Intraoperative Cone Beam-Computed Tomography With Navigation (O-ARM) Versus Conventional Fluoroscopy (C-ARM)

被引:107
|
作者
Tabaraee, Ehsan [1 ]
Gibson, Anthony G. [1 ]
Karahalios, Dean G. [2 ]
Potts, Eric A. [3 ]
Mobasser, Jean-Pierre [3 ]
Burch, Shane [1 ]
机构
[1] Univ Calif San Francisco, San Francisco, CA 94143 USA
[2] Rush Med Coll, NorthShore Med, Chicago, IL 60612 USA
[3] Goodman Campbell Brain & Spine, Indianapolis, IN USA
关键词
radiation; O-ARM; image-guided surgery; radiation exposure; intraoperative imaging; PEDICLE SCREW PLACEMENT; RADIATION-EXPOSURE; IMAGE-GUIDANCE; THORACIC SPINE; ACCURACY; LUMBAR; THORACOLUMBAR; SURGEON; SYSTEM; COMPLICATIONS;
D O I
10.1097/BRS.0b013e3182a51d1e
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Cadaveric laboratory study. Objective. To compare the accuracy, efficiency, and safety of intraoperative cone beam-computed tomography with navigation (O-ARM) with traditional intraoperative fluoroscopy (C-ARM) for the placement of pedicle screws. Summary of Background Data. Radiation exposure remains a concern with traditional methods of intraoperative imaging in spine surgery. The use of O-ARM has been proposed for more accurate and efficient spinal instrumentation. Understanding radiation imparted to patients and surgeons by O-ARM is important for assessing risks and benefits of this technology, especially in light of evolving indications. Methods. Four surgeons placed 160 pedicle screws on 8 cadavers without deformity. Eighty pedicle screws were placed using O-ARM and C-ARM each. Instrumentation was placed bilaterally in the thoracic (T1-T6) spine and lumbosacral junction (L5-S1) using a standard open technique, whereas minimally invasive surgery technique was used at the lumbar 3 to 4 (L3-L4) level. A "postoperative" computed tomography (CT) scan was performed on cadavers where instrumentation was done using the C-ARM. An independent musculoskeletal radiologist assessed final images for screw position. Time required to set up and instrumentation was recorded. Dosimeters were placed on multiple aspects of cadavers and surgeons to record radiation exposure. Results. There were no differences in breach rate between the O-ARM and C-ARM groups (5 vs. 7,. 2 = 0.63, P = 0.4). The setup time for the O-ARM group was longer than that for the C-ARM group (592 vs. 297 s, P < 0.05). However, the average total time was statistically the same (1629 vs. 1639 s, P = 0.96). Radiation exposure was higher for surgeons in the C-ARM group and cadavers in the O-ARM group. When a "postoperative" CT scan was included in the estimation of the total radiation exposure, there was less of difference between the groups, but still more for the O-ARM group. Conclusion. In cadavers without deformity, O-ARM use results in similar breach rates as C-ARM for the placement of pedicle screws. Time for instrumentation is shorter with the O-ARM, but requires a longer setup time. The O-ARM exposes less radiation to the surgeon, but higher doses to the cadaver.
引用
收藏
页码:1953 / 1958
页数:6
相关论文
共 50 条
  • [31] O-arm navigation versus C-arm guidance for pedicle screw placement in spine surgery: a systematic review and meta-analysis
    Feng, Weili
    Wang, Weihao
    Chen, Shubiao
    Wu, Kezhou
    Wang, Hu
    INTERNATIONAL ORTHOPAEDICS, 2020, 44 (05) : 919 - 926
  • [32] Geometric calibration of a mobile C-arm for intraoperative cone-beam CT
    Daly, M. J.
    Siewerdsen, J. H.
    Cho, Y. B.
    Jaffray, D. A.
    Irish, J. C.
    MEDICAL PHYSICS, 2008, 35 (05) : 2124 - 2136
  • [33] Intraoperative assessment of hindfoot alignment using C-arm fluoroscopy
    Park, Young Uk
    Lee, Jong Wha
    Chung, Jun Young
    Choi, Wan-Sun
    Kim, Taehun
    Seo, Young Wook
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2024, 29 (05) : 1248 - 1254
  • [34] Use of robotic C-arm cone-beam computed tomography in surgical stabilization of rib fractures
    Yen, Ming-Hong
    Liu, Tsu-Hao
    Liu, Jung -Sen
    Yim, Shelly
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2023, 54 (12):
  • [35] Dosimetric Evaluation of O-Arm Cone Beam Computed Tomography (CBCT) Scan Modes for Pediatric Head Scans
    Shankar, A.
    Olguin, E.
    Bova, F.
    Arreola, M.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [36] Survey of terminology used for the intraoperative direction of C-arm fluoroscopy
    Pally, Elliott
    Kreder, Hans J.
    CANADIAN JOURNAL OF SURGERY, 2013, 56 (02) : 109 - 112
  • [37] Value of C-Arm Cone Beam Computed Tomography Image Fusion in Maximizing the Versatility of Endovascular Robotics
    Chinnadurai, Ponraj
    Duran, Cassidy
    Al-Jabbari, Odeaa
    Abu Saleh, Walid K.
    Lumsden, Alan
    Bismuth, Jean
    ANNALS OF VASCULAR SURGERY, 2016, 30 : 138 - 148
  • [38] C-arm cone-beam computed tomography in interventional oncology: technical aspects and clinical applications
    Floridi, Chiara
    Radaelli, Alessandro
    Abi-Jaoudeh, Nadine
    Grass, Micheal
    De Lin, Ming
    Chiaradia, Melanie
    Geschwind, Jean-Francois
    Kobeiter, Hishman
    Squillaci, Ettore
    Maleux, Geert
    Giovagnoni, Andrea
    Brunese, Luca
    Wood, Bradford
    Carrafiello, Gianpaolo
    Rotondo, Antonio
    RADIOLOGIA MEDICA, 2014, 119 (07): : 521 - 532
  • [39] C-arm cone-beam computed tomography in interventional oncology: technical aspects and clinical applications
    Chiara Floridi
    Alessandro Radaelli
    Nadine Abi-Jaoudeh
    Micheal Grass
    Ming De Lin
    Melanie Chiaradia
    Jean-Francois Geschwind
    Hishman Kobeiter
    Ettore Squillaci
    Geert Maleux
    Andrea Giovagnoni
    Luca Brunese
    Bradford Wood
    Gianpaolo Carrafiello
    Antonio Rotondo
    La radiologia medica, 2014, 119 : 521 - 532
  • [40] Application of C-arm computed tomography in cardiology
    Rieber, J.
    Rohkohl, C.
    Lauritsch, G.
    Rittger, H.
    Meissner, O.
    RADIOLOGE, 2009, 49 (09): : 862 - 867