Determination of coal ash content by the combined x-ray fluorescence and scattering spectrum

被引:12
|
作者
Mikhailov, I. F. [1 ]
Baturin, A. A. [1 ]
Mikhailov, A. I. [1 ]
Borisova, S. S. [1 ]
Fomina, L. P. [1 ]
机构
[1] Natl Tech Univ, Kharkov Polytech Inst, Dept Phys Met & Semicond, Kharkov, Ukraine
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2018年 / 89卷 / 02期
关键词
ELEMENT CONTENT; RAYLEIGH; RATIO;
D O I
10.1063/1.4993101
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An alternative method is proposed for the determination of the inorganic constituent mass fraction (ash) in solid fuel by the ratio of Compton and Rayleigh X-ray scattering peaks IC/IR subject to the iron fluorescence intensity. An original X-ray optical scheme with a Ti/Mo (or Sc/Cu) double-layer secondary radiator allows registration of the combined fluorescence-and-scattering spectrum at the specified scattering angle. An algorithm for linear calibration of the Compton-to-Rayleigh IC/IR ratio is proposed which uses standard samples with two certified characteristics: mass fractions of ash (A(d)) and iron oxide (W-Fe2O3). Ash mass fractions have been determined for coals of different deposits in the wide range of Ad from 9.4% to 52.7% mass andW(Fe2O3) from 0.3% to 4.95% mass. Due to the high penetrability of the probing radiation with energy E > 17 keV, the sample preparation procedure is rather simplified in comparison with the traditional method of Ad determination by the sum of fluorescence intensities of all constituent elements. Published by AIP Publishing.