Proton radiography and proton computed tomography based on time-resolved dose measurements

被引:59
|
作者
Testa, Mauro [1 ]
Verburg, Joost M. [1 ,2 ]
Rose, Mark [3 ]
Min, Chul Hee [1 ]
Tang, Shikui [1 ]
Bentefour, El Hassane [1 ]
Paganetti, Harald [1 ]
Lu, Hsiao-Ming [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiat Oncol, Boston, MA 02114 USA
[2] Eindhoven Univ Technol, Sch Med Phys & Engn, NL-5600 MB Eindhoven, Netherlands
[3] Sun Nucl Corp, Melbourne, FL 32940 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 22期
关键词
VIVO RANGE VERIFICATION; SPATIAL-RESOLUTION; BEAM RADIOGRAPHY; THERAPY; SYSTEM; RADIOTHERAPY; DENSITY; GEANT4;
D O I
10.1088/0031-9155/58/22/8215
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present a proof of principle study of proton radiography and proton computed tomography (pCT) based on time-resolved dose measurements. We used a prototype, two-dimensional, diode-array detector capable of fast dose rate measurements, to acquire proton radiographic images expressed directly in water equivalent path length (WEPL). The technique is based on the time dependence of the dose distribution delivered by a proton beam traversing a range modulator wheel in passive scattering proton therapy systems. The dose rate produced in the medium by such a system is periodic and has a unique pattern in time at each point along the beam path and thus encodes the WEPL. By measuring the time dose pattern at the point of interest, the WEPL to this point can be decoded. If one measures the time-dose patterns at points on a plane behind the patient for a beam with sufficient energy to penetrate the patient, the obtained 2D distribution of the WEPL forms an image. The technique requires only a 2D dosimeter array and it uses only the clinical beam for a fraction of second with negligible dose to patient. We first evaluated the accuracy of the technique in determining the WEPL for static phantoms aiming at beam range verification of the brain fields of medulloblastoma patients. Accurate beam ranges for these fields can significantly reduce the dose to the cranial skin of the patient and thus the risk of permanent alopecia. Second, we investigated the potential features of the technique for real-time imaging of a moving phantom. Real-time tumor tracking by proton radiography could provide more accurate validations of tumor motion models due to the more sensitive dependence of proton beam on tissue density compared to x-rays. Our radiographic technique is rapid (similar to 100 ms) and simultaneous over the whole field, it can image mobile tumors without the problem of interplay effect inherently challenging for methods based on pencil beams. Third, we present the reconstructed pCT images of a cylindrical phantom containing inserts of different materials. As for all conventional pCT systems, the method illustrated in this work produces tomographic images that are potentially more accurate than x-ray CT in providing maps of proton relative stopping power (RSP) in the patient without the need for converting x-ray Hounsfield units to proton RSP. All phantom tests produced reasonable results, given the currently limited spatial and time resolution of the prototype detector. The dose required to produce one radiographic image, with the current settings, is similar to 0.7 cGy. Finally, we discuss a series of techniques to improve the resolution and accuracy of radiographic and tomographic images for the future development of a full-scale detector.
引用
收藏
页码:8215 / 8233
页数:19
相关论文
共 50 条
  • [41] Issues in proton computed tomography
    Sadrozinski, HFW
    Bashkirov, V
    Bruzzi, M
    Johnson, LR
    Keeney, B
    Ross, G
    Schulte, RW
    Seiden, A
    Shahnazi, K
    Williams, DC
    Zhang, L
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 511 (1-2): : 275 - 281
  • [42] PROTON COMPUTED-TOMOGRAPHY
    HANSON, KM
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (01) : 1635 - 1640
  • [43] Proton Radiography with FLASH Dose Rates
    Stanforth, A.
    Charyyev, S.
    Arrue, J.
    Duce, M.
    Erickson, A.
    Dhabaan, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E704 - E704
  • [44] Toward proton computed tomography
    Sadrozinski, HFW
    Bashkirov, V
    Keeney, B
    Johnson, LR
    Peggs, SG
    Ross, G
    Satogata, T
    Schulte, RWM
    Seiden, A
    Shanazi, K
    Williams, DC
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) : 3 - 9
  • [45] Time-resolved dosimetry for validation of 4D dose calculation in PBS proton beam radiotherapy
    Kostiukhina, N.
    Palmans, H.
    Stock, M.
    Georg, D.
    Knaeusl, B.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S122 - S123
  • [46] Audit of radiation dose delivered in time-resolved four-dimensional computed tomography in a radiotherapy department
    Hubbard, Patricia
    Callahan, Jason
    Cramb, Jim
    Budd, Ray
    Kron, Tomas
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2015, 59 (03) : 346 - 352
  • [47] Volume and enthalpy changes of proton transfers in the bacteriorhodopsin photocycle studied by millisecond time-resolved photopressure measurements
    Liu, Yan
    Edens, Gregory J.
    Grzymski, Joseph
    Mauzerall, David
    BIOCHEMISTRY, 2008, 47 (29) : 7752 - 7761
  • [48] Assembling and Test of a Proton Computed Radiography Apparatus
    Menichelli, David
    Bruzzi, Mara
    Bucciolini, Marta
    Candiano, Giuliana
    Cirrone, Giulio A. P.
    Civinini, Carlo
    Cuttone, Giacomo
    Lo Presti, Domenico
    Marrazzo, Livia
    Pallotta, Stefania
    Randazzo, Nunzio
    Scaringella, Monica
    Sipala, Valeria
    Talamonti, Cinzia
    Brianzi, Mirko
    Tesi, Mauro
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 4181 - +
  • [49] A proton Computed Tomography based medical imaging system
    Scaringella, M.
    Bruzzi, M.
    Bucciolini, M.
    Carpinelli, M.
    Cirrone, G. A. P.
    Civinini, C.
    Cuttone, G.
    Lo Presti, D.
    Pallotta, S.
    Pugliatti, C.
    Randazzo, N.
    Romano, F.
    Sipala, V.
    Stancampiano, C.
    Talamonti, C.
    Vanzi, E.
    Zani, M.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [50] Time-resolved Fourier transform infrared spectroscopy of the polarizable proton continua and the proton pump mechanism of bacteriorhodopsin
    Wang, JP
    El-Sayed, MA
    BIOPHYSICAL JOURNAL, 2001, 80 (02) : 961 - 971