Interference of the Hope Hemoglobin With Hemoglobin A1c Results
被引:9
|作者:
Chakraborty, Sutirtha
[1
,2
]
Chanda, Dalia
[3
]
Gain, Mithun
[1
,2
]
Krishnan, Prasad
[4
]
机构:
[1] Peerless Hosp, Dept Biochem, Kolkata, India
[2] BK Roy Res Ctr, Dept Biochem, Kolkata, India
[3] BP Poddar Hosp & Med Res Ltd, Dept Biochem, Kolkata, India
[4] Natl Neurosci Ctr, Dept Neurosurg, Kolkata, India
来源:
关键词:
HbA1c;
hemoglobin Hope;
interference;
capillary electrophoresis;
HPLC;
immunoassay;
ELECTROPHORESIS;
DIAGNOSIS;
D O I:
10.1309/LME82XNY6SYVWDYQ
中图分类号:
R446 [实验室诊断];
R-33 [实验医学、医学实验];
学科分类号:
1001 ;
摘要:
Hemoglobin A1c (HbA1c) is now considered to be the marker of choice in diagnosis and management of diabetes mellitus, based on the results of certain landmark clinical trials. Herein, we report the case of a 52-year-old ethnic Southeast Asian Indian man with impaired glucose tolerance whose glycated hemoglobin (ie, HbA1c) levels, as measured via Bio-Rad D10 high-performance liquid chromatography (HPLC) and Roche Tina-quant immunoassay were 47.8% and 44.0%, respectively. No variant hemoglobin (Hb) peak was observed via the 010 chromatogram. We assayed the patient specimen on the Sebia MINICAP capillary electrophoresis platform; the HbA1c level was 6.8%, with a large variant Hb peak of 42.0%. This finding suggested the possible presence of the heterozygous Hb Hope, which can result in spuriously elevated HbA1c results on HPLC and turbidimetric immunoassays. Although the capillary electrophoresis system was able to identify the variant, the A1c results should not be considered accurate due to overlapping of the variant and adult Hb peaks on the electrophoretogram reading. Hb Hope is usually clinically silent but can present such analytical challenges. Through this case study, we critically discuss the limitations of various HbA1c assay methods, highlighting the fact that laboratory professionals need to be aware of occurrences of Hb Hope, to help ensure patient safety.
引用
收藏
页码:221 / 225
页数:5
相关论文