Integration of Microdialysis Sampling and Microchip Electrophoresis with Electrochemical Detection

被引:60
|
作者
Mecker, Laura C. [1 ]
Martin, R. Scott [1 ]
机构
[1] St Louis Univ, Dept Chem, St Louis, MO 63103 USA
基金
美国国家卫生研究院;
关键词
D O I
10.1021/ac801614r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Here we describe the fabrication, optimization, and application of a microfluidic device that integrates microdialysis (MD) sampling, microchip electrophoresis (ME), and electrochemical detection (EC). The manner in which the chip is produced is reproducible and enables the fixed alignment of the MD/ME and ME/EC interfaces. Poly(dimethylsiloxane) (PDMS)-based valves were used for the discrete injection of sample from the hydrodynamic MD dialysate stream into a separation channel for analysis with ME. To enable the integration of ME with EC detection, a palladium decoupler was used to isolate the high voltages associated with electrophoresis from micrometer-sized carbon ink detection electrodes. Optimization of the ME/EC interface was needed to allow the use of biologically appropriate perfusate buffers containing high salt content. This optimization included changes in the fabrication procedure, increases in the decoupler surface area, and a programmed voltage shutoff. The ability of the MD/ME/EC system to sample a biological system was demonstrated by using a linear probe to monitor the stimulated release of dopamine from a confluent layer of PC 12 cells. To our knowledge, this is the first report of a microchip-based system that couples microdialysis sampling with microchip electrophoresis and electrochemical detection.
引用
收藏
页码:9257 / 9264
页数:8
相关论文
共 50 条
  • [21] Bioanalytical applications of microchip capillary electrophoresis with electrochemical detection.
    Lunte, SM
    Martin, RS
    Gawron, A
    Lacher, N
    Fogarty, B
    Reagan, F
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2001, 221 : U102 - U102
  • [22] New analytical portable instrument for microchip electrophoresis with electrochemical detection
    Fernandez-la-Villa, Ana
    Pozo-Ayuso, Diego F.
    Castano-Alvarez, Mario
    ELECTROPHORESIS, 2010, 31 (15) : 2641 - 2649
  • [23] Microchip capillary electrophoresis with electrochemical detection for monitoring environmental pollutants
    Chen, G
    Lin, YH
    Wang, J
    CURRENT ANALYTICAL CHEMISTRY, 2006, 2 (01) : 43 - 50
  • [24] Application of microchip electrophoresis with electrochemical detection to environmental aldehyde monitoring
    Dossi, Nicolo
    Susmel, Sabina
    Toniolo, Rosanna
    Pizzariello, Andrea
    Bontempelli, Gino
    ELECTROPHORESIS, 2009, 30 (19) : 3465 - 3471
  • [25] Electrochemical Detection Using an Engraved Microchip Capillary Electrophoresis Platform
    de Abreu, Fabiane Caxico
    Costa, Elton Elias M.
    ELECTROANALYSIS, 2016, 28 (09) : 2104 - 2108
  • [26] Polymer based microchip for combined capillary electrophoresis and electrochemical detection
    Rashid, M.
    Auger, V.
    Williams, L.
    Dou, Y.
    Pruneanu, S.
    Ali, Z.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 111 - +
  • [27] Monitoring environmental pollutants by microchip capillary electrophoresis with electrochemical detection
    Chen, G
    Lin, YH
    Wang, J
    TALANTA, 2006, 68 (03) : 497 - 503
  • [28] Simultaneous electrochemical and electrochemiluminescence detection for microchip and conventional capillary electrophoresis
    Qiu, HB
    Yin, XB
    Yan, JL
    Zhao, XC
    Yang, XR
    Wang, EK
    ELECTROPHORESIS, 2005, 26 (03) : 687 - 693
  • [29] Microchip electrophoresis and electrochemical detection: A review on a growing synergistic implementation
    Costa, Brenda Maria De Castro
    Griveau, Sophie
    d'Orlye, Fanny
    Bedioui, Fethi
    da Silva, Jose Alberto Fracassi
    Varenne, Anne
    ELECTROCHIMICA ACTA, 2021, 391
  • [30] Microchip capillary electrophoresis device with on-line microdialysis sampling for near real-time monitoring
    Huynh, B
    Fogarty, B
    Martin, S
    Lunte, S
    Micro Total Analysis Systems 2004, Vol 1, 2005, (296): : 548 - 550