Separation of benzodiazepines by electrospray ionization ion mobility spectrometry-mass spectrometry

被引:28
|
作者
Matz, LM [1 ]
Hill, HH [1 ]
机构
[1] Washington State Univ, Dept Chem, Pullman, WA 99164 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
electrospray; ion mobility spectrometry; benzodiazepines;
D O I
10.1016/S0003-2670(02)00021-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Benzodiazepines are a commonly abused class of drugs; requiring analytical techniques that can separate and detect the drugs in a rapid time period. In this paper, the two-dimensional separation of five benzodiazepines was shown by electrospray ionization (ESI) ion mobility spectrometry (IMS)-mass spectrometry (MS). In this study, both the two dimensions of separation (m/z and mobility) and the high resolution of our IMS instrument enabled confident identification of each of the five benzodiazepines studied. This was a significant improvement over previous IMS studies that could not separate many of the analytes due to low instrumental resolution. The benzodiazepines that contain a hydroxyl group in their molecular structure (lorazepam and oxazepam) were found to form both the protonated molecular ion and dehydration product as predominant ions. Experiments to isolate the parametric reasons for the dehydration ion formation showed that it was not the result of corona discharge processes or the potential applied to the needle. However, the potential difference between the needle and first drift ring did influence both the relative intensity ratios of the two ions and the ion sensitivity. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:235 / 245
页数:11
相关论文
共 50 条
  • [21] Improvement of the Ion Transfer Efficiency in Ion Mobility Spectrometry-Mass Spectrometry
    El-Shafie, Mahmoud Y.
    Bebawi, Sally
    Zomor, Hussein H.
    Gunzer, Frank
    2016 IEEE WORKSHOP ON ENVIRONMENTAL, ENERGY, AND STRUCTURAL MONITORING SYSTEMS (EESMS), 2016,
  • [22] Integrating the potential of ion mobility spectrometry-mass spectrometry in the separation and structural characterisation of lipid isomers
    Camunas-Alberca, Sandra M.
    Moran-Garrido, Maria
    Saiz, Jorge
    Gil-de-la-Fuente, Alberto
    Barbas, Coral
    Gradillas, Ana
    FRONTIERS IN MOLECULAR BIOSCIENCES, 2023, 10
  • [23] Ambient molecular imaging by laser ablation electrospray ionization mass spectrometry with ion mobility separation
    Li, Hang
    Smith, Brian K.
    Mark, Laszlo
    Nemes, Peter
    Nazarian, Javad
    Vertes, Akos
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2015, 377 : 681 - 689
  • [24] Ion Mobility Spectrometry-Mass Spectrometry Analysis for the Site of Aromatic Hydroxylation
    Shimizu, Atsushi
    Chiba, Masato
    DRUG METABOLISM AND DISPOSITION, 2013, 41 (07) : 1295 - 1299
  • [25] Perspective on Developments in Ion Mobility Spectrometry-Mass Spectrometry Combined with Ion Spectroscopy
    Wang N.
    Su W.
    Zhang D.
    Jiang Y.
    Fang X.
    Yue L.
    Journal of Chinese Mass Spectrometry Society, 2022, 43 (05) : 635 - 642
  • [26] ELECTROSPRAY-IONIZATION ION MOBILITY SPECTROMETRY
    WITTMER, D
    LUCKENBILL, BK
    HILL, HH
    CHEN, YH
    ANALYTICAL CHEMISTRY, 1994, 66 (14) : 2348 - 2355
  • [27] The potential for clinical applications using a new ionization method combined with ion mobility spectrometry-mass spectrometry
    Inutan, Ellen D.
    Wager-Miller, James
    Narayan, Srinivas B.
    Mackie, Ken
    Trimpin, Sarah
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2013, 16 (02) : 145 - 159
  • [28] Electrospray Ionization Ion Mobility Mass Spectrometry of Human Brain Gangliosides
    Sarbu, Mirela
    Robu, Adrian C.
    Ghiulai, Roxana M.
    Vukelic, Zeljka
    Clemmer, David E.
    Zamfir, Alina D.
    ANALYTICAL CHEMISTRY, 2016, 88 (10) : 5166 - 5178
  • [29] Sterically hindered phenols in negative ion mobility spectrometry-mass spectrometry
    Laakia, Jaakko
    Pedersen, Christian Schack
    Adamov, Alexey
    Viidanoja, Jyrki
    Sysoev, Alexey
    Kotiaho, Tapio
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (19) : 3069 - 3076
  • [30] Tetraalkylammonium halides as chemical standards for positive electrospray ionization with ion mobility spectrometry/mass spectrometry
    Viidanoja, J
    Sysoev, A
    Adamov, A
    Kotiaho, T
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (21) : 3051 - 3055