Eight years of left ventricle pacing due to inadvertent malposition of a transvenous pacemaker lead in the left ventricle

被引:17
|
作者
Vanhercke, Daniel [1 ]
Heytens, Wendy [1 ]
Verloove, Hugues [1 ]
机构
[1] AZ Sint Lucas, Dept Cardiol, BE-9000 Ghent, Belgium
来源
关键词
Left ventricle; Pacemaker; Malpositioned pacemaker lead;
D O I
10.1093/ejechocard/jen187
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The inadvertent malposition of a pacemaker lead in the left ventricle is a rare and underdiagnosed pacemaker complication. A 78-year-old woman was admitted to our Emergency Department for progressive dyspnea, 8 years after transvenous pacemaker implantation. Routine 12-lead electrocardiography revealed a right bundle branch block on the paced beats, and lateral chest X-ray showed posterior deflection of the pacemaker lead, suggesting a pacemaker electrode in the left ventricle. Echocardiography confirmed that the pacing lead had migrated through the foramen ovale into the left ventricle. After review of the literature and taking into account the age and emboligenic-free history of the patient under platelet therapy, we decided to postpone the surgical removal of the malpositioned lead in the left ventricle until the need arose for cardiac surgery for additional reasons, such as valve and/or coronary illness. The general therapeutic strategy remained unchanged due to the risk of bleeding complications with anticoagulation in the patient. This case illustrates the prominent role of echocardiography in the diagnosis of unusual cardiac events.
引用
收藏
页码:825 / 827
页数:3
相关论文
共 50 条
  • [1] Malposition of transvenous pacing lead in the left ventricle
    Raghavan, C
    Cashion, WR
    Spencer, WH
    CLINICAL CARDIOLOGY, 1996, 19 (04) : 335 - 338
  • [2] MALPOSITION OF TRANSVENOUS PACING LEAD IN THE LEFT-VENTRICLE
    GHANI, M
    THAKUR, RK
    BOUGHNER, D
    MORILLO, CA
    YEE, R
    KLEIN, GJ
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (09): : 1800 - 1807
  • [3] Malposition of the pacemaker lead in the left ventricle
    Benito Martin, Eva M.
    Rubin Lopez, Jose M.
    Corros Vicente, Cecilia
    De La Hera Galarza, Jesus M.
    Martin Fernandez, Maria
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2013, 32 (7-8) : 633 - 635
  • [4] MALPOSITION OF TRANSVENOUS PACING LEAD IN THE LEFT-VENTRICLE - RADIOGRAPHIC FINDINGS
    BAUERSFELD, UK
    THAKUR, RK
    GHANI, M
    YEE, R
    KLEIN, GJ
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 162 (02) : 290 - 292
  • [5] Incidence, predictors, and outcome of inadvertent malposition of transvenous pacing or defibrillation lead in the left heart
    Ohlow, Marc-Alexander
    Roos, Marcus
    Lauer, Bernward
    Von Korn, Hubertus
    Geller, Johann Christoph
    EUROPACE, 2016, 18 (07): : 1049 - 1054
  • [6] Inadvertent implantation of a pacemaker lead in the left ventricle: A new challenge in cardiology
    Calvagna, Giuseppe Mario
    Patane, Salvatore
    Ceresa, Fabrizio
    Fontana, Alessandro
    Sicuso, Giovanni
    Vinci, Eugenio
    Muscio, Gianfranco
    Vasquez, Ludovico
    Patane, Francesco
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 202 : 914 - 917
  • [7] INADVERTENT MALPOSITION OF A TRANSVENOUS-INSERTED PACING LEAD IN THE LEFT-VENTRICULAR CHAMBER
    SHARIFI, M
    SORKIN, R
    SHARIFI, V
    LAKIER, JB
    AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (01): : 92 - 95
  • [8] Six uneventful years with a pacing lead in the left ventricle
    Ling, Lee Fong
    Lever, Harry
    HEART RHYTHM, 2013, 10 (04) : 614 - 615
  • [9] Inadvertent malposition of a permanent pacemaker ventricular lead into the left ventricle which was initially missed and diagnosed two years later: A case report
    Zaher M.F.
    Azab B.N.
    Bogin M.B.
    Bekheit S.G.
    Journal of Medical Case Reports, 5 (1)
  • [10] Do all roads lead to Rome? Treatment of malposition pacemaker lead in the left ventricle
    Schmiady, Martin Oliver
    Hofmann, Michael
    Maisano, Francesco
    Morjan, Mohammed
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2020, 57 (05) : 1009 - 1010