Wavelength-Based Neutron/Gamma Ray Discrimination in CLYC

被引:0
|
作者
Bell, Zane W. [1 ]
Hornback, D. E. [1 ]
Hu, M. Z. [1 ]
Neal, J. S. [1 ]
机构
[1] Oak Ridge Natl Lab, Oak Ridge, TN 37831 USA
关键词
GAMMA-RAYS;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
We investigated the optical components of Cs(2)LiYC1(6):Ce (CLYC) as a possible path to particle discrimination by measuring the effects of optical filters on the shape of the waveform and comparing filtered waveforms to unfiltered ones. Crystals packaged with two optical windows were viewed by two photomultiplier tubes operated in coincidence to be certain that the signals from the same event were analyzed. We applied long-pass optical filters with cut-off wavelengths from 280 to 455 mu, and bandpass filters with cutoff wavelengths at 313, 350, 385, and 425 mu. Waveforms from gamma rays were significantly sharpened by the 313 nm bandpass filter because this filter cut out self-trapped exciton luminescence, while neutron waveforms were largely unaffected by filtering. Cumulative distribution functions generated from average filtered waveforms indicate that the neutron/gamma assignment can be made 44 ns by applying a Kolmogorov-Smirnov test comparing an acquired waveform to the average filtered waveforms.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Neutron measurement and inversion based on CLYC scintillator
    Hou, Donghui
    Zhang, Shenyi
    Yang, Yigang
    Wang, Qibiao
    Zhang, Binquan
    Yu, Qinglong
    Beijing Hangkong Hangtian Daxue Xuebao/Journal of Beijing University of Aeronautics and Astronautics, 2021, 47 (01): : 106 - 114
  • [22] Wavelength-based analog-to-digital conversion
    Zmuda, H
    Hayduk, MJ
    Bussjager, R
    Toughlian, EN
    PHOTONICS FOR SPACE AND RADIATION ENVIRONMENTS II, 2001, 4547 : 134 - 145
  • [23] A FAST NEUTRON TIME-OF-FLIGHT SPECTROMETER WITH NEUTRON-GAMMA RAY DISCRIMINATION
    WAKUTA, Y
    TAWARA, H
    HYAKUTAKE, M
    SONODA, M
    NUCLEAR INSTRUMENTS & METHODS, 1969, 71 (02): : 133 - +
  • [24] A low-power neutron and gamma-ray detector for environmental radiation monitoring using CLYC scintillator
    Zhao, K.
    Huang, Y.
    Shan, C.
    Feng, C.
    Wang, Z.
    Chang, K.
    Xie, W.
    Liu, F.
    Liu, S.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (09)
  • [25] Neutron and gamma-ray discrimination by a pressurized helium-4 based scintillation detector
    Dutta, Shubham
    Ghosh, Sayan
    Saha, Satyajit
    Journal of Instrumentation, 2024, 19 (11)
  • [26] Pulse Shape Discrimination for CLYC Based Handheld Instruments
    Gueorguiev, A.
    Glodo, J.
    Tower, J.
    Hawrami, R.
    Shirwadkar, U.
    O'Dougherty, P.
    Shah, K. S.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 32 - 37
  • [27] Design and Fabrication of CLYC-Based Rotational Modulation Collimator (RMC) System for Gamma-Ray/Neutron Dual-Particle Imager
    Kim, Hyun Suk
    Lee, Jooyub
    Choi, Sanghun
    Bang, Young-bong
    Ye, Sung-Joon
    Kim, Geehyun
    JOURNAL OF RADIATION PROTECTION AND RESEARCH, 2021, 46 (03): : 112 - 119
  • [28] PULSE SHAPE ANALYZER FOR FAST NEUTRON-GAMMA RAY DISCRIMINATION
    BRIENT, CE
    YOUNG, RL
    NELSON, CE
    NUCLEAR INSTRUMENTS & METHODS, 1972, 98 (02): : 329 - &
  • [29] Polysiloxane scintillators for neutron and gamma-ray pulse shape discrimination
    Lim, Allison
    Arrue, Jonathan
    Rose, Paul B., Jr.
    Sellinger, Alan
    Erickson, Anna
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XXII, 2020, 11494
  • [30] FINE STRUCTURE IN NEUTRON-GAMMA RAY PULSE SHAPE DISCRIMINATION
    STONGE, R
    AMOS, TM
    GALONSKY, A
    JOLLY, RK
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1971, 16 (10): : 1172 - &