Targeting accuracy of an image guided gating system for stereotactic body radiotherapy

被引:16
|
作者
Tenn, SE
Solberg, TD
Medin, PM
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[2] Jonsson Comprehens Canc Ctr, Los Angeles, CA 90034 USA
[3] Univ Nebraska, Med Ctr, Dept Radiat Oncol, Omaha, NE USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 23期
关键词
D O I
10.1088/0031-9155/50/23/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently, a commercial system capable of x-ray image guided patient positioning and respiratory gated delivery has become available. Here we describe the operational principles of this system and investigate its geometric targeting accuracy under controlled conditions. The system tracks breathing via infrared (IR) detection of reflective markers located on the patient's abdomen. Localization kilovoltage (kV) x-rays are triggered from within the gated delivery window portion of the breathing trace and after positioning, the tumour will cross the linac isocentre during gated delivery. We tested geometric accuracy of this system by localizing and delivering gated fields to a moving phantom. Effects of phantom speed, gating window location, timing errors and phantom rotations on positioning and gating accuracy were investigated. The system delivered gated fields to both a moving and static phantom with equal accuracy. The position of the gating window affects accuracy only to the extent that an asymmetric breathing motion could affect dose distribution within its boundaries. Positioning errors were found to be less then 0.5 +/- 0.2 mm for phantom rotations up to 5 degrees. We found and corrected a synchronization error caused by a faulty x-ray duration setting and detected a 60 +/- 20 ms time delay in our linear accelerator.
引用
收藏
页码:5443 / 5462
页数:20
相关论文
共 50 条
  • [31] CT-guided stereotactic targeting accuracy of osteoid osteoma
    Widmann, Gerlig
    Schullian, Peter
    Fasser, Martin
    Niederwanger, Christian
    Bale, Reto
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2013, 9 (03): : 274 - 279
  • [32] Image-Guided Liver Stereotactic Body Radiotherapy Using VMAT and Real-Time Adaptive Tumor Gating: Evaluation of the Efficacy and Toxicity for Hepatocellular Carcinoma
    Cantaloube, Marie
    Castan, Florence
    Creoff, Morgane
    Prunaretty, Jessica
    Bordeau, Karl
    Michalet, Morgan
    Assenat, Eric
    Guiu, Boris
    Pageaux, Georges-Philippe
    Ychou, Marc
    Ailleres, Norbert
    Fenoglietto, Pascal
    Azria, David
    Riou, Olivier
    CANCERS, 2021, 13 (19)
  • [33] Accuracy of a patient positioning system for image-guided frameless stereotactic IMRT treatments
    Dohm, O.
    Hoenes, A.
    Paulsen, F.
    Alber, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S288 - S288
  • [34] Image guided stereotactic approach of central nervous system lesions:: accuracy, morbidity, mortality
    Hisatugo, MKI
    Stávale, JN
    Bidó, JO
    Ferraz, FP
    ARQUIVOS DE NEURO-PSIQUIATRIA, 1999, 57 (3A) : 615 - 620
  • [35] Stereotactic Body Radiation Therapy for Thoracic Malignancies: An Analysis of MR-Guided Radiation Therapy Gating Tracking Accuracy
    Przybysz, D., Jr.
    Mazur, T. R.
    Gach, H. M.
    Robinson, C. G.
    Roach, M. C., Jr.
    Green, O. L.
    Mutic, S.
    Bradley, J. D.
    Cai, B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E665 - E666
  • [36] Image Guided Stereotactic Radiotherapy for the Treatment of Lung Oligometastatic Patients
    Masciullo, S.
    Masi, L.
    Menichelli, C.
    Bonucci, I.
    Doro, R.
    Casamassima, F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S509 - S510
  • [37] Image-guided stereotactic intensity modulated radiosurgery/radiotherapy
    Shiu, A
    Chang, E
    Ye, J
    Dong, L
    Lii, M
    Liu, H
    Maor, M
    Parker, B
    Court, L
    Mohan, R
    MEDICAL PHYSICS, 2002, 29 (06) : 1239 - 1239
  • [38] RELIABILITY OF THE BONY ANATOMY IN IMAGE-GUIDED STEREOTACTIC RADIOTHERAPY
    Guckenberger, M.
    Flentje, M.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S178 - S178
  • [39] Improving accuracy for stereotactic body radiotherapy treatments of spinal metastases
    Rijken, James
    Jordan, Barry
    Crowe, Scott
    Kairn, Tanya
    Trapp, Jamie
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 453 - 462
  • [40] Clinical feasibility of using an epid in cine mode for image-guided verification of stereotactic body radiotherapy
    Berbeco, Ross I.
    Hacker, Fred
    Ionascu, Dan
    Mamon, Harvey J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (01): : 258 - 266