Validity of segmental multiple-frequency bioelectrical impedance analysis to estimate body composition of adults across a range of body mass indexes

被引:181
|
作者
Shafer, Kimberly J. [1 ]
Siders, William A. [1 ]
Johnson, LuAnn K. [1 ]
Lukaski, Henry C. [1 ]
机构
[1] USDA, Grand Forks Human Nutr Res Ctr, Grand Forks, ND USA
关键词
Bioelectrical impedance analysis; Body composition; Body mass index; Dual x-ray absorptiometry; X-RAY ABSORPTIOMETRY; CROSS-CALIBRATION; OBESE; FAT; ACCURACY; OVERWEIGHT; PREVENTION; RISK; DXA;
D O I
10.1016/j.nut.2008.07.004
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: We compared body composition estimates using an eight-electrode, segmental, multiple-frequency bioelectrical impedance analysis (segmental MF-BIA) and dual x-ray absorptiometry (DXA) in a group of healthy adults with a range of body mass indexes (BMIs). Methods: Percentage of body fat (%BF), fat-free mass, and fat mass assessed by DXA and segmental MF-BIA in 132 healthy adults were classified by normal (N; 18.5-24.9 kg/m(2)), overweight (OW; 25-29.9 kg/m(2)), and obese (OB; 30-39.9 kg/m(2)) BMI. Results: Compared with DXA, segmental MF-BIA overestimated %BF in the OB BMI group (3.4%; P < 0.0001). MF-BIA overestimated %BF among men (0.75%; P < 0.006) and women (0.87%; P < 0.006) and underestimated it in the N BMI group (-1.56%; P < 0.0001); %BF was not different between methods in the OW BMI group. Error in %BF determined by segmental MF-BIA and DXA increased as %BF increased (r = 0.42, P < 0.0001). Waist circumference was the only significant predictor of systematic error in %BF between MF-BIA and DXA (r = 0.60, P < 0.0001). Conclusion: Eight-electrode, segmental MF-BIA is a valid method to estimate %BF in adults with BMI classified as N and OW, but not as OB. Estimation of trunk resistance with current segmental MF-BIA devices may explain the underestimation of %BF in the adults with OB BMI. Further examination of the effect of waist circumference and body fat distribution on the accuracy of BIA measurements is warranted. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:25 / 32
页数:8
相关论文
共 50 条
  • [1] Body composition: validity of segmental bioelectrical impedance analysis
    LaForgia, Joe
    Gunn, Simon M.
    Withers, Robert T.
    ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION, 2008, 17 (04) : 586 - 591
  • [2] ASSESSMENT OF BODY COMPOSITION IN PATIENTS WITH CHRONIC KIDNEY DISEASE USING MULTIPLE-FREQUENCY BIOELECTRICAL IMPEDANCE ANALYSIS
    Andronesi, Andreea
    Robe, Cristina
    Berbecar, Vlad
    Andronesi, Danut
    Gamala, Andreea
    Obrisca, Bogdan
    Ismail, Gener
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2017, 32
  • [3] Estimation of body water compartments in cirrhosis by multiple-frequency bioelectrical-impedance analysis
    Lehnert, ME
    Clarke, DD
    Gibbons, JG
    Ward, LC
    Golding, SM
    Shepherd, RW
    Cornish, BH
    Crawford, DHG
    NUTRITION, 2001, 17 (01) : 31 - 34
  • [4] Equations to estimate body composition using bioelectrical impedance in Chilean adults
    Schifferli, Ingrid
    Jose Orellana-Caceres, Juan
    Morales, Gladys
    Inostroza, Jorge
    Carrasco, Fernando
    REVISTA MEDICA DE CHILE, 2020, 148 (10) : 1435 - 1443
  • [5] VALIDITY OF BIOELECTRICAL IMPEDANCE TO ESTIMATE BODY-COMPOSITION IN CARDIAC AND PULMONARY PATIENTS
    KATCH, FI
    SOLOMON, RT
    SHAYEVITZ, M
    SHAYEVITZ, B
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1986, 43 (06): : 972 - 973
  • [6] Body mass index, bioelectrical impedance, and body composition
    Lukaski, HC
    NUTRITION, 2001, 17 (01) : 55 - 56
  • [7] Data analysis in multiple-frequency bioelectrical impedance analysis
    Cornish, BH
    Ward, LC
    PHYSIOLOGICAL MEASUREMENT, 1998, 19 (02) : 275 - 283
  • [8] Use of different segmental multi-frequency bioelectrical impedance devices for analysis of body composition in young adults: comparison with bioelectrical spectroscopy
    Guedes, Dartagnan Pinto
    Calabrese, Jean Carlos
    Pirolli, Paulo Marcelo
    NUTRICION HOSPITALARIA, 2019, 36 (03) : 618 - 625
  • [9] Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans
    Bracco, D
    Thiebaud, D
    Chiolero, RL
    Landry, M
    Burckhardt, P
    Schutz, Y
    JOURNAL OF APPLIED PHYSIOLOGY, 1996, 81 (06) : 2580 - 2587
  • [10] Using Multiple Frequency Bioelectrical Impedance to Determine Regional Body Composition
    Davidson, Robert
    Jones, Brian
    Danhoff, Guy
    FASEB JOURNAL, 2015, 29