Assessment of an ultrasonic dermal scanner for skin thickness measurements

被引:10
|
作者
Kong, Linghua [1 ]
Caspall, Jayme [1 ]
Duckworth, Mark [1 ]
Spriale, Stephen [1 ]
机构
[1] Georgia Inst Technol, Coll Architecture, Ctr Assist Technol & Environm Access, Atlanta, GA 30332 USA
关键词
ultrasound technical assessment; dermal scanning; skin thickness;
D O I
10.1016/j.medengphy.2007.10.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The measurement precision of a 20 MHz ultrasound dermal scanning system, Episcano(R), for its use in the measurement of skin thickness. Results from measurements of relatively ideal homogeneous, uniformly thick plastic plates indicate that the scanner can accurately measure to depths of 8-34 inm from the surface of the transducer or approximately 18 mm from the surface of the latex probe-cover with a high level of precision, less than 1% of the mean thickness. The precision was determined to be dependent on the depth of the scan. Using the measured optimal gain of 45%, a sample rate of 0.0 13 mu s, and an A-scan record length of 1024 points, we determined the thickness resolution of the device is on the order of 0.003 mm for 2 mm thick layers and 0.01 mm for 5 mm thick layers. We conclude that variations greater than this value for clinical dermal thickness measurements are not due to instrument precision, but must result from the limitations of analyzing the data from real tissue. (C) 2007 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:804 / 807
页数:4
相关论文
共 50 条
  • [1] Assessment of Uncertainty in Modelled Ultrasonic Thickness Measurements
    Markucic, Damir
    Runje, Biserka
    Stepanic, Josip
    MATERIALS TESTING, 2011, 53 (04) : 233 - 238
  • [2] ULTRASONIC THICKNESS MEASUREMENTS
    NEWMAN, DR
    RYDEN, J
    LAMB, LT
    INSTRUMENTS & CONTROL SYSTEMS, 1970, 43 (12): : 71 - &
  • [3] Handheld High-Resolution Ultrasonic Scanner for Quantitative Assessment of Skin Conditions
    Seviaryn, Fedar
    Malyarenko, Eugene
    Schreiner, Gregory
    Seviaryna, Ina
    Maev, Roman Gr.
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 2380 - 2382
  • [4] An Ultrasonic Micro-Scanner for Thickness Assessment of the Vestibular Jawbone: In-Vitro Accuracy Evaluation
    Habor, Daniel
    Vollborn, Thorsten
    Chuembou, Fabrice
    Radermacher, Klaus
    Degen, Katharina
    Marotti, Juliana
    Wolfart, Stefan
    Heger, Stefan
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 745 - 748
  • [5] THE ULTRASONIC EVALUATION OF SKIN THICKNESS IN SCLEREDEMA
    COLE, GW
    HANDLER, SJ
    BURNETT, K
    JOURNAL OF CLINICAL ULTRASOUND, 1981, 9 (09) : 501 - 503
  • [6] ULTRASONIC MEASUREMENTS OF CHEST WALL THICKNESS
    RAMSDEN, D
    HEALTH PHYSICS, 1977, 33 (04): : 346 - 346
  • [7] ULTRASONIC MEASUREMENTS OF CHEST WALL THICKNESS
    GARG, SP
    HEALTH PHYSICS, 1977, 32 (01): : 34 - +
  • [8] ULTRASONIC MEASUREMENTS OF SUBCUTANEOUS FAT THICKNESS
    RAMIREZ, ME
    AMERICAN JOURNAL OF PHYSICAL ANTHROPOLOGY, 1990, 81 (02) : 283 - 284
  • [9] Ultrasonic instrumentation techniques for thickness measurements
    Chauhan, S.K.
    Singh, V.R.
    IETE Technical Review (Institution of Electronics and Telecommunication Engineers, India), 1992, 9 (03):
  • [10] ULTRASONIC THICKNESS MEASUREMENTS - AN EQUIPMENT APPRAISAL
    BOBBIN, J
    MATERIALS EVALUATION, 1972, 30 (02) : A23 - &