The Traceability Improvement and Comparison of Bell Prover as the Indonesian National Standard of Gas Volume Flow Rate

被引:3
|
作者
Prakosa, J. A. [1 ]
Su, C. M. [2 ]
Wang, W. B. [2 ]
Sirenden, B. H. [1 ]
Zaid, G. [3 ]
Darmayanti, N. C. E. [3 ]
机构
[1] Indonesian Inst Sci LIPI, Res Ctr Phys, Tangerang Selatan 15314, Banten, Indonesia
[2] Ind Technol Res Inst ITRI, Ctr Measurement Stand CMS, 30 Ta Hsueh Rd, Hsinchu 30080, Taiwan
[3] Natl Standardizat Agcy Indonesia BSN, Tangerang Selatan 15314, Banten, Indonesia
来源
关键词
Improvement; Traceability; Comparison; Bell prover; Gas flow rate; Turbine;
D O I
10.1007/s12647-020-00402-4
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Traceability is urgent to guarantee the truth of measurement, particularly for National Standard. Although a bell prover facility has established as the national low pressure gas flow standard of Indonesia since 2008, the measurement capability (15 to 1400 L/min, U-95 = 0.20%) of the bell prover has not been traceable to SI Units. Moreover, displacement values provided by the bell prover's manufacturer by rotary encoder method are still used. The limit detector by optical sensor to measure displacement is worthy to achieve dimensional traceability which can be switched by ultrasonic sensor for reasonable solution. The aim of this study is to improve the traceability of the bell prover by reasonable solution and then validate the gas volume flow rate measurement by inter-laboratory comparison. The Center for Measurement Standards (CMS), Industrial Technology Research Institute (ITRI), was invited to participate in the bilateral comparison, and a gas turbine flow meter was used as the transfer standard. The bell prover has been successfully traceable to SI units, particularly displacement measurement through flying start/stop method by limit of ultrasonic sensor (250 to 1200 L/min, U-95 = 0.80% passing assessment report). Despite having larger uncertainty than manufacturer specification, the results are better for more in of layer comparison. Because the effect of turbine drift should be a concern, the gap of measurement time for inter-laboratory comparison is recommended to minimize as many as possible. Furthermore, the measurement range of bell prover should be reduced as reasonability purpose.
引用
收藏
页码:81 / 87
页数:7
相关论文
共 50 条
  • [1] The Traceability Improvement and Comparison of Bell Prover as the Indonesian National Standard of Gas Volume Flow Rate
    J. A. Prakosa
    C. M. Su
    W. B. Wang
    B. H. Sirenden
    G. Zaid
    N. C. E. Darmayanti
    MAPAN, 2021, 36 : 81 - 87
  • [2] Ensuring metrological traceability of volume units and gas flow rate to the National Measurement Standard Base of Ukraine
    Barchuk, A.
    Serediuk, D.
    Kepeshchuk, T.
    Lemishka, V.
    Pelikan, Yu.
    UKRAINIAN METROLOGICAL JOURNAL, 2025, (01): : 26 - 33
  • [3] Improvement and Analysis of Low Pressure Gas Flow Calibration System - Bell Prover
    Yen, Chih-Shan
    Lin, Ying-Chun
    Lin, Win-Ti
    Fan, Sheng-Cyuan
    19th International Flow Measurement Conference 2022, FLOMEKO 2022, 2022,
  • [4] Measurement methods for the outer diameter of bell type prover of gas flowmeter standard
    Wang, Chi
    Cui, Li-Shui
    Jiliang Xuebao/Acta Metrologica Sinica, 2007, 28 (03): : 239 - 242
  • [5] Design and Simulation of Automatic Control Valve for Gas Flow Meter Calibrator of Bell Prover
    Prakosa, Jalu. A.
    Kozlova, Liudmila P.
    PROCEEDINGS OF THE 2018 IEEE CONFERENCE OF RUSSIAN YOUNG RESEARCHERS IN ELECTRICAL AND ELECTRONIC ENGINEERING (EICONRUS), 2018, : 966 - 969
  • [6] Characterization of a piston prover gas flow standard in variable ambient temperature conditions
    Bobovnik, G.
    Zibret, P.
    Kutin, J.
    MEASUREMENT, 2021, 177
  • [7] Business Process Improvement Design of Indonesian National Standard (SNI) Product Certification
    Saputra, Mas'ud Adhi
    Meyilani, Rina
    PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON AUTOMATION, COGNITIVE SCIENCE, OPTICS, MICRO ELECTRO-MECHANICAL SYSTEM, AND INFORMATION TECHNOLOGY (ICACOMIT), 2015, : 46 - 51
  • [8] Leak rate calibration with gas flow standard apparatus
    Zhang, DX
    Liu, SL
    Li, DT
    Yang, M
    Liu, QA
    Li, WK
    VACUUM, 1996, 47 (6-8) : 515 - 517
  • [9] Comparisons of national standards in the field of gas flow rate and volume, gas flow rates from 20 to 6500 m3/h
    Mingaleev, A. V.
    Yakovlev, A. B.
    Mickan, B.
    Bardonov, A. I.
    Serediuk, D.
    Stankevicius, A.
    Busuladzic, I.
    METROLOGIA, 2020, 57 (1A)
  • [10] A primary standard for the volume flow rate of natural gas under high pressure based on laser Doppler velocimetry
    Mickan, B.
    Strunck, V.
    METROLOGIA, 2014, 51 (05) : 459 - 475