摘要
Radioimmunoassay (RIA) is the "gold standard" method for evaluation of serum cortisol concentration. The VIDAS cortisol test is an enzyme-linked fluorescent assay designed for the MiniVidas system. The aim of this study was to compare the VIDAS method with RIA for measurement of bovine serum cortisol concentration. Cortisol concentrations were evaluated in 40 cows using both VIDAS and RIA methods, the latter as the reference method. A paired Student's t-test, Pearson's correlation analysis, Bland-Altman plot, and Deming regression analysis were used to compare the two methods. There was no statistically significant difference between mean serum cortisol concentrations measured by VIDAS or RIA methods (P = 0.6570). Both methods were able to detect significant differences in mean low and high cortisol concentrations (P < 0.00014 RIA and P < 0.0016 VIDAS). The correlation coefficient was low, but a Bland-Altman plot and Deming regression analysis show neither constant nor proportional error. The VIDAS method produced slightly higher values than RIA, but the difference was small and in no case did the mean value move the normal range. Results suggest that VIDAS method is suitable for the determination of bovine serum cortisol concentration in studies of large numbers of animals.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] THE DEVELOPMENT OF A DIRECT HOMOLOGOUS RADIOIMMUNOASSAY FOR SERUM CORTISOL
    STALLA, GK
    GIESEMANN, G
    MULLER, OA
    WOOD, WG
    SCRIBA, PC
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1981, 19 (07): : 427 - 434
  • [23] BOVINE FSH IN SERUM MEASURED BY RADIOIMMUNOASSAY
    AKBAR, AM
    REICHERT, LE
    DUNN, TG
    KALTENBACH, CC
    NISWENDER, GD
    JOURNAL OF ANIMAL SCIENCE, 1973, 37 (01) : 299 - 299
  • [24] Direct radioimmunoassay for the measurement of urinary cortisol.
    LupiChen, N
    Portuesi, S
    Barrande, C
    Cailla, H
    Fournier, C
    Mappus, E
    Cuilleron, CY
    CLINICAL CHEMISTRY, 1996, 42 (06) : 57 - 57
  • [25] COMPARISON OF COMPETITIVE PROTEIN-BINDING ANALYSIS AND RADIOIMMUNOASSAY FOR DETERMINATION OF CORTISOL IN SERUM AND URINE
    DEMERS, LM
    DERCK, DD
    CLINICAL BIOCHEMISTRY, 1977, 10 (03) : 104 - 108
  • [26] COMPARISON OF METHODS FOR SERUM APOPROTEIN MEASUREMENT
    MEUFFELS, MT
    WATSON, DC
    HINDMARSH, JT
    CLINICAL BIOCHEMISTRY, 1987, 20 (04) : 290 - 290
  • [27] MEASUREMENT OF SERUM OSTEONECTIN BY RADIOIMMUNOASSAY
    MALAVAL, L
    DELMAS, PD
    MEUNIER, PJ
    JOURNAL OF BONE AND MINERAL RESEARCH, 1986, 1 (01) : 126 - 126
  • [28] MEASUREMENT OF SERUM FERRITIN BY RADIOIMMUNOASSAY
    GOLDIE, DJ
    THOMAS, MJ
    ANNALS OF CLINICAL BIOCHEMISTRY, 1978, 15 (MAR) : 102 - 108
  • [29] RADIOIMMUNOASSAY FOR MEASUREMENT OF SERUM THYMIDINE
    CHRISTIN.M
    STOLLAR, D
    HUGHES, WL
    FEDERATION PROCEEDINGS, 1972, 31 (02) : A807 - &
  • [30] RADIOIMMUNOASSAY FOR MEASUREMENT OF SERUM THYMIDINE
    HUGHES, WL
    CHRISTINE, M
    STOLLAR, BD
    ANALYTICAL BIOCHEMISTRY, 1973, 55 (02) : 468 - 478