Real-time automated thickness measurement of the in vivo human tympanic membrane using optical coherence tomography

被引:51
|
作者
Hubler, Zita [1 ]
Shemonski, Nathan D. [2 ,3 ]
Shelton, Ryan L. [2 ]
Monroy, Guillermo L. [1 ,2 ]
Nolan, Ryan M. [2 ]
Boppart, Stephen A. [1 ,2 ,3 ,4 ]
机构
[1] Univ Illinois, Dept Bioengn, Urbana, IL USA
[2] Univ Illinois, Beckman Inst Adv Sci & Technol, Urbana, IL USA
[3] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL USA
[4] Univ Illinois, Dept Internal Med, Urbana, IL USA
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
Optical coherence tomography (OCT); otitis media (OM); ear infection; tympanic membrane (TM); real-time; automated;
D O I
10.3978/j.issn.2223-4292.2014.11.32
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Otitis media (OM), an infection in the middle ear, is extremely common in the pediatric population. Current gold-standard methods for diagnosis include otoscopy for visualizing the surface features of the tympanic membrane (TM) and making qualitative assessments to determine middle ear content. OM typically presents as an acute infection, but can progress to chronic OM, and after numerous infections and antibiotic treatments over the course of many months, this disease is often treated by surgically inserting small tubes in the TM to relieve pressure, enable drainage, and provide aeration to the middle ear. Diagnosis and monitoring of OM is critical for successful management, but remains largely qualitative. Methods: We have developed an optical coherence tomography (OCT) system for high-resolution, depth-resolved, cross-sectional imaging of the TM and middle ear content, and for the quantitative assessment of in vivo TM thickness including the presence or absence of a middle ear biofilm. A novel algorithm was developed and demonstrated for automatic, real-time, and accurate measurement of TM thickness to aid in the diagnosis and monitoring of OM and other middle ear conditions. The segmentation algorithm applies a Hough transform to the OCT image data to determine the boundaries of the TM to calculate thickness. Results: The use of OCT and this segmentation algorithm is demonstrated first on layered phantoms and then during real-time acquisition of in vivo OCT from humans. For the layered phantoms, measured thicknesses varied by approximately 5 mu m over time in the presence of large axial and rotational motion. In vivo data also demonstrated differences in thicknesses both spatially on a single TM, and across normal, acute, and chronic OM cases. Conclusions: Real-time segmentation and thickness measurements of image data from both healthy subjects and those with acute and chronic OM demonstrate the use of OCT and this algorithm as a robust, quantitative, and accurate method for use during real-time in vivo human imaging.
引用
收藏
页码:69 / 77
页数:9
相关论文
共 50 条
  • [41] REAL-TIME PARYLENE COATING THICKNESS MEASUREMENT USING OPTICAL REFLECTOMETRY
    KIM, H
    POWERS, T
    MASON, J
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1982, 21 (03): : 900 - 901
  • [42] Real-Time In Vivo Assessment of Retinal Reattachment in Humans using Swept-Source Optical Coherence Tomography
    Bansal, Aditya
    Lee, Wei Wei
    Felfeli, Tina
    Muni, Rajeev H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [43] Real-Time In Vivo Assessment of Retinal Reattachment in Humans Using Swept-Source Optical Coherence Tomography
    Bansal, Aditya
    Lee, Wei Wei
    Felfeli, Tina
    Muni, Rajeev H.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2021, 227 : 265 - 274
  • [44] Optical assessment of the &ITin vivo&IT tympanic membrane status using a handheld optical coherence tomography-based otoscope
    Park, Kibeom
    Cho, Nam Hyun
    Jeon, Mansik
    Lee, Sang Heun
    Jang, Jeong Hun
    Boppart, Stephen A.
    Jung, Woonggyu
    Kim, Jeehyun
    ACTA OTO-LARYNGOLOGICA, 2018, 138 (04) : 367 - 374
  • [45] Real-Time Subglottic Stenosis Imaging Using Optical Coherence Tomography in the Rabbit
    Lin, Jennifer L.
    Yau, Amy Y.
    Boyd, Jonathon
    Hamamoto, Ashley
    Su, Erica
    Tracey, Lauren
    Heidari, Andrew E.
    Wang, Alex H.
    Ahuja, Gurpreet
    Chen, Zhongping
    Wong, Brian J.
    JAMA OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2013, 139 (05) : 502 - 509
  • [46] Choroidal thickness measurement in children using optical coherence tomography
    Garnier, M. Bidaut
    Schwartz, C.
    Puyraveau, M.
    Montard, M.
    Delbosc, B.
    Saleh, M.
    ACTA OPHTHALMOLOGICA, 2013, 91
  • [47] CHOROIDAL THICKNESS MEASUREMENT IN CHILDREN USING OPTICAL COHERENCE TOMOGRAPHY
    Bidaut-Garnier, Melanie
    Schwartz, Claire
    Puyraveau, Marc
    Montard, Michel
    Delbosc, Bernard
    Saleh, Maher
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2014, 34 (04): : 768 - 774
  • [48] Contactless measurement of fabric thickness using optical coherence tomography
    Sabuncu, Metin
    Ozdemir, Hakan
    JOURNAL OF THE TEXTILE INSTITUTE, 2022, 113 (05) : 713 - 717
  • [49] Optical Coherence Tomography of the Tympanic Membrane and Middle Ear: A Review
    Tan, Hsern Ern Ivan
    Maria, Peter Luke Santa
    Wijesinghe, Philip
    Kennedy, Brendan Francis
    Allardyce, Benjamin James
    Eikelboom, Robert Henry
    Atlas, Marcus David
    Dilley, Rodney James
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2018, 159 (03) : 424 - 438
  • [50] In-vivo sensing of the vibrations and thickness of the human tympanum with real-time profilometry using low-cost line-field spectral domain optical coherence tomography
    George, Denny Melkay M.
    Nandakumar, Hari
    Koushik, V
    Josyula, Sai Prashanth
    Chettri, Prajal
    Srivastava, Shailesh
    SADHANA-ACADEMY PROCEEDINGS IN ENGINEERING SCIENCES, 2024, 49 (02):