol box intubation with C-MAC video laryngoscope and direct laryngoscopy-A randomised controlled trial
    Puthenveettil, Nitu
    Rahman, Sajan
    Vijayaraghavan, Sundeep
    Suresh, Sneha
    Kadapamannil, Dilesh
    Paul, Jerry
    INDIAN JOURNAL OF ANAESTHESIA, 2021, 65 (02) : 133 - 138
  • [26] Tracheal intubation using the mobile C-MAC video laryngoscope or direct laryngoscopy for patients with a simulated difficult airway
    Byhahn, C.
    Iber, T.
    Zacharowski, K.
    Weber, C. F.
    Ruesseler, M.
    Schalk, R.
    Meininger, D.
    MINERVA ANESTESIOLOGICA, 2010, 76 (08) : 577 - 583
  • [27] Comparative evaluation of C-MAC and McGrath MAC videolaryngoscopes with Macintosh direct laryngoscope for endotracheal intubation in adult patients undergoing elective surgeries
    Abhyankar, Prachi
    Sabharwal, Nikki
    Gupta, Anju
    Das, Akshaya K.
    JOURNAL OF ANAESTHESIOLOGY CLINICAL PHARMACOLOGY, 2023, 39 (03) : 422 - 428
  • [28] Clinical evaluation of C-Mac D-Blade video laryngoscope: randomized comparison with C-Mac Blade and direct laryngoscopy
    Hostic, V.
    Maldini, B.
    Novotny, Z.
    Hodzovic, I.
    BRITISH JOURNAL OF ANAESTHESIA, 2016, 116 (06) : E905 - E906
  • [29] The C-MAC videolaryngoscope compared with conventional laryngoscopy for rapid sequence intubation at the emergency department: study protocol
    Simon Sulser
    Dirk Ubmann
    Martin Brueesch
    Georg Goliasch
    Burkhardt Seifert
    Donat R Spahn
    Kurt Ruetzler
    Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 23
  • [30] The C-MAC videolaryngoscope compared with conventional laryngoscopy for rapid sequence intubation at the emergency department: study protocol
    Sulser, Simon
    Ubmann, Dirk
    Brueesch, Martin
    Goliasch, Georg
    Seifert, Burkhardt
    Spahn, Donat R.
    Ruetzler, Kurt
    SCANDINAVIAN JOURNAL OF TRAUMA RESUSCITATION & EMERGENCY MEDICINE, 2015, 23