Observational Study of Humidified High-Flow Nasal Cannula Compared with Nasal Continuous Positive Airway Pressure

被引:157
|
作者
Lampland, Andrea L. [1 ]
Plumm, Brenda [1 ]
Meyers, Patricia A. [1 ]
Worwa, Cathy T. [1 ]
Mammel, Mark C. [1 ,2 ]
机构
[1] Childrens Hosp & Clin Minnesota, Infant Diagnost & Res Ctr, St Paul, MN USA
[2] Univ Minnesota, Sch Med, Dept Pediat, Div Neonatol, Minneapolis, MN 55455 USA
来源
JOURNAL OF PEDIATRICS | 2009年 / 154卷 / 02期
关键词
EXPIRATORY PRESSURE; PHARYNGEAL PRESSURE; ESOPHAGEAL PRESSURE; INFANTS; BIRTH; CPAP; INTUBATION; MANAGEMENT;
D O I
10.1016/j.jpeds.2008.07.021
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objectives To conduct an in Nitro evaluation of a humidified high-flow nasal cannula (HFNC) system at different flows, cannula sizes, and air leaks and also an in vivo analysis of mean end-expiratory esophageal pressure (EEEP) from nasal continuous positive airway pressure at 6 cm H2O (NCPA+6) versus HFNC. Study design In the in vitro study, we measured HFNC system pressure and flow, with varying degrees of leak and with and without the use of a pressure-limiting valve. In the in vivo study, we measured EEEP in 15 newborns on NCPAP+6 and then on HFNC at 6 L/minute, with flow decreased by 1 L/minute every 30 minutes. Heart rate, respiratory rate, fraction of inspired oxygen, arterial oxygen saturation, respiratory distress syndrome score, and EEEP were recorded for each intervention. Data analysis was done using repeated-measures analysis of variance and linear regression. Results In the in Nitro study. in the absence of leads, the pressures were limited by the pressure-limiting valve only at flows : 2 L/minute. With leaks of 30% and 50%, delivered pressures were always < 3 cm H2O. In the in vivo study, respiratory rate increased from baseline (NCPAP+6) as flow decreased (P < .02). Intrapatient and interpatient coefficients of variation were always high. Conclusions A pressure-limiting valve is necessary in a HFNC system. Although mean EEEP levels were similar in NCPAP+6 and HFNC, tachypnea developed as flow diminished. This system apparently cannot predict EEEP, because of interpatient and intrapatient variation. (J Pediatr 2009;154:177-82)
引用
收藏
页码:177 / 182
页数:6
相关论文
共 50 条
  • [1] Nasal Continuous Positive Airway Pressure or High-Flow Nasal Cannula?
    Ramanathan, R.
    Biniwale, M.
    NEONATOLOGY, 2014, 106 (03) : 268 - 269
  • [2] Physiologic Basis for Nasal Continuous Positive Airway Pressure, Heated and Humidified High-Flow Nasal Cannula, and Nasal Ventilation
    Dysart, Kevin C.
    CLINICS IN PERINATOLOGY, 2016, 43 (04) : 621 - +
  • [3] Nasal Continuous Positive Airway Pressure and High-Flow Nasal Cannula Today
    Dani, Carlo
    CLINICS IN PERINATOLOGY, 2021, 48 (04) : 711 - 724
  • [4] High-flow nasal cannula compared with continuous positive airway pressure: a bench and physiological study
    Vieira, Fernando
    Bezerra, Frank Silva
    Coudroy, Remi
    Schreiber, Annia
    Telias, Irene
    Dubo, Sebastian
    Cavalot, Giulia
    Pereira, Sergio Martins
    Piraino, Thomas
    Brochard, Laurent Jean
    JOURNAL OF APPLIED PHYSIOLOGY, 2022, 132 (06) : 1580 - 1590
  • [5] Comparison of the Heated Humidified High-flow Nasal Cannula with Nasal Continuous Positive Airway Pressure as Primary Respiratory Support for Preterm Neonates: A Prospective Observational Study
    Skariah, Tisha Ann
    Dias, Laveena
    Lewis, Leslie Edward
    IRANIAN JOURNAL OF NEONATOLOGY, 2019, 10 (03) : 51 - 57
  • [6] Heated Humidified High-Flow Nasal Cannulae: A Good Alternative to Nasal Continuous Positive Airway Pressure
    Ignacio, Lena
    AlFaleh, Khalid
    JOURNAL OF CLINICAL NEONATOLOGY, 2014, 3 (03) : 139 - U56
  • [7] Positive distending pressure generated by high flow nasal cannula as compared to nasal continuous positive airway pressure
    Lampland, A.
    Plumm, B.
    Worwa, C.
    Meyers, P.
    Mammel, M. C.
    JOURNAL OF INVESTIGATIVE MEDICINE, 2008, 56 (01) : 140 - 141
  • [8] High-flow nasal cannula for children not compliant with continuous positive airway pressure
    Amaddeo, Alessandro
    Khirani, Sonia
    Frapin, Annick
    Teng, Theo
    Griffon, Lucie
    Fauroux, Brigitte
    SLEEP MEDICINE, 2019, 63 : 24 - 28
  • [9] High-flow nasal cannula compared with continuous positive airway pressure in the treatment of obstructive sleep apnea
    Chung-Chieh Yu
    Chih-Yu Huang
    Chung-Ching Hua
    Huang-Pin Wu
    Sleep and Breathing, 2022, 26 : 549 - 558
  • [10] High-flow nasal cannula compared with continuous positive airway pressure in the treatment of obstructive sleep apnea
    Yu, Chung-Chieh
    Huang, Chih-Yu
    Hua, Chung-Ching
    Wu, Huang-Pin
    SLEEP AND BREATHING, 2022, 26 (02) : 549 - 558