Technique, radiation safety and image quality for chest X-ray imaging through glass and in mobile settings during the COVID-19 pandemic

被引:27
|
作者
Brady, Zoe [1 ,2 ]
Scoullar, Heather [1 ]
Grinsted, Ben [1 ]
Ewert, Kyle [1 ]
Kavnoudias, Helen [1 ,2 ,3 ]
Jarema, Alexander [1 ]
Crocker, James [1 ]
Wills, Rob [1 ]
Houston, Gillian [1 ]
Law, Meng [1 ,2 ,4 ]
Varma, Dinesh [1 ,3 ]
机构
[1] Alfred Hlth, Dept Radiol, Melbourne, Vic, Australia
[2] Monash Univ, Dept Neurosci, Melbourne, Vic, Australia
[3] Monash Univ, Dept Surg, Melbourne, Vic, Australia
[4] Monash Univ, Dept Elect & Comp Syst Engn, Melbourne, Vic, Australia
关键词
Chest imaging; X-ray; Pandemic; Radiation safety; Portable; COVID-19;
D O I
10.1007/s13246-020-00899-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The COVID-19 pandemic in 2020 has led to preparations within our hospital for an expected surge of patients. This included developing a technique to perform mobile chest X-ray imaging through glass, allowing the X-ray unit to remain outside of the patient's room, effectively reducing the cleaning time associated with disinfecting equipment. The technique also reduced the infection risk of radiographers. We assessed the attenuation of different types of glass in the hospital and the technique parameters required to account for the glass filtration and additional source to image distance (SID). Radiation measurements were undertaken in a simulated set-up to determine the appropriate position for staff inside and outside the room to ensure occupational doses were kept as low as reasonably achievable. Image quality was scored and technical parameter information collated. The alternative to imaging through glass is the standard portable chest X-ray within the room. The radiation safety requirements for this standard technique were also assessed. Image quality was found to be acceptable or borderline in 90% of the images taken through glass and the average patient dose was 0.02 millisieverts (mSv) per image. The majority (67%) of images were acquired at 110 kV, with an average 5.5 mAs and with SID ranging from 180 to 300 cm. With staff positioned at greater than 1 m from the patient and at more than 1 m laterally from the tube head outside the room to minimise scatter exposure, air kerma values did not exceed 0.5 microgray (mu Gy) per image. This method has been implemented successfully.
引用
收藏
页码:765 / 779
页数:15
相关论文
共 50 条
  • [11] COVID-19 pandemic and the effect of increased utilisation of mobile X-ray examinations on radiation dose to radiographers
    Yeung, Phoebe
    Pinson, Jo-Anne
    Lawson, Michael
    Leong, Christopher
    Badawy, Mohamed Khaldoun
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2022, 69 (02) : 147 - 155
  • [12] Evaluation of Admission Chest X-Ray Findings in Patients With Respiratory Infection During the COVID-19 Pandemic
    Velissaris, Dimitrios
    Lagadinou, Maria
    Paraskevas, Themistoklis
    Oikonomou, Eleousa
    Karamouzos, Vasileios
    Sampsonas, Fotios
    Marangos, Markos
    Karanikolas, Menelaos
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (09)
  • [13] Modifications to mobile chest radiography technique during the COVID-19 pandemic - implications of X-raying through side room windows
    England, A.
    Littler, E.
    Romani, S.
    Cosson, P.
    RADIOGRAPHY, 2021, 27 (01) : 193 - 199
  • [14] Robust Technique to Detect COVID-19 using Chest X-ray Images
    Channa, Asma
    Popescu, Nirvana
    Malik, Najeeb Ur Rehman
    2020 INTERNATIONAL CONFERENCE ON E-HEALTH AND BIOENGINEERING (EHB), 2020,
  • [15] COVID-19 chest X-ray image classification in the presence of noisy labels*
    Ying, Xiaoqing
    Liu, Hao
    Huang, Rong
    DISPLAYS, 2023, 77
  • [16] Comparison of Chest Ultrasound and Standard X-Ray Imaging in COVID-19 Patients
    Shumilov, Evgenii
    Hosseini, Ali Seif Amir
    Petzold, Golo
    Treiber, Hannes
    Lotz, Joachim
    Ellenrieder, Volker
    Kunsch, Steffen
    Neesse, Albrecht
    ULTRASOUND INTERNATIONAL OPEN, 2020, 06 (02) : E36 - E40
  • [17] Pediatric chest x-ray in covid-19 infection
    Serrano, C. Oterino
    Alonso, E.
    Andres, M.
    Buitrago, N. M.
    Perez Vigara, A.
    Parron Pajares, M.
    Cuesta Lopez, E.
    Garzon Moll, G.
    Martin Espin, I
    Bueno Barriocanal, M.
    De Ceano-Vivas la Calle, M.
    Calvo Rey, C.
    Bret-Zurita, M.
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 131
  • [18] A dataset of COVID-19 x-ray chest images
    Fraiwan, Mohammad
    Khasawneh, Natheer
    Khassawneh, Basheer
    Ibnian, Ali
    DATA IN BRIEF, 2023, 47
  • [19] Implementing a Novel Through-Glass Chest Radiography Technique for COVID-19 Patients: Image Quality, Radiation Dose Optimization, and Practical Considerations
    Chan, Jessica
    Auffermann, William
    Jenkins, Peter
    Streitmatter, Seth
    Phuong-Anh Duong
    CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2022, 51 (01) : 38 - 45
  • [20] COVID-19 chest X-ray image analysis by threshold-based segmentation
    Al-Zyoud, Walid
    Erekat, Dana
    Saraiji, Rama
    HELIYON, 2023, 9 (03)