共 50 条
Determination of anti-oxidative histidine dipeptides in poultry by microchip capillary electrophoresis with contactless conductivity detection
被引:21
|作者:
Jozanovic, Marija
[1
]
Hajdukovic, Mateja
[1
]
Galovic, Olivera
[1
,2
]
Kralik, Gordana
[2
,3
]
Kralik, Zlata
[2
,3
]
Sakac, Nikola
[1
]
Medvidovic-Kosanovic, Martina
[1
]
Sak-Bosnar, Milan
[1
,2
]
机构:
[1] Josip Juraj Strossmayer Univ Osijek, Dept Chem, Cara Hadrijana 8A, HR-31000 Osijek, Croatia
[2] Josip Juraj Strossmayer Univ Osijek, Ctr Excellence, HR-31000 Osijek, Croatia
[3] Josip Juraj Strossmayer Univ Osijek, Fac Agr, HR-31000 Osijek, Croatia
来源:
关键词:
Anserine;
Carnosine;
(CD)-D-4 detector;
Histidine dipeptides;
Microchip electrophoresis;
Poultry meat;
CARNOSINE-RELATED PEPTIDES;
EXCHANGE CHROMATOGRAPHY;
LIQUID-CHROMATOGRAPHY;
MASS-SPECTROMETRY;
ORIGIN PROTEINS;
ANIMAL ORIGIN;
PERFORMANCE;
ANSERINE;
MEAT;
SUPPLEMENTATION;
D O I:
10.1016/j.foodchem.2016.10.122
中图分类号:
O69 [应用化学];
学科分类号:
081704 ;
摘要:
A home-made microchip electrophoresis (MCE) device was used to quantitate two biologically important histidine dipeptides, carnosine and anserine, using capacitively coupled contactless conductivity detection ((CD)-D-4), at pH 2.7. The (CD)-D-4 detector exhibited a linear response to both carnosine and anserine in the range of 0-200 mu M for the individual dipeptides and in the range of 0-100 mu M for each dipeptide when both were present as a mixture. The limit of detections (LOD) for the dipeptides in the mixture were 0.10 mu M for carnosine and 0.16 mu M for anserine. Standard addition was used to detemine the accuracy of the method. For carnosine and anserine the recoveries were in the range of 96.7 +/- 4.9-106.0 +/- 7.5% and 95.3 +/- 4.5-105.0 +/- 5.1% in thigh muscle and 97.5 +/- 5.1-105.0 +/- 7.5% and 95.3 +/- 5.4-97.3 +/- 5.6% in breast muscle, respectively. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1658 / 1665
页数:8
相关论文