Observational Study of Humidified High-Flow Nasal Cannula Compared with Nasal Continuous Positive Airway Pressure

被引:157
|
作者
Lampland, Andrea L. [1 ]
Plumm, Brenda [1 ]
Meyers, Patricia A. [1 ]
Worwa, Cathy T. [1 ]
Mammel, Mark C. [1 ,2 ]
机构
[1] Childrens Hosp & Clin Minnesota, Infant Diagnost & Res Ctr, St Paul, MN USA
[2] Univ Minnesota, Sch Med, Dept Pediat, Div Neonatol, Minneapolis, MN 55455 USA
来源
JOURNAL OF PEDIATRICS | 2009年 / 154卷 / 02期
关键词
EXPIRATORY PRESSURE; PHARYNGEAL PRESSURE; ESOPHAGEAL PRESSURE; INFANTS; BIRTH; CPAP; INTUBATION; MANAGEMENT;
D O I
10.1016/j.jpeds.2008.07.021
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objectives To conduct an in Nitro evaluation of a humidified high-flow nasal cannula (HFNC) system at different flows, cannula sizes, and air leaks and also an in vivo analysis of mean end-expiratory esophageal pressure (EEEP) from nasal continuous positive airway pressure at 6 cm H2O (NCPA+6) versus HFNC. Study design In the in vitro study, we measured HFNC system pressure and flow, with varying degrees of leak and with and without the use of a pressure-limiting valve. In the in vivo study, we measured EEEP in 15 newborns on NCPAP+6 and then on HFNC at 6 L/minute, with flow decreased by 1 L/minute every 30 minutes. Heart rate, respiratory rate, fraction of inspired oxygen, arterial oxygen saturation, respiratory distress syndrome score, and EEEP were recorded for each intervention. Data analysis was done using repeated-measures analysis of variance and linear regression. Results In the in Nitro study. in the absence of leads, the pressures were limited by the pressure-limiting valve only at flows : 2 L/minute. With leaks of 30% and 50%, delivered pressures were always < 3 cm H2O. In the in vivo study, respiratory rate increased from baseline (NCPAP+6) as flow decreased (P < .02). Intrapatient and interpatient coefficients of variation were always high. Conclusions A pressure-limiting valve is necessary in a HFNC system. Although mean EEEP levels were similar in NCPAP+6 and HFNC, tachypnea developed as flow diminished. This system apparently cannot predict EEEP, because of interpatient and intrapatient variation. (J Pediatr 2009;154:177-82)
引用
收藏
页码:177 / 182
页数:6
相关论文
共 50 条
  • [31] A comparison of nasal trauma in preterm infants extubated to either heated humidified high-flow nasal cannulae or nasal continuous positive airway pressure
    C. L. Collins
    C. Barfield
    R. S. C. Horne
    P. G. Davis
    European Journal of Pediatrics, 2014, 173 : 181 - 186
  • [32] Heated, humidified high-flow nasal cannula therapy
    Weiner, Daniel J.
    McDonough, Joseph
    Perez, Myrza
    Evans, Jacquelyn
    Fox, William
    Hedgman, Angela
    Tyler, Lisa
    Panitch, Howard B.
    PEDIATRICS, 2008, 121 (06) : 1293 - +
  • [33] Work of breathing indices in infants with respiratory insufficiency receiving high-flow nasal cannula and nasal continuous positive airway pressure
    B E de Jongh
    R Locke
    A Mackley
    J Emberger
    D Bostick
    J Stefano
    E Rodriguez
    T H Shaffer
    Journal of Perinatology, 2014, 34 : 27 - 32
  • [34] Work of breathing indices in infants with respiratory insufficiency receiving high-flow nasal cannula and nasal continuous positive airway pressure
    de Jongh, B. E.
    Locke, R.
    Mackley, A.
    Emberger, J.
    Bostick, D.
    Stefano, J.
    Rodriguez, E.
    Shaffer, T. H.
    JOURNAL OF PERINATOLOGY, 2014, 34 (01) : 27 - 32
  • [35] High-flow nasal cannula oxygen and nasal continuous positive airway pressure and full oral feeding in infants with bronchopulmonary dysplasia
    Shetty, Sandeep
    Hunt, Katie
    Douthwaite, Amy
    Athanasiou, Maria
    Hickey, Ann
    Greenough, Anne
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2016, 101 (05): : F408 - F411
  • [36] To Compare the Efficacy of Heated Humidified High-Flow Nasal Cannula and Continuous Positive Airway Pressure in Post-Extubation Period in VLBW Infants
    Garg, Bhawan Deep
    Bajaj, Naveen
    Sharma, Deepak
    JOURNAL OF TROPICAL PEDIATRICS, 2018, 64 (04) : 262 - 266
  • [37] The Difference between High Flow Nasal Cannula and Continuous Positive Airway Pressure
    Goldstein, Mitchell R.
    Kadri, Munaf
    Merritt, T. Allen
    Fayard, Elba
    Peverini, Ricardo
    PEDIATRICS, 2018, 142
  • [38] High flow nasal cannula: an alternative to continuous positive airway pressure in cats
    Baudin, Florent
    Pouzot-Nevoret, Celine
    Hocine, Leila
    Allaouchiche, Bernard
    Staffieri, Francesco
    JOURNAL OF FELINE MEDICINE AND SURGERY, 2021, 23 (04) : 405 - 406
  • [39] Effectiveness of Humidified High Flow Nasal Cannula Versus Continuous Nasal Positive Airway Pressure in Managing Respiratory Failure in Preterm Infants: An Emergency Department Study
    Mahboob, Duaa Yousof
    Hassan, Amber
    Naheed, Faiza
    Ali Shah, Arshad
    Siddiqui, Maria Fareed
    BIOMEDICINES, 2025, 13 (03)
  • [40] High Flow Nasal Cannula as a Method for Rapid Weaning From Nasal Continuous Positive Airway Pressure
    Badiee, Zohreh
    Eshghi, Alireza
    Mohammadizadeh, Majid
    INTERNATIONAL JOURNAL OF PREVENTIVE MEDICINE, 2015, 6