Transient and steady-state dark current mechanisms in polycrystalline mercuric iodide X-ray imaging detectors

被引:12
|
作者
Kabir, M. Z. [1 ]
机构
[1] Concordia Univ, Dept Elect & Comp Engn, Montreal, PQ H3G 1M8, Canada
关键词
X-ray imaging detectors; Polycrystalline mercuric iodide; Dark current; Carrier injections; Thermal generation; GAMMA-RAY; PHOTOCONDUCTORS; IMAGERS; HGI2; SI;
D O I
10.1016/j.nima.2013.11.014
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A theoretical model for describing bias-dependent time transient and steady-state dark current behaviors in polycrystalline mercuric iodide (poly-HgI2) based X-ray image detectors is developed. The model considers carrier injection from the metal electrode, bulk carrier depletion process, and bulk thermal generation current from the mid-gap states. The transient dark current is mainly determined by the initial carrier depletion process. At a very low applied field (less than 0.05 V/mu m), the steady-state dark current is almost equal to the bulk thermal generation current. However, the injection current increases sharply with increasing the applied field. The steady-state dark current in poly-HgI2 detectors at normal operating field (similar to 1 V/mu m) is mainly controlled by the Schottky emission of electrons from the metal/HgI2 contact. The fitting of the physics-based model to the experimental results estimates the effective barrier height and interface defect states for injecting electrons from the metal to poly-HgI2 layer in various poly-HgI2 detectors. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:156 / 160
页数:5
相关论文
共 50 条
  • [21] Large area mercuric iodide and lead iodide X-ray detectors for medical and non-destructive industrial imaging
    Zentai, G.
    Schieber, M.
    Partain, L.
    Pavlyuchkova, R.
    Proano, C.
    JOURNAL OF CRYSTAL GROWTH, 2005, 275 (1-2) : E1327 - E1331
  • [22] Large area mercuric iodide thick film X-ray detectors for fluoroscopic (on-line) imaging
    Zentai, G
    Partain, L
    Pavlyuchkova, R
    Proano, C
    Virshup, G
    Breen, BN
    Zuck, A
    Reisman, B
    Taieb, A
    Schieber, M
    SMART NONDESTRUCTIVE EVALUATION FOR HEALTH MONITORING OF STRUCTURAL AND BIOLOGICAL SYSTEMS, 2002, 4702 : 446 - 453
  • [23] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Schieber, M
    Hermon, H
    Zentai, G
    Partain, L
    Street, RA
    Gilboa, H
    RADIOLOGY, 2001, 221 : 464 - 464
  • [24] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Gilboa, H
    Zuck, A
    Dagan, O
    Vilensky, A
    Breen, BN
    Taieb, A
    Reisman, B
    Hermon, H
    Zentai, G
    Partain, L
    Street, R
    Ready, S
    X-RAY AND GAMMA-RAY DETECTORS AND APPLICATIONS IV, 2002, 4784 : 315 - 325
  • [25] A working device model of mercuric iodide x-ray detectors for XRF applications
    Bao, XJ
    Natarajan, M
    Henderson, J
    1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 75 - 79
  • [26] Steady-state and transient current transport in undoped polycrystalline diamond films
    Jauhiainen, A
    Bengtsson, S
    Engstrom, O
    JOURNAL OF APPLIED PHYSICS, 1997, 82 (10) : 4966 - 4976
  • [28] Quantitative evaluation of mercuric iodide and selenium for x-ray imaging device
    Cha, BY
    Park, JK
    Kang, SS
    Shin, JW
    Kim, JY
    Lee, HW
    Nam, SH
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4496 - 4500
  • [29] Red mercuric iodide crystals obtained by isothermal solution evaporation: Characterization for mammographic X-ray imaging detectors
    Caldeira, A. M. F.
    Ugucioni, J. C.
    Mulato, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 737 : 87 - 91
  • [30] Recent advances in gamma- and X-ray spectrometry by means of mercuric iodide detectors
    Ponpon, JP
    Sieskind, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2): : 173 - 178