In vitro evaluation of current thoracic aortic stent-grafts for real-time MR-guided placement

被引:16
|
作者
Eggebrecht, H
Zenge, M
Ladd, ME
Erbel, R
Quick, HH
机构
[1] Univ Duisburg Essen, Dept Cardiol, D-45122 Essen, Germany
[2] Univ Duisburg Essen, Dept Diagnost & Intervent Radiol & Neuroradiol, W German Heart Ctr Essen, D-45122 Essen, Germany
关键词
magnetic resonance imaging; real-time MRI; MR guidance; thoracic aorta; stent-graft; susceptibility artifact;
D O I
10.1583/05-1707.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To systematically evaluate the magnetic resonance imaging (MRI) characteristics of current thoracic aortic stent-graft devices before, during, and after in vitro deployment as a step toward real-time MRI-guided stent placement. Methods: Six stent-graft devices used for thoracic aortic repair were examined in a dedicated phantom model using a 1.5-T MRI scanner. First, the delivery systems with the mounted stent-graft were examined using real-time fast imaging with steady-state precession (TrueFISP) with Cartesian and radial k-space filling. TrueFISP imaging was subsequently used for real-time monitoring of stent-graft expansion. The deployed stent-grafts were then examined in a water bath containing gadolinium (1:40) with high-resolution Tjweighted 3D fast low-angle shot (FLASH) sequences. The images were analyzed for artifacts, radiofrequency caging effects, and device visualization quality. Results: Three delivery systems with mounted stent-grafts did not contain ferromagnetic elements and were well visualized. Imaging with radial k-space filling showed fewer artifacts than Cartesian imaging. Movement of the delivery system and stent-graft expansion of these devices were successfully demonstrated at a rate of up to 6 frames per second. Evaluation of the expanded stent-grafts revealed only minor susceptibility artifacts without relevant signal attenuation in the stent-graft lumen for 5 nitinol-based stent-grafts. Only a stainless steel-based stent-graft was associated with severe artifacts, thwarting visualization of its lumen or surroundings. Conclusion: The present study shows that 3 nitinol-based thoracic stent-graft devices are potentially suited for real-time MRI-guided placement with respect to both the delivery system and the stent-graft itself. These observations provide the basis for the evaluation of MRI-guided stent-graft placement in vivo.
引用
收藏
页码:62 / 71
页数:10
相关论文
共 50 条
  • [1] Real-time MR fluoroscopy for MR-guided iliac artery stent placement
    Buecker, A
    Neuerburg, JM
    Adam, GB
    Glowinski, A
    Schaeffter, T
    Rasche, V
    van Vaals, JJ
    Molgaard-Nielsen, A
    Guenther, RW
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 12 (02) : 616 - 622
  • [2] Real-time MR fluoroscopy for MR-guided iliac artery stent placement
    Buecker, A
    Neuerburg, JM
    Adam, GB
    Glowinski, A
    Schaeffter, T
    Rasche, V
    van Vaals, JJ
    Molgaard-Nielsen, A
    Guenther, RW
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 12 (04) : 616 - 622
  • [3] Real-time MR-guided placement of an atrial septal defect closure device: In vitro experiments
    Buecker, A
    Freudenthal, FP
    Schaeffter, T
    Van Vaals, JJ
    Guenther, RW
    RADIOLOGY, 2001, 221 : 396 - 397
  • [4] Transluminal placement of endovascular stent-grafts for the treatment of descending thoracic aortic dissection or aneurysm
    Shim, WH
    Lee, DH
    Yoon, YS
    Lee, DY
    Cho, SY
    AMERICAN JOURNAL OF CARDIOLOGY, 1999, 84 (6A): : 43P - 44P
  • [5] Real-time MR-guided transarterial aortic valve implantation (tavi): in vivo evaluation in swine
    Quick H.H.
    Kahlert P.
    Eggebrecht H.
    Kaiser G.M.
    Parohl N.
    Albert J.
    Schäfer L.
    McDougall I.
    Erbel R.
    Ladd M.E.
    Journal of Cardiovascular Magnetic Resonance, 13 (Suppl 1)
  • [6] TRANSLUMINAL PLACEMENT OF ENDOVASCULAR STENT-GRAFTS FOR THE TREATMENT OF DESCENDING THORACIC AORTIC-ANEURYSMS
    DAKE, MD
    MILLER, DC
    SEMBA, CP
    MITCHELL, RS
    WALKER, PJ
    LIDDELL, RP
    NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (26): : 1729 - 1734
  • [7] Thoracic Aortic Stent-Grafts: Utility of Multidetector CT for Pre- and Postprocedure Evaluation
    Bean, Marchelle J.
    Johnson, Pamela T.
    Roseborough, Glen S.
    Black, James H.
    Fishman, Elliot K.
    RADIOGRAPHICS, 2008, 28 (07) : 1835 - 1851
  • [8] Real-time visualization of instruments for MR-guided interventions
    Ladd, M
    Debatin, J
    6TH WORLD CONGRESS OF ENDOSCOPIC SURGERY, PTS 1 AND 2, 1998, : 681 - 686
  • [9] Technical Challenges of Real-Time Adaptive MR-Guided Radiotherapy
    Thorwarth, Daniela
    Low, Daniel A.
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [10] Real-Time Catheter Tracking and Visualization in MR-Guided Brachytherapy
    Wang, W.
    Gao, Y.
    Mehrtash, A.
    Seethmaraju, R.
    Kapur, T.
    Vishwanathan, A.
    Schmidt, E.
    Cormack, R.
    MEDICAL PHYSICS, 2013, 40 (06)