Improving the Arterial Input Function in Dynamic Contrast Enhanced MRI by Fitting the Signal in the Complex Plane

被引:11
|
作者
Simonis, Frank F. J. [1 ]
Sbrizzi, Alessandro [2 ]
Beld, Ellis [1 ]
Lagendijk, Jan J. W. [1 ]
van den Berg, Cornelis A. T. [1 ]
机构
[1] Univ Med Ctr Utrecht, Imaging Div, Dept Radiotherapy, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiol, Utrecht, Netherlands
关键词
DCE-MRI; arterial input function; complex signal; fitting; DCE-MRI; T-1; BLOOD; PHASE; QUANTIFICATION; RELAXATION; TISSUE; MODEL; TESLA;
D O I
10.1002/mrm.26023
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dynamic contrast enhanced (DCE) imaging is a widely used technique in oncologic imaging. An essential prerequisite for obtaining quantitative values from DCE-MRI is the determination of the arterial input function (AIF). However, it is very challenging to accurately estimate the AIF using MR. A comprehensive model, which uses complex data instead of either magnitude or phase, was developed to improve AIF estimation. Theory and Methods: The model was first applied to simulated data. Subsequently, the accuracy of the estimated contrast agent concentration was validated in a phantom. Finally the method was applied to existing DCE scans of 13 prostate cancer patients. Results: The complex signal method combines the complementary strengths of the magnitude and phase method, increasing the precision and accuracy of concentration estimation in simulated and phantom data. The in vivo AIFs show a good agreement between arterial voxels (standard deviation in the peak and tail equal 0.4 mM and 0.12 mM, respectively). Furthermore, the dynamic behavior closely followed the AIF obtained with DCE-CT in the same patients (mean correlation coefficient: 0.92). Conclusion: By using the complex signal, the AIF estimation becomes more accurate and precise. This might enable patient specific AIFs, thereby improving the quantitative values obtained from DCE-MRI. (C) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:1236 / 1245
页数:10
相关论文
共 50 条
  • [41] Delay of Aortic Arterial Input Function Time Improves Detection of Malignant Vertebral Body Lesions on Dynamic Contrast-Enhanced MRI Perfusion
    Camelo, Felipe
    Peck, Kyung K.
    Saha, Atin
    Arevalo-Perez, Julio
    Lyo, John K.
    Tisnado, Jamie
    Lis, Eric
    Karimi, Sasan
    Holodny, Andrei I.
    CANCERS, 2023, 15 (08)
  • [42] Impact of Arterial Input Function Selection on the Accuracy of Dynamic Contrast-Enhanced MRI Quantitative Analysis for the Diagnosis of Clinically Significant Prostate Cancer
    Azahaf, Mustapha
    Haberley, Marc
    Betrouni, Nacim
    Ernst, Olivier
    Behal, Helene
    Duhamel, Alain
    Ouzzane, Adil
    Puech, Philippe
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 43 (03) : 737 - 749
  • [43] Impact of the arterial input function on microvascularization parameter measurements using dynamic contrast-enhanced ultrasonography
    Gauthier, Marianne
    Pitre-Champagnat, Stephanie
    Tabarout, Farid
    Leguerney, Ingrid
    Polrot, Melanie
    Lassau, Nathalie
    WORLD JOURNAL OF RADIOLOGY, 2012, 4 (07): : 291 - 301
  • [44] Dynamic contrast-enhanced MRI using Gd-DTPA: Interindividual variability of the arterial input function and consequences for the assessment of kinetics in tumors
    Port, RE
    Knopp, MV
    Brix, G
    MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (06) : 1030 - 1038
  • [45] Editorial for "Influence of Different Measurement Methods of Arterial Input Function on Quantitative Dynamic Contrast-Enhanced MRI Parameters in Head and Neck Cancer"
    Anzai, Yoshimi
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2023, 58 (01) : 133 - 134
  • [46] Pharmacokinetic modeling of dynamic contrast-enhanced MRI using a reference region and input function tail
    Ahmed, Zaki
    Levesque, Ives R.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 83 (01) : 286 - 298
  • [47] Use of Cardiac Output to Improve Measurement of Input Function in Quantitative Dynamic Contrast-Enhanced MRI
    Zhang, Jeff L.
    Rusinek, Henry
    Bokacheva, Louisa
    Chen, Qun
    Storey, Pippa
    Lee, Vivian S.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (03) : 656 - 665
  • [48] Determination of arterial input function in dynamic susceptibility contrast MRI using group independent component analysis technique
    Chen, Sharon
    Liu, Ho-Ling
    Yang, Yihong
    Hsu, Yuan-Yu
    Chuang, Keh-Shih
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02): : 617 - 621
  • [49] Automatic determination of the arterial input function in dynamic susceptibility contrast MRI: comparison of different reproducible clustering algorithms
    Jiandong Yin
    Jiawen Yang
    Qiyong Guo
    Neuroradiology, 2015, 57 : 535 - 543
  • [50] Automatic determination of the arterial input function in dynamic susceptibility contrast MRI: comparison of different reproducible clustering algorithms
    Yin, Jiandong
    Yang, Jiawen
    Guo, Qiyong
    NEURORADIOLOGY, 2015, 57 (05) : 535 - 543