Evaluation of spectrophotometric methods for determination of orthophosphates by sequential injection analysis

被引:41
|
作者
Munoz, A [1 ]
Torres, FM [1 ]
Estela, JM [1 ]
Cerda, V [1 ]
机构
[1] UNIV BALEARIC ISL,FAC SCI,DEPT CHEM,E-07071 PALMA DE MALLORCA,SPAIN
关键词
phosphate; sequential injection analysis; wastewater;
D O I
10.1016/S0003-2670(97)00286-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Three methods for the determination of orthophosphate have been studied by using the same sequential injection analysis manifold, namely the vanadomolybdate, the Malachite Green and the molybdenurn blue methods. The calibration graphs were linear up to 18.0, 0.4 and 4.0 mg P l(-1), respectively. Relative standard deviations and detection limits for vanadomolybdate, Malachite Green and molybdenum blue were 2.1%, 18% and 1.7% and 0.15, 0.01 and 0.01 mg Pl(-1), respectively. The three methods were applied to the determination of phosphate in several water samples, both from natural and waste origin at a frequency of about 30 determinations per hour.
引用
收藏
页码:21 / 29
页数:9
相关论文