Evaluation of the automated coagulation analyzer SYSMEX CA 6000

被引:17
|
作者
Quehenberger, P [1 ]
Kapiotis, S [1 ]
Handler, S [1 ]
Ruzicka, K [1 ]
Speiser, W [1 ]
机构
[1] Univ Vienna, Clin Inst Med & Chem Lab Diagnost, A-1090 Vienna, Austria
关键词
evaluation; coagulation analyzer; SYSMEX CA 6000;
D O I
10.1016/S0049-3848(99)00069-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In the present study the coagulation analyzer SYSMEX CA 6000 (TOA Medical Electronics Co., Kobe, Japan), an analyzer equipped with a photo-optical clot detection unit and a cap-piercing system, was evaluated with respect to its technical characteristics in the determination of standard coagulation tests (prothrombin time, activated partial thromboplastin time, thrombin time, fibrinogen, and antithrombin) and in the determination of coagulation single factor activities. In the normal and in the pathological range the intraassay coefficients of variation (CV) and interassay CV for most parameters were below 5% (exceptions: intraassay CV 5.4% for prolonged thrombin time; intraassay CV 9.26% and interassay CV 10.7% for decreased antithrombin; interassay CV 5.62% for fibrinogen in the normal range, intraassay CV 10.1% for fibrinogen greater than 7.0 g/L; intraassay CV 6.36% and interassay CV 11.7% for decreased fibrinogen; interassay CV 11.6% for prolonged activated partial thromboplastin time; interassay CV 6.12% for decreased factor VII). Interference studies with lipemic, icteric, and hemolytic samples showed just minor influences of these abnormal sample characteristics on prothrombin time, activated partial thromboplastin time, fibrinogen, and antithrombin measurements when compared to the results obtained by using mechanical clot detection (STA, Stage Diagnostica, Asnieres-Sur-Seine, France). No carryover was detected in alternating measurements of heparinized (3 U/mL unfractionated heparin) and normal plasma samples. Measurement of the activities of clotting factors V, VII, VIII, and IX showed a good correlation (r=0.993 to r=0.977) between SYSMEX CA 6000 and STA. Our results demonstrate that using SYSMEX CA 6000 analyzer basal routine coagulation testing as well as specialized tests for single factor activities can be performed with satisfactory precision; in particular, the cap-piercing system has no negative effect on the performance of the analyzer. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:65 / 71
页数:7
相关论文
共 50 条
  • [1] Summary evaluation of the Sysmex CA-6000 coagulation analyzer
    Gosselin, RC
    Owings, JT
    Larkin, E
    Cannon, J
    CLINICAL CHEMISTRY, 1997, 43 (09) : 1783 - 1785
  • [2] Evaluation of the automated coagulation analyzer Sysmex® CA-7000
    Fischer, F
    Appert-Flory, A
    Jambou, D
    Toulon, P
    THROMBOSIS RESEARCH, 2006, 117 (06) : 721 - 729
  • [3] Evaluation of the automated coagulation analyzer Sysmex® CA-7000
    Dorn-Beineke, A
    Dempfle, CE
    Bertsch, T
    Wisser, H
    THROMBOSIS RESEARCH, 2005, 116 (02) : 171 - 179
  • [4] Evaluation of the Sysmex CN-6000 Coagulation Analyzer for Routine and Specialized Coagulation Testing in a Central Laboratory
    Kocijancic, Marija
    Roche, Janina
    Peter, Andreas
    Hoerber, Sebastian
    CLINICAL LABORATORY, 2025, 71 (03) : 493 - 501
  • [5] Performance Evaluation of the Sysmex CS-5100 Automated Coagulation Analyzer
    Chen, Liming
    Chen, Yu
    CLINICAL LABORATORY, 2015, 61 (07) : 653 - 660
  • [6] Performance evaluation of the Sysmex CN-6000-automated coagulation analyzers for routine coagulation tests
    Kim, Hyerim
    Jo, Su-Yeon
    Kim, In-Suk
    BLOOD COAGULATION & FIBRINOLYSIS, 2023, 34 (02) : 93 - 98
  • [7] Clinical evaluation of the Sysmex® CA-7000 coagulation analyzer.
    Engelhardt, W
    Gringel, E
    Hafner, G
    Kretschmer, V
    Lichte, A
    Moessmer, G
    Sinha, P
    Ziemer, S
    CLINICAL CHEMISTRY, 2002, 48 (06) : A156 - A156
  • [8] Evaluation of the automated coagulation analyzer Sysmex® CS-2100i (Siemens)
    Martin-Toutain, Isabelle
    Jobic, Lionel
    Mancic, Tessy
    Ankri, Annick
    ANNALES DE BIOLOGIE CLINIQUE, 2011, 69 (06) : 699 - 704
  • [9] VALIDATION OF THE SYSMEX CN6000 COAGULATION ANALYZER FOR COAGULATION AND PLATELET AGGREGATION TESTS
    Ang, Mui Kia
    Sum, Christina Lai Lin
    Tan, Hwee Tat
    Lim, Shu Ping
    Fan, Bingwen Eugene
    Ong, Kiat Hoe
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 : 9 - 9
  • [10] Evaluation of Sysmex Fully Automated CS Series Coagulation Analyzer at a Tertiary Care Hospital
    Gupta, Monali
    Melinkeri, Sameer Ramesh
    Prabhakaran, Anusree
    Choksey, Urmi
    BLOOD, 2017, 130