Treatment vault shielding for a flattening filter-free medical linear accelerator

被引:29
|
作者
Kry, Stephen F. [1 ]
Howell, Rebecca M. [1 ]
Polf, Jerimy [1 ]
Mohan, Radhe [1 ]
Vassiliev, Oleg N. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 05期
关键词
UNFLATTENED PHOTON BEAMS; DOSIMETRIC PROPERTIES; MONTE-CARLO;
D O I
10.1088/0031-9155/54/5/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The requirements for shielding a treatment vault with a Varian Clinac 2100 medical linear accelerator operated both with and without the flattening filter were assessed. Basic shielding parameters, such as primary beam tenth-value layers (TVLs), patient scatter fractions, and wall scatter fractions, were calculated using Monte Carlo simulations of 6, 10 and 18 MV beams. Relative integral target current requirements were determined from treatment planning studies of several disease sites with, and without, the flattening filter. The flattened beam shielding data were compared to data published in NCRP Report No. 151, and the unflattened beam shielding data were presented relative to the NCRP data. Finally, the shielding requirements for a typical treatment vault were determined for a single-energy ( 6 MV) linac and a dual-energy ( 6 MV/18 MV) linac. With the exception of large-angle patient scatter fractions and wall scatter fractions, the vault shielding parameters were reduced when the flattening filter was removed. Much of this reduction was consistent with the reduced average energy of the FFF beams. Primary beam TVLs were reduced by 12%, on average, and small-angle scatter fractions were reduced by up to 30%. Head leakage was markedly reduced because less integral target current was required to deliver the target dose. For the treatment vault examined in the current study, removal of the flattening filter reduced the required thickness of the primary and secondary barriers by 10-20%, corresponding to 18 m(3) less concrete to shield the single-energy linac and 36 m3 less concrete to shield the dual-energy linac. Thus, a shielding advantage was found when the linac was operated without the flattening filter. This translates into a reduction in occupational exposure and/or the cost and space of shielding.
引用
收藏
页码:1265 / 1273
页数:9
相关论文
共 50 条
  • [1] Radiation safety survey on a flattening filter-free medical accelerator
    Vassiliev, O. N.
    Titt, U.
    Kry, S. F.
    Mohan, R.
    Gillin, M. T.
    RADIATION PROTECTION DOSIMETRY, 2007, 124 (02) : 187 - 190
  • [2] A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements
    Dalaryd, Marten
    Kragl, Gabriele
    Ceberg, Crister
    Georg, Dietmar
    McClean, Brendan
    af Wetterstedt, Sacha
    Wieslander, Elinore
    Knoos, Tommy
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23): : 7333 - 7344
  • [3] Impact of a flattening filter free linear accelerator on structural shielding design
    Jank, Julia
    Kragl, Gabriele
    Georg, Dietmar
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2014, 24 (01): : 38 - 48
  • [4] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Mesbahi, Asghar
    Nejad, Farshad Seyed
    RADIATION MEDICINE, 2008, 26 (06): : 331 - 336
  • [5] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Asghar Mesbahi
    Farshad Seyed Nejad
    Radiation Medicine, 2008, 26 : 331 - 336
  • [6] Radiotherapy Vault Shielding Evaluation Method for a Flattening Filter-Free (FFF) Linac-Practical Considerations and Recommendations
    Mihailidis, D.
    MEDICAL PHYSICS, 2015, 42 (06) : 3490 - 3490
  • [7] Flattening Filter Free RapidArc Treatment with Trilogy MX Linear Accelerator
    Hrbacek, J.
    Lang, S.
    Dossenbach, T.
    Bocanek, J.
    Klock, S.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [8] Acceptance criteria for flattening filter-free photon beam from standard medical electron linear accelerator: AERB task group recommendations
    Sahani, G.
    Sharma, S. D.
    Sharma, P. K. Dash
    Deshpande, D. D.
    Negi, P. S.
    Sathianarayanan, V. K.
    Rath, G. K.
    JOURNAL OF MEDICAL PHYSICS, 2014, 39 (04) : 206 - 211
  • [9] Development of a dose-rate dosimeter using a silicon photodiode for a medical linear accelerator in a 10 MV flattening filter-free mode
    Yamaguchi, Satoshi
    Ariga, Hisanori
    Yoshioka, Kunihiro
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2024, 95 (05):
  • [10] A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment
    Ali Wagdy
    Khaled EI Shahat
    Huda Ashry
    Amaal El Shershaby
    Ehab Abdo
    OncologyandTranslationalMedicine, 2017, 3 (06) : 260 - 266