Readout of plastic scintillators with cooled large-area avalanche photodiodes

被引:1
|
作者
Mykulyak, A
Kapusta, M
Lynen, U
Moszynski, M
Müller, WFJ
Orth, H
Schwarz, C
Szawlowski, M
Trautmann, W
Trzcinski, A
Wolski, D
Zwieglinski, B
机构
[1] Andrzej Soltan Inst Nucl Studies, PL-00681 Warsaw, Poland
[2] Gesell Schwerionenforsch mbH, D-64291 Darmstadt, Germany
[3] Adv Photon Inc, Camarillo, CA 93012 USA
关键词
multifragmentation of heavy relativistic projectiles; fragment time-of-flight measurements with BC-408 plastic scintillators; amplitude and time response of a BC-408+LAAPD combination; LAAPD gain increase by its cooling; limiting time resolution;
D O I
10.1016/j.nima.2004.01.003
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Time-of-flight measurements in multifragmentation of heavy, relativistic projectiles require very good time resolution and at the same time the detecting system must cope with a dynamic range of up to 8000: 1. We look into the possibility of application of large-area avalanche photodiodes (LAAPDs) as alternative light sensors to meet the above requirements. The paper presents the results of our amplitude and time response studies of a plastic scintillator BC-408 readout with a phi16 mm LAAPD using radioactive sources. The measurements were performed using two different setups. The best time resolution has been obtained by exploiting LAAPD cooling to increase its gain beyond that accessible at room temperature. We reach 610 ps (FWHM) at -26degreesC for the corresponding LAAPD gain of approximate to900 and the Sr-90/Y-90 beta-ray source. We hope to reach below the desired 400 ps with the latter setup at the higher light levels available with multifragmentation products. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:425 / 434
页数:10
相关论文
共 50 条
  • [21] SIMPLE SCHEME FOR LIGHT COLLECTION FROM LARGE-AREA PLASTIC SCINTILLATORS
    FIELDS, T
    JANKOWSKI, D
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 215 (1-2): : 131 - 133
  • [22] Energy resolution of scintillation detectors with large area avalanche photodiodes and photomultipliers light readout
    Moszynski, M
    Kapusta, M
    Wolski, D
    Szawlowski, M
    Klamra, W
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 206 - 211
  • [23] X-ray spectrometry with Peltier-cooled large area avalanche photodiodes
    Fernandes, LMP
    Lopes, JAM
    dos Santos, JMF
    Conde, CAN
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 213 : 267 - 271
  • [24] Beta spectroscopy with a large-area avalanche photodiode module and a plastic scintillator
    Kriss, AA
    Hamby, DM
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (03): : 553 - 559
  • [25] Can large-area avalanche photodiodes be used for a clinical PET/MRI block detector?
    Peng, Hao
    Olcott, P. D.
    Spanoudaki, Virginia
    Levin, Craig S.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3948 - 3953
  • [26] Application of large-area avalanche photodiodes to X-ray spectrometry of muonic atoms
    Fernandes, LMP
    Antognini, A
    Boucher, M
    Conde, CAN
    Huot, O
    Knowles, PE
    Kottmann, F
    Ludhova, L
    Mulhauser, F
    Pohl, R
    dos Santos, JMF
    Schaller, LA
    Taqqu, D
    Veloso, JFCA
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2003, 58 (12) : 2255 - 2260
  • [27] AN AMPLIFIER FOR USE WITH LARGE-AREA PHOTODIODES
    DYER, GR
    UCKAN, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (12): : 3909 - 3910
  • [28] Xenon GPSC high-pressure operation with large-area avalanche photodiode readout
    Coelho, L. C. C.
    Lopes, J. A. M.
    Covita, D. S.
    Conceicao, A. S.
    dos Santos, J. M. F.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 575 (03): : 444 - 448
  • [29] RECENT ADVANCES IN LARGE AREA AVALANCHE PHOTODIODES
    SQUILLANTE, MR
    REIFF, G
    ENTINE, G
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) : 563 - 566
  • [30] Highly segmented large-area hybrid photodiodes with bialkali photocathodes and enclosed VLSI readout electronics
    Braem, A
    Chesi, E
    Dulinski, W
    Filthaut, F
    Go, A
    Joram, C
    Séguinot, J
    Weilhammer, P
    Wenzel, H
    Wicht, P
    Ypsilantis, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 442 (1-3): : 128 - 135