Incident angle-dependent gain correction for reflection images on ultrasound computed tomography

被引:0
|
作者
Tsubota, Yushi [1 ]
Terada, Takahide [1 ]
Suzuki, Atsuro [1 ]
Wu, Wenjing [1 ]
Yamanaka, Kazuhiro [1 ]
Yamashita, Hiroko [2 ]
Kato, Fumi [2 ]
Nishida, Mutsumi [2 ]
Satoh, Megumi [2 ]
Kawabata, Ken-ichi [1 ]
机构
[1] Hitachi Ltd, Res & Dev Grp, 1-280 Higashi Koigakubo, Kokubunji, Tokyo 1858601, Japan
[2] Hokkaido Univ Hosp, Kita Ku, Kita 14,Nishi 5, Sapporo, Hokkaido 0608648, Japan
关键词
Ultrasound computed tomography; reflection image; gain correction;
D O I
10.1117/12.2548555
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We are performing clinical studies on breast cancer examinations at Hokkaido University Hospital with an ultrasound computed tomography (USCT) system. Our studies have revealed that some reflection images exhibit intensity inhomogeneity because ultrasound waves, shot by a 1-D ring array transducer, go non-vertically into the object surface. This trend significantly increases the burden of interpretation. Therefore, we developed a calibration method to remove this heterogeneity based on the distribution of the incident angle of waves that are estimated from the slope of the subject surface morphologically extracted from multi-slice reflection images. Results showed that applying this correction method to clinical images enabled the image contrast and uniformity to be successfully recovered.