The electrically evoked auditory change complex: Preliminary results from Nucleus cochlear implant users

被引:43
|
作者
Brown, Carolyn J. [1 ,2 ]
Etler, Christine [2 ]
He, Shuman [1 ,2 ]
O'Brien, Sara [2 ]
Erenberg, Sheryl [2 ]
Kim, Jae-Ryong [2 ,3 ]
Dhuldhoya, Aayesha N. [1 ]
Abbas, Paul J. [1 ,2 ]
机构
[1] Univ Iowa, Dept Speech Pathol & Audiol, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Otolaryngol Head & Neck Surg, Iowa City, IA 52242 USA
[3] Dong A Univ, Coll Med, Dept Otolaryngol Head & Neck Surg, Pusan, South Korea
来源
EAR AND HEARING | 2008年 / 29卷 / 05期
关键词
D O I
10.1097/AUD.0b013e31817a98af
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objectives: The purpose of this study was to determine if changes in the position of the stimulating electrode in the cochlea could be used to elicit the electrically evoked auditory change complex (EACC) from Nucleus cochlear implant users. Design: Nine postlingually deafened adults participated in this study. Each study participant had been using his or her Nucleus CI24 cochlear implant for at least 3 mos before testing. The speech processor was bypassed and the output of the implanted receiver/stimulator was controlled directly. The stimulus was a 600 msec burst of a biphasic pulse train (1000 pps). In control conditions, the stimulating electrode was held constant and stimulation continued throughout the 600 msec recording interval. In experimental conditions, the EACC was elicited by introducing a change in the stimulating electrode 300 msec after the onset of the pulse train. The EACC was recorded using surface electrodes. Three recordings of 100 sweeps each were obtained for each stimulus condition. Bandpass filtering (1-100 Hz) was used to minimize contamination of the recordings by stimulus artifact. Averaged responses were then smoothed using a 40-msec wide boxcar filter and standard peak picking procedures were used to analyze these responses in the time domain. Results: In each case, a clear onset response (P1-N1-P2) was recorded. In the experimental conditions, a second evoked potential, the EACC, was also recorded after the change in stimulating electrode. This second response had general morphological characteristics that were very similar to those of the onset response. Increasing the separation between the two stimulating electrodes in the experimental conditions resulted in a general trend toward increased EACC amplitudes. Conclusions: This report describes results of a set of experiments in which the speech processor of the cochlear implant was bypassed and the EACC was recorded in response to a change in stimulating electrode position. EACC amplitude was shown to increase as the separation between the two stimulating electrodes increased. Although preliminary in nature, these results demonstrate the feasibility of recording the EACC in response to changes in stimulating electrode position from individual cochlear implant users.
引用
收藏
页码:704 / 717
页数:14
相关论文
共 50 条
  • [31] Functional evaluation of auditory system in patients with cochlear implant using electrically evoked auditory brainstem responses
    Wang, L.
    Zhang, Q.
    Wang, Q.
    Dong, M.
    Zeng, Y.
    ACOUSTICAL PHYSICS, 2009, 55 (06) : 857 - 865
  • [32] Functional evaluation of auditory system in patients with cochlear implant using electrically evoked auditory brainstem responses
    L. Wang
    Q. Zhang
    Q. Wang
    M. Dong
    Y. Zeng
    Acoustical Physics, 2009, 55 : 857 - 865
  • [33] Characterization of cochlear implant artifacts in electrically evoked auditory steady-state responses
    Deprez, Hanne
    Gransier, Robin
    Hofmann, Michael
    van Wieringen, Astrid
    Wouters, Jan
    Moonen, Marc
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2017, 31 : 127 - 138
  • [34] Effect of anesthetic agents on electrically evoked auditory responses in pediatric cochlear implant surgery
    Michelle White
    Crawford Mark
    Zaarour Christian
    E. Propst
    J. O’Rourke
    K. Gordon
    A. James
    B. Papsin
    Canadian Journal of Anaesthesia, 2007, 54 (Suppl 1) : 44283 - 44283
  • [35] Electrically evoked mismatch negativity responses to loudness and pitch cues in cochlear implant users
    Wagner, Luise
    Ladek, Anna S.
    Plontke, Stefan K.
    Rahne, Torsten
    SCIENTIFIC REPORTS, 2023, 13 (01):
  • [36] Electrically evoked compound action potentials in cochlear implant users with preoperative residual hearing
    Liebscher, Tim
    Hornung, Joachim
    Hoppe, Ulrich
    FRONTIERS IN HUMAN NEUROSCIENCE, 2023, 17
  • [37] Relationships Between Electrically Evoked Potentials and Loudness Growth in Bilateral Cochlear Implant Users
    Kirby, Benjamin
    Brown, Carolyn
    Abbas, Paul
    Etler, Christine
    O'Brien, Sara
    EAR AND HEARING, 2012, 33 (03): : 389 - 398
  • [38] Electrically evoked mismatch negativity responses to loudness and pitch cues in cochlear implant users
    Luise Wagner
    Anna S. Ladek
    Stefan K. Plontke
    Torsten Rahne
    Scientific Reports, 13 (1)
  • [39] Preliminary Results of the Relationship Between the Binaural Interaction Component of the Electrically Evoked Auditory Brainstem Response and Interaural Pitch Comparisons in Bilateral Cochlear Implant Recipients
    He, Shuman
    Brown, Carolyn J.
    Abbas, Paul J.
    EAR AND HEARING, 2012, 33 (01): : 57 - 68
  • [40] A longitudinal study of electrode impedance, the electrically evoked compound action potential, and behavioral measures in Nucleus 24 cochlear implant users
    Hughes, ML
    Werff, KRV
    Brown, CJ
    Abbas, PJ
    Kelsay, DMR
    Teagle, HFB
    Lowder, MW
    EAR AND HEARING, 2001, 22 (06): : 471 - 486