Burden of atrial fibrillation and poor rate control detected by continuous monitoring and the risk for heart failure hospitalization

被引:45
|
作者
Sarkar, Shantanu [1 ]
Koehler, Jodi [1 ]
Crossley, George H. [2 ,3 ]
Tang, W. H. Wilson [4 ]
Abraham, William T. [5 ]
Warman, Eduardo N. [1 ]
Whellan, David J. [6 ]
机构
[1] Medtronic Inc, Mounds View, MN USA
[2] St Thomas Res Inst, Nashville, TN USA
[3] Univ Tennessee, Coll Med, Nashville, TN USA
[4] Cleveland Clin, Cleveland, OH 44106 USA
[5] Ohio State Univ, Columbus, OH 43210 USA
[6] Thomas Jefferson Univ, Philadelphia, PA 19107 USA
关键词
VENTRICULAR SYSTOLIC DYSFUNCTION; QUALITY-OF-CARE; MANAGEMENT; TRIAL; MORTALITY; PROGRAM; DRONEDARONE; ABLATION; NETWORK; STROKE;
D O I
10.1016/j.ahj.2012.06.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Atrial fibrillation (AF) on electrocardiogram has been identified as a risk factor for hospitalizations in patients with heart failure (HF). We investigated whether continuous AF monitoring can identify when patients with HF are at risk for hospitalization. Methods In this retrospective analysis of data from 4 studies enrolling patients with HF with cardiac resynchronization therapy defibrillator devices with >= 90 days of follow-up (n = 1561), patients were identified as having AF if they had >= 1 day of >5 minutes of AF and >1 hour of total AF during entire follow-up. In patients with AF, device recorded AF burden (AFb) and ventricular rate during AF (VRAF) over the last 30 days was classified on a monthly basis into 3 evaluation groups: (1) >= 1 day of high burden of paroxysmal AF (>= 6 hours) or persistent AF (all 30 days with AFb >23 hours) with poor rate control (VRAF >90 beats/min), (2) >= 1 day of high burden of paroxysmal AF with good rate control (VRAF <= 90 beats/min), and (3) no days with high burden of AF (AFb <6 hours) or persistent AF with good rate control. Each group was compared with monthly evaluations in patients without AF using an Anderson-Gill model for occurrence of HF hospitalizations in the next 30 days. Results Patients with AF (n = 519, 33%) have a greater risk (hazard ratio [HR] 2.0, P < .001) for impending HF hospitalizations during entire follow-up compared with patients with no AF. One day of high burden of paroxysmal AF with good rate control in the last 30 days increases risk for HF hospitalization in the next 30 days (HR 3.4, P < .001). The risk increases further (HR 5.9, P < .001) with 1 day of poor rate control during persistent AF or high burden paroxysmal AF in last 30 days. Conclusion Evaluation of AFb and rate control information on a monthly basis can identify patients at risk for HF hospitalization in the next 30 days. (Am Heart J 2012;164:616-24.)
引用
收藏
页码:616 / 624
页数:9
相关论文
共 50 条
  • [1] BURDEN OF ATRIAL FIBRILLATION AND POOR RATE CONTROL DETECTED BY CONTINUOUS MONITORING VIA IMPLANTED DEVICES IDENTIFIES WHEN A PATIENT IS AT RISK FOR HEART FAILURE HOSPITALIZATION
    Sarkar, Shantanu
    Koehler, Jodi
    Crossley, George H.
    Tang, Wai Hong Wilson
    Abraham, William T.
    Whellan, David J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (14) : E107 - E107
  • [2] Device Detected Atrial Fibrillation Poses Significant Risk of Heart Failure Related Hospitalization
    Crossley, George H.
    Boyle, Andrew
    Mead, R. Hardwin
    Bergeman, Tracy
    CIRCULATION, 2014, 130
  • [3] Continuous Monitoring of Atrial Fibrillation in Heart Failure
    Herweg, Bengt
    Barold, S. Serge
    Steinberg, Jonathan S.
    HEART FAILURE CLINICS, 2013, 9 (04) : 385 - +
  • [4] Device-detected silent atrial fibrillation is associated to a high hospitalization rate for heart failure
    El Bouazzaoui, R.
    Davy, J. M.
    Massin, F.
    Cung, T. T.
    Khoueiry, Z.
    Laugaudin, G.
    Zhao, X.
    Thomann, S.
    Pasquie, J. -L.
    EUROPEAN HEART JOURNAL, 2016, 37 : 304 - 304
  • [5] Atrial Fibrillation (AF) detected by repeat handheld ECG screening and the risk of heart failure hospitalization
    Sun, W.
    Yan, B.
    EUROPEAN HEART JOURNAL, 2024, 45
  • [6] Atrial fibrillation and heart failure in continuous distance monitoring
    Santini, Luca
    Papavasileiou, Lida Pieretta
    Santini, Massimo
    GIORNALE ITALIANO DI CARDIOLOGIA, 2013, 14 (04) : 253 - 261
  • [7] Continuous heart rate monitoring in atrial fibrillation with multispectral photoplethysmography
    Devaki, V.
    Lakshmanan, Madan Kumar
    Aravindan, R.
    Dhandapani, V. E.
    Jayanthi, T.
    SENSORS AND ACTUATORS A-PHYSICAL, 2024, 380
  • [8] Ventricular Rate Control of Atrial Fibrillation in Heart Failure
    Rienstra, Michiel
    Van Gelder, Isabelle C.
    HEART FAILURE CLINICS, 2013, 9 (04) : 397 - +
  • [9] Post-operative atrial fibrillation and risk of heart failure hospitalization
    Goyal, Parag
    Kim, Michael
    Krishnan, Udhay
    Mccullough, Stephen A.
    Cheung, Jim W.
    Kim, Luke K.
    Pandey, Ambarish
    Borlaug, Barry A.
    Horn, Evelyn M.
    Safford, Monika M.
    Kamel, Hooman
    EUROPEAN HEART JOURNAL, 2022, 43 (31) : 2971 - 2980
  • [10] Mediterranean diet and risk of first hospitalization for heart failure and atrial fibrillation
    Bonaccio, M.
    Di Castelnuovo, A.
    Costanzo, S.
    Ruggiero, E.
    Persichillo, M.
    De Curtis, A.
    Cerletti, C.
    Donati, M. B.
    de Gaetano, G.
    Iacoviello, Licia
    EUROPEAN JOURNAL OF PUBLIC HEALTH, 2019, 29