In vitro evaluation of the imaging accuracy of C-arm conebeam CT in cerebral perfusion imaging

被引:4
|
作者
Ganguly, A. [1 ]
Fieselmann, A. [2 ,3 ,4 ]
Boese, J. [4 ]
Rohkohl, C. [2 ,4 ]
Hornegger, J. [2 ,3 ]
Fahrig, R. [1 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Univ Erlangen Nurnberg, Dept Comp Sci, Pattern Recognit Lab, D-91058 Erlangen, Germany
[3] Univ Erlangen Nurnberg, Erlangen Grad Sch Adv Opt Technol SAOT, D-91058 Erlangen, Germany
[4] Siemens AG, Healthcare Sect, D-91301 Forchheim, Germany
基金
美国国家卫生研究院;
关键词
perfusion CT; C-arm; conebeam CT; interventional imaging; x-ray; arterial input function; venous input function; sinusoid; HIGH-RESOLUTION MEASUREMENT; TRACER BOLUS PASSAGES; RECONSTRUCTION; STROKE;
D O I
10.1118/1.4757910
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors have developed a method to enable cerebral perfusion CT imaging using C-arm based conebeam CT (CBCT). This allows intraprocedural monitoring of brain perfusion during treatment of stroke. Briefly, the technique consists of acquiring multiple scans (each scan comprised of six sweeps) acquired at different time delays with respect to the start of the x-ray contrast agent injection. The projections are then reconstructed into angular blocks and interpolated at desired time points. The authors have previously demonstrated its feasibility in vivo using an animal model. In this paper, the authors describe an in vitro technique to evaluate the accuracy of their method for measuring the relevant temporal signals. Methods: The authors' evaluation method is based on the concept that any temporal signal can be represented by a Fourier series of weighted sinusoids. A sinusoidal phantom was developed by varying the concentration of iodine as successive steps of a sine wave. Each step corresponding to a different dilution of iodine contrast solution contained in partitions along a cylinder. By translating the phantom along the axis at different velocities, sinusoidal signals at different frequencies were generated. Using their image acquisition and reconstruction algorithm, these sinusoidal signals were imaged with a C-arm system and the 3D volumes were reconstructed. The average value in a slice was plotted as a function of time. The phantom was also imaged using a clinical CT system with 0.5 s rotation. C-arm CBCT results using 6, 3, 2, and 1 scan sequences were compared to those obtained using CT. Data were compared for linear velocities of the phantom ranging from 0.6 to 1 cm/s. This covers the temporal frequencies up to 0.16 Hz corresponding to a frequency range within which 99% of the spectral energy for all temporal signals in cerebral perfusion imaging is contained. Results: The errors in measurement of temporal frequencies are mostly below 2% for all multiscan sequences. For single scan sequences, the errors increase sharply beyond 0.10 Hz. The amplitude errors increase with frequency and with decrease in the number of scans used. Conclusions: Our multiscan perfusion CT approach allows low errors in signal frequency measurement. Increasing the number of scans reduces the amplitude errors. A two-scan sequence appears to offer the best compromise between accuracy and the associated total x-ray and iodine dose. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4757910]
引用
收藏
页码:6652 / 6659
页数:8
相关论文
共 50 条
  • [1] C-Arm Conebeam CT Perfusion Imaging in the Angiographic Suite: A Comparison with Multidetector CT Perfusion Imaging
    Niu, K.
    Yang, P.
    Wu, Y.
    Struffert, T.
    Doerfler, A.
    Schafer, S.
    Royalty, K.
    Strother, C.
    Chen, G. -H.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (07) : 1303 - 1309
  • [2] C-arm CT Perfusion Imaging in the Interventional Suite
    Fieselmann, Andreas
    Manhart, Michael
    CURRENT MEDICAL IMAGING, 2013, 9 (02) : 96 - 101
  • [3] Evaluating the Feasibility of C-arm CT for Brain Perfusion Imaging: An in vitro Study
    Ganguly, A.
    Fieselmann, A.
    Boese, J.
    Rohkohl, C.
    Hornegger, J.
    Fahrig, R.
    MEDICAL IMAGING 2010: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, 2010, 7625
  • [4] Cerebral CT Perfusion Using an Interventional C-Arm Imaging System: Cerebral Blood Flow Measurements
    Ganguly, A.
    Fieselmann, A.
    Marks, M.
    Rosenberg, J.
    Boese, J.
    Deuerling-Zheng, Y.
    Straka, M.
    Zaharchuk, G.
    Bammer, R.
    Fahrig, R.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (08) : 1525 - 1531
  • [5] CT Imaging with a mobile C-arm prototype
    Cheryauka, Arvi
    Tubbs, Dave
    Langille, Vinton
    Kalya, Prabhanjana
    Smith, Brady
    Cherone, Rocco
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [6] Performance evaluation of a C-Arm CT perfusion phantom
    Boese, Axel
    Gugel, Sebastian
    Serowy, Steffen
    Purmann, Jonas
    Rose, Georg
    Beuing, Oliver
    Skalej, Martin
    Kyriakou, Yiannis
    Yu Deuerling-Zheng
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2013, 8 (05) : 799 - 807
  • [7] Performance evaluation of a C-Arm CT perfusion phantom
    Axel Boese
    Sebastian Gugel
    Steffen Serowy
    Jonas Purmann
    Georg Rose
    Oliver Beuing
    Martin Skalej
    Yiannis Kyriakou
    Yu Deuerling-Zheng
    International Journal of Computer Assisted Radiology and Surgery, 2013, 8 : 799 - 807
  • [8] Diagnostic Accuracy of CT Perfusion Imaging in the Evaluation of Acute Cerebral Ischemia
    Patel, S.
    Pappas, G.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (05)
  • [9] Closed Sinusoid Trajectory for C-Arm CT Imaging
    Yang, Haiquan
    Li, Meihua
    Koizumi, Kazuhito
    Kudo, Hiroyuki
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3480 - 3484
  • [10] Circular tomosynthesis for neuro perfusion imaging on an interventional C-arm
    Claus, Bernhard E.
    Langan, David A.
    Al Assad, Omar
    Wang, Xin
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412