A new stand-alone anterior lumbar interbody fusion device: Biomechanical comparison with established fixation techniques

被引:50
|
作者
Cain, CMJ
Schleicher, P
Gerlach, R
Pflugmacher, R
Scholz, M
Kandziora, F
机构
[1] Adelaide Spine clin, Adelaide, SA 5000, Australia
[2] Humboldt Univ, Univ Klinikum Charite, Unfall & Wiederherstellungschirurg, Berlin, Germany
关键词
anterior lumbar fusion; stand alone; lumbar fixation; biomechanical evaluation;
D O I
10.1097/01.brs.0000187897.25889.54
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Established lumbar fixation methods were assessed biomechanically, and a comparison was made with a new stand-alone anterior lumbar interbody cage device incorporating integrated anterior fixation. Objectives. To compare the stability of a new stand-alone anterior implant (Test-device) with established fixation methods to assess its suitability for clinical use. Our hypothesis being that the Test-device would provide stability comparable to that provided by an anterior cage when supplemented with posterior pedicle screw fixation. Summary of Background Data. It is accepted that the use of rigid pedicle screw instrumentation increases the chance of achieving a solid fusion, but its use may be associated with a significant increase in postoperative morbidity caused by disruption of the posterior musculature. It is also evident that this increased fusion rate is generally not associated with increased clinical success. This dilemma has led to a search for a solution and to the development of the Test-device anterior lumbar interbody device. Methods. The kinematic properties of either the L3 - L4 or L4 - L5 lumbar motion segment of 8 cadaveric lumbar spines have been tested using the following sequence of fixation: intact, Test-device, Test-device and translaminar facet screws (TS), Cage and TS, Cage and Universal Spine System (USS), and Cage and small stature USS. Results. All fixation techniques except the cage and TS decreased (P < 0.05) range of motion (ROM), neutral zone (NZ), and elastic zone (EZ), and increased (P < 0.05) stiffness in comparison to the intact motion segment in all test modes. There was a significant increase (P < 0.01) in the ROM, NZ, and EZ, and decrease in the stiffness of the cage and TS group in comparison to all other stabilization techniques in flexion and rotation. There was no significant difference in the ROM, NZ, EZ, and stiffness between the Test-device and cage and USS groups in flexion, extension, and bending. The Test-device resulted in a significantly lower EZ (P < 0.05) and a significantly higher stiffness (P < 0.05) in rotation than all other fixation methods. Conclusions. The Test-device alone provided similar and the Test-device and TS higher stability than the pedicle screw constructs evaluated. These results support progression to clinical trials using the Test-device as a stand-alone implant.
引用
收藏
页码:2631 / 2636
页数:6
相关论文
共 50 条
  • [41] Biomechanical comparison: stability of lateral-approach anterior lumbar interbody fusion and lateral fixation compared with anterior-approach anterior lumbar interbody fusion and posterior fixation in the lower lumbar spine
    Kim, SM
    Lim, TJ
    Paterno, J
    Park, J
    Kim, DH
    JOURNAL OF NEUROSURGERY-SPINE, 2005, 2 (01) : 62 - 68
  • [42] Impact of cage position on biomechanical performance of stand-alone lateral lumbar interbody fusion: a finite element analysis
    Nan, Chong
    Ma, Zhanbei
    Liu, Yuxiu
    Ma, Liang
    Li, Jiaqi
    Zhang, Wei
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [43] Impact of cage position on biomechanical performance of stand-alone lateral lumbar interbody fusion: a finite element analysis
    Chong Nan
    Zhanbei Ma
    Yuxiu Liu
    Liang Ma
    Jiaqi Li
    Wei Zhang
    BMC Musculoskeletal Disorders, 23
  • [44] When to Consider Stand-Alone Lateral Lumbar Interbody Fusion: Is There a Role for a Comeback With New Implants?
    Menezes, Cristiano Magalhaes
    Menezes, Erica Godinho
    Asghar, Jahangir
    Guiroy, Alfredo
    INTERNATIONAL JOURNAL OF SPINE SURGERY, 2022, 16 : S69 - S75
  • [45] In Reply to the Letter to the Editor Regarding "Biomechanical Comparison of Stand-Alone and Bilateral Pedicle Screw Fixation for Oblique Lumbar Interbody Fusion Surgery-A Finite Element Analysis"
    Fang, Guofang
    Sang, Hongxun
    WORLD NEUROSURGERY, 2021, 150 : 250 - 250
  • [46] Non-Union Rate With Stand-Alone Lateral Lumbar Interbody Fusion
    Watkins, Robert
    Watkins, Robert, III
    Hanna, Robert
    MEDICINE, 2014, 93 (29) : e275
  • [47] Long-Term Durability of Stand-Alone Lateral Lumbar Interbody Fusion
    Agarwal, Nitin
    White, Michael D.
    Roy, Souvik
    Ozpinar, Alp
    Alan, Nima
    Lavadi, Raj Swaroop
    Okonkwo, David O.
    Hamilton, D. Kojo
    Kanter, Adam S.
    NEUROSURGERY, 2023, 93 (01) : 60 - 65
  • [48] Effect of supplemental translaminar facet screw fixation on the stability of stand-alone anterior lumbar interbody fusion cages under physiologic compressive preloads
    Phillips, FM
    Cunningham, B
    Carandang, G
    Ghanayem, AJ
    Voronov, L
    Havey, RM
    Patwardhan, AG
    SPINE, 2004, 29 (16) : 1731 - 1736
  • [49] Clinical and radiologic outcomes of stand-alone anterior lumbar interbody fusion at L4-L5
    Szadkowski, Marc
    Bahroun, Sami
    Aleksic, Ivan
    Vande Kerckhove, Michiel
    Ramos-Pascual, Sonia
    Fiere, Vincent
    D'Astorg, Henri
    INTERDISCIPLINARY NEUROSURGERY-ADVANCED TECHNIQUES AND CASE MANAGEMENT, 2021, 26
  • [50] Stand-alone anterior lumbar interbody fusion for treatment of degenerative spondylolisthesis (vol 22, pg 1619, 2015)
    Rao, Prashanth J.
    Ghent, Finn
    Phan, Kevin
    Lee, Sun Loong Keegan
    Reddy, Rajesh
    Mobbs, Ralph J.
    JOURNAL OF CLINICAL NEUROSCIENCE, 2019, 60 : 176 - 176