Finite element analysis and experimental validation of polycentric prosthetic knee

被引:7
|
作者
Mohanty, Rajesh Kumar [1 ,2 ]
Mohanty, Ramesh Chandra [3 ]
Sabut, Sukanta Kumar [4 ]
机构
[1] Centurion Univ Technol & Management, Bhubaneswar 752050, Odisha, India
[2] Swami Vivekanand Natl Inst Rehabil Training & Res, Dept Prosthet & Orthot, Cuttack 754010, Odisha, India
[3] Centurion Univ Technol & Management, Dept Mech Engn, Bhubaneswar 752050, Odisha, India
[4] Kalinga Inst Ind Technol, Sch Elect Engn, Bhubaneswar 751024, Odisha, India
关键词
Experimental; Finite element analysis; Prosthetic knee; Static strength; Simulation; Validation; DESIGN;
D O I
10.1016/j.matpr.2022.02.509
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The objective of this work is to perform simulation and experimental validation of a polycentric prosthetic knee through finite element analysis. The prosthetic knee may be subjected to a load of 4.7 to 5.6 times body weight during various activities which are taken into consideration during its design with a higher factor of safety. Static strength test has been performed to ensure its structural integrity as per the International Organization for Standardization ISO 10328:2006 standard. ANSYS 20.1 workbench has been used for its simulation to study the stress distribution, after obtaining accurate dimensioning of knee components and generating a 3D computer-aided design model. An experimental structural strength test has been performed to validate the findings of the simulation. The results indicate that the tested prosthetic knee sample is mildly sufficient to outstrip the static strength test. The validation result has confirmed the accuracy of the finite element model. The mean absolute error percentage between finite element-based simulation and the experimental result is estimated to be 10%. Copyright (c) 2022 Elsevier Ltd. All rights reserved. Selection and peer-review under responsibility of the scientific committee of the 4th International Conference on Advances in Mechanical Engineering and Nanotechnology.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 50 条
  • [41] The Finite Element Analysis of Standard Knee Prosthesis
    Dong, Li-min
    Tian, Xi-yan
    Gu, You-quan
    Zhang, Chun-qiu
    MEMS, NANO AND SMART SYSTEMS, PTS 1-6, 2012, 403-408 : 2535 - 2538
  • [42] Finite element analysis of unicompartmental knee arthroplasty
    Hopkins, Andrew R.
    New, Andrew M.
    Rodriguez y Baena, Ferdinando
    Taylor, Mark
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (01) : 14 - 21
  • [43] A Finite Element Analysis of an Osteoarthritis Knee Brace
    Stamenovic, D.
    Kojic, M.
    Stojanovic, B.
    Hunter, D.
    JOURNAL OF THE SERBIAN SOCIETY FOR COMPUTATIONAL MECHANICS, 2008, 2 (02) : 29 - 41
  • [44] Improvement of a Knee Prosthesis Mechanism through Experimental Stress Analysis and the Finite Element Method
    Pasaguayo, L.
    Ciaccia, M.
    2017 IEEE SECOND ECUADOR TECHNICAL CHAPTERS MEETING (ETCM), 2017,
  • [45] Experimental validation of a finite element model of a composite tibia
    Gray, H. A.
    Zavatsky, A. B.
    Taddei, F.
    Cristofolini, L.
    Gill, H. S.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H3) : 315 - 324
  • [46] Experimental validation of finite element codes for welding deformations
    Aarbogh, H. M.
    Hamide, M.
    Fjaer, H. G.
    Mo, A.
    Bellet, M.
    JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2010, 210 (13) : 1681 - 1689
  • [47] Finite element simulation and experimental validation of fretting wear
    McColl, IR
    Ding, J
    Leen, SB
    WEAR, 2004, 256 (11-12) : 1114 - 1127
  • [48] Experimental results for the validation of dynamic finite element models
    Huber, G.
    Paetzold, H.
    Pueschel, K.
    Morlock, M. M.
    HUMAN VIBRATION: EFFECTS ON HEALTH - PERFORMANCE - COMFORT, 2007, 2002 : 249 - 264
  • [49] Experimental validation of a finite element model of the temporomandibular joint
    DeVocht, JW
    Goel, VK
    Zeitler, DL
    Lew, D
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2001, 59 (07) : 775 - 778
  • [50] A systematic review on design technology and application of polycentric prosthetic knee in amputee rehabilitation
    Rajesh Kumar Mohanty
    R C Mohanty
    Sukanta Kumar Sabut
    Physical and Engineering Sciences in Medicine, 2020, 43 : 781 - 798