Critical evaluation of connectivity-based point of care testing systems of glucose in a hospital environment

被引:11
|
作者
Flore, Katelijne M. J. [1 ]
Fiers, Tom [1 ]
Delanghe, Joris R. [1 ]
机构
[1] Ghent Univ Hosp, Clin Chem Lab, B-9000 Ghent, Belgium
关键词
critically ill patients; hyperglycemia; point of care testing (POCT); quality control; tight glucose monitoring;
D O I
10.1515/CCLM.2008.332
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: In recent years a number of point of care testing (POCT) glucometers were introduced on the market. We investigated the analytical variability (lot-to-lot variation, calibration error, inter-instrument and inter-operator variability) of glucose POCT systems in a university hospital environment and compared these results with the analytical needs required for tight glucose monitoring. Methods: The reference hexokinase method was compared to different POCT systems based on glucose oxidase (blood gas instruments) or glucose dehydrogenase (handheld glucometers). Based upon daily internal quality control data, total errors were calculated for the various glucose methods and the analytical variability of the glucometers was estimated. Results: The total error of the glucometers exceeded by far the desirable analytical specifications (based on a biological variability model). Lot-to-lot variation, inter-instrument variation and inter-operator variability contributed approximately equally to total variance. As in a hospital environment, distribution of hematocrit values is broad, converting blood glucose into plasma values using a fixed factor further increases variance. The percentage of outliers exceeded the ISO 15197 criteria in a broad glucose concentration range. Conclusions: Total analytical variation of handheld glucometers is larger than expected. Clinicians should be aware that the variability of glucose measurements obtained by blood gas instruments is lower than results obtained with handheld glucometers on capillary blood. Clin Chem Lab Med 2008; 46: 1763-8.
引用
收藏
页码:1763 / 1768
页数:6
相关论文
共 50 条
  • [31] Connectivity - The millennium challenge for point-of-care testing
    Kost, GJ
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2000, 124 (08) : 1108 - 1110
  • [32] Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care hospital ambulatory patients in the United States and Canada
    Kost, GJ
    Vu, HT
    Lee, JH
    Bourgeois, P
    Kiechle, FL
    Martin, C
    Miller, SS
    Okorodudu, AO
    Podczasy, JJ
    Webster, R
    Whitlow, KJ
    CRITICAL CARE MEDICINE, 1998, 26 (03) : 581 - 590
  • [33] Point-of-care glucose testing - Effects of critical care variables, influence of reference instruments, and a modular glucose meter design
    Louie, RF
    Tang, ZP
    Sutton, DV
    Lee, JH
    Kost, GJ
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2000, 124 (02) : 257 - 266
  • [34] Drug interference with glucose measurements on six handheld glucose meters and a portable glucose analyzer for point-of-care testing in critical care.
    Tang, ZP
    Du, XG
    Louie, RF
    Kost, GJ
    CLINICAL CHEMISTRY, 1999, 45 (06) : A31 - A32
  • [35] Point of Care Glucose Testing in Neonatal Hypoglycemia
    Movalia, Mayur K.
    Ogino, Mark T.
    POINT OF CARE, 2006, 5 (03): : 95 - 99
  • [36] Blood glucose determination: Point of care testing
    Bennett, BD
    SOUTHERN MEDICAL JOURNAL, 1997, 90 (07) : 678 - 680
  • [37] Special Editions on Point of Care Glucose Testing
    Lewandrowski, Kent
    POINT OF CARE, 2006, 5 (02): : 47 - 48
  • [38] Microneedle-Based Glucose Sensor Platform: From Vitro to Wearable Point-of-Care Testing Systems
    Ju, Jian
    Li, Lin
    Regmi, Sagar
    Zhang, Xinyu
    Tang, Shixing
    Biosensors, 2022, 12 (08):
  • [39] Microneedle-Based Glucose Sensor Platform: From Vitro to Wearable Point-of-Care Testing Systems
    Ju, Jian
    Li, Lin
    Regmi, Sagar
    Zhang, Xinyu
    Tang, Shixing
    BIOSENSORS-BASEL, 2022, 12 (08):
  • [40] Multicenter survey for blood contamination on hospital point-of-care glucose testing meters.
    Louie, RF
    Lau, MJ
    Tran, NK
    Tang, Z
    Lee, JH
    Kost, GJ
    CLINICAL CHEMISTRY, 2003, 49 (06) : A152 - A153