This paper describes a general-purpose photoelastic fiber optic accelerometer having a sensitivity of 2 V/g and a resolution of 2.5 mg over a frequency range extending from 0 to 600 Hz. The unit has a linear amplitude range of 0 to 5g and is sealed in an aluminum case having outside dimensions of 41 mm x 35 mm x 20 mm. It relies on a force transducer made from an optically birefringent material and measurements are made based on the principles of photoelasticity. A low-cost LED is employed as an incoherent light source; multimode optical fibers with a hard plastic cladding are used to transmit signals between the transducer and the conditioning electronics. (C) 1997 Society of Photo-Optical Instrumentation Engineers.
机构:
Institute of Molecular Medicine and Shanghai Key Laboratory for Nucleic Acid Chemistry and Nanomedicine,Renji Hospital,School of Medicine,Shanghai Jiao Tong UniversityInstitute of Molecular Medicine and Shanghai Key Laboratory for Nucleic Acid Chemistry and Nanomedicine,Renji Hospital,School of Medicine,Shanghai Jiao Tong University