Critical evaluation of connectivity-based point of care testing systems of glucose in a hospital environment

被引:11
|
作者
Flore, Katelijne M. J. [1 ]
Fiers, Tom [1 ]
Delanghe, Joris R. [1 ]
机构
[1] Ghent Univ Hosp, Clin Chem Lab, B-9000 Ghent, Belgium
关键词
critically ill patients; hyperglycemia; point of care testing (POCT); quality control; tight glucose monitoring;
D O I
10.1515/CCLM.2008.332
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: In recent years a number of point of care testing (POCT) glucometers were introduced on the market. We investigated the analytical variability (lot-to-lot variation, calibration error, inter-instrument and inter-operator variability) of glucose POCT systems in a university hospital environment and compared these results with the analytical needs required for tight glucose monitoring. Methods: The reference hexokinase method was compared to different POCT systems based on glucose oxidase (blood gas instruments) or glucose dehydrogenase (handheld glucometers). Based upon daily internal quality control data, total errors were calculated for the various glucose methods and the analytical variability of the glucometers was estimated. Results: The total error of the glucometers exceeded by far the desirable analytical specifications (based on a biological variability model). Lot-to-lot variation, inter-instrument variation and inter-operator variability contributed approximately equally to total variance. As in a hospital environment, distribution of hematocrit values is broad, converting blood glucose into plasma values using a fixed factor further increases variance. The percentage of outliers exceeded the ISO 15197 criteria in a broad glucose concentration range. Conclusions: Total analytical variation of handheld glucometers is larger than expected. Clinicians should be aware that the variability of glucose measurements obtained by blood gas instruments is lower than results obtained with handheld glucometers on capillary blood. Clin Chem Lab Med 2008; 46: 1763-8.
引用
收藏
页码:1763 / 1768
页数:6
相关论文
共 50 条
  • [1] Critical Care Glucose Point-of-Care Testing
    Narla, S. N.
    Jones, M.
    Hermayer, K. L.
    Zhu, Y.
    ADVANCES IN CLINICAL CHEMISTRY, VOL 76, 2016, 76 : 97 - 121
  • [2] One Touch II Hospital Data Dock connectivity for glucose testing at the point-of-care.
    Hampton, B
    CLINICAL CHEMISTRY, 2000, 46 (06) : A8 - A8
  • [3] Remote connectivity for point-of-care glucose testing.
    Weber, S
    CLINICAL CHEMISTRY, 1998, 44 : A67 - A67
  • [4] Remote network (Ethernet) connectivity for glucose testing at the point-of-care.
    Weber, S
    CLINICAL CHEMISTRY, 1999, 45 (06) : A35 - A35
  • [5] Outcomes and benefits of connectivity for point-of-care glucose testing.
    Bauwens, L
    CLINICAL CHEMISTRY, 2001, 47 (06) : A178 - A178
  • [6] Point of care testing informatics - The critical care-hospital interface
    Halpern, NA
    Brentjens, T
    CRITICAL CARE CLINICS, 1999, 15 (03) : 577 - +
  • [7] Connectivity-based cylinder detection in unorganized point clouds
    Araujo, Abner M. C.
    Oliveira, Manuel M.
    PATTERN RECOGNITION, 2020, 100
  • [8] Reliability of Critical Values Obtained By Point Of Care Glucose Testing
    Schifman, R.
    Renner, S.
    Cooper, E.
    Douros, E.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A98 - A98
  • [9] Connectivity-based error evaluation for ellipse fitting
    Masuzaki, Tomonari
    Sugaya, Yasuyuki
    2015 14TH IAPR INTERNATIONAL CONFERENCE ON MACHINE VISION APPLICATIONS (MVA), 2015, : 118 - 121
  • [10] Icodextrin:: A major problem for glucose dehydrogenase-based glucose point of care testing systems
    Flore, K.
    Delanghe, J.
    ACTA CLINICA BELGICA, 2006, 61 (06): : 351 - 354